FDA (Devices) recall action 76255

Resource Optimization & Innovation Llc recalled 135 products on 2017-04-05

The FDA (Devices) publishes a recall like this one as 135 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 135 pages.

Products
135
Announced
2017-04-05
Critical
0
Units
39,233

Why these products were recalled

Custom procedure trays contain a component, a light glove, that has been recalled by the manufacturer because it can split during application to a light handle breaking the sterile field.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–25 of 135

  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Laparotomy Procedure Trays Due to Glove Defect

    Resource Optimization & Innovation LLC is recalling 30 sterile laparotomy procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 880342004, Sterile, GS0776D - Laparotomy - OSU
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Lap PK Spohn Custom Procedure Trays

    Resource Optimization & Innovation LLC is recalling 3,000 sterile custom procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800401002, Sterile, GS0470B - Lap PK - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Robotic Hysterectomy Kits Due to Light Glove Splitting

    Resource Optimization & Innovation LLC is recalling 249 sterile robotic hysterectomy kits because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800416002, Sterile, GY0485B - Robotic Hysterectomy Pk - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Lap Chole Procedure Kits Due to Glove Defect

    Resource Optimization & Innovation LLC is recalling 2,016 sterile lap chole procedure kits because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800409002, Sterile, GS0478B - Lap Chole PK - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Custom Procedure Trays Due to Splitting Light Gloves

    Resource Optimization & Innovation LLC is recalling 96 Regard custom procedure trays because the included light gloves can split, breaking the sterile field.

    Product
    Regard, Item Number: 800413003, Sterile, CV0482C - Vascular PK - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Sterile Arthroscopy Kits Due to Glove Splitting Risk

    Resource Optimization & Innovation LLC is recalling 110 sterile arthroscopy kits because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800422003, Sterile, OR0491C - Arthroscopy Pk - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Spohn Laparoscopy Kits Due to Light Glove Splitting

    Resource Optimization & Innovation LLC is recalling 92 Spohn laparoscopy kits because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800417002, Sterile, GY0486B - Laparoscopy Pk - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Custom Procedure Trays Due to Glove Splitting

    Resource Optimization & Innovation LLC is recalling 280 sterile custom procedure trays because a light glove component can split during application, breaking the sterile field.

    Product
    Regard, Item Number: 800558004, Sterile, OR0647D - Extremity - Childrens Santa Rosa
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Custom Procedure Trays with Defective Light Gloves

    Resource Optimization & Innovation Llc is recalling 71 custom procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800569002, Sterile, CV0659B - Open Heart A&B - Medical Center Santa Rosa
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    FDA Recalls 460 Sterile Lap Chole Procedure Kits Due to Glove Defect

    Resource Optimization & Innovation LLC is recalling 460 sterile lap chole procedure kits because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800528005 Sterile, GS0612E - Lap Chole PK - Westover Hills
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Custom Procedure Trays Due to Light Glove Splitting

    Resource Optimization & Innovation LLC is recalling 356 Regard custom procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800557001, Sterile, LD0646A - C Section - Childrens Santa Rosa
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Custom Procedure Trays with Defective Light Gloves

    Resource Optimization & Innovation LLC is recalling 30 custom procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800562001, Sterile, NU0651A - Neuro - Childrens Santa Rosa
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Custom Procedure Trays Due to Glove Splitting

    Resource Optimization & Innovation Llc is recalling 48 kits of Regard sterile custom procedure trays because a light glove component can split during application, breaking the sterile field.

    Product
    Regard, Item Number: 880327005, Sterile, CV0758E - Open Heart A&B - OSU
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Abdominal Procedure Kits Due to Glove Defect

    Resource Optimization & Innovation LLC is recalling 210 sterile abdominal procedure kits because a light glove component may split during use, breaking the sterile field.

    Product
    Regard, Item Number: 800533001, Sterile, GS0606A - Abdominal Pk - Westover Hills
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Abdominal PK Procedure Trays Due to Glove Splitting

    Resource Optimization & Innovation Llc is recalling 290 sterile abdominal procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800533003 Sterile, GS0606C - Abdominal PK - Westover Hills
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Laparotomy Kits Due to Glove Splitting Risk

    Resource Optimization & Innovation LLC is recalling 78 sterile laparotomy kits because a light glove component may split, breaking the sterile field.

    Product
    Regard, Item Number: 880342003, Sterile, GS0776C - Laparotomy - OSU
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Ortho PK Spohn Procedure Trays

    Resource Optimization & Innovation LLC is recalling 840 sterile procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800407004, Sterile, OR0476D - Ortho PK - Spohn
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2017-04-05

    FDA Recalls 170 Sterile Laparoscopic Procedure Trays Due to Glove Splitting

    Resource Optimization & Innovation LLC is recalling 170 sterile laparoscopic procedure trays because a light glove component can split during application, breaking the sterile field.

    Product
    Regard, Item Number: 880341003, Sterile, GS0775C - Laparoscopic - OSU
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Sterile Arthroscopy Kits Due to Light Glove Splitting

    Resource Optimization & Innovation LLC is recalling 162 sterile arthroscopy kits because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800422004, Sterile, OR0491D - Arthroscopy Pk - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Spohn AV Access Procedure Kits Due to Glove Defect

    Resource Optimization & Innovation LLC is recalling 118 Spohn AV Access Procedure Kits because a light glove component may split during use, breaking the sterile field.

    Product
    Regard, Item Number: 800425002, Sterile, CV0494B - AV Access Pk - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Resource Optimization Recalls 15 Sterile Custom Procedure Trays

    Resource Optimization & Innovation LLC is recalling 15 sterile custom procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800526003 Sterile, ET0610C - Head and Neck - Westover Hills
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Regard Sterile Vaginal Kits Due to Glove Splitting Hazard

    Resource Optimization & Innovation LLC is recalling 148 sterile vaginal kits because a light glove component may split, breaking the sterile field.

    Product
    Regard, Item Number: 800418002, Sterile, GY0487B - Major Vaginal Pk - Spohn
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    FDA Recalls Robotic Hysterectomy Kits Due to Sterile Field Breach Risk

    Resource Optimization & Innovation LLC is recalling 236 robotic hysterectomy kits because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800567002, Sterile, GY0656B- Robotic Hysterectomy - Alamo
    Category
    Distribution
    2 states
  • HighFDA (Devices)··2017-04-05

    Recall of Custom Procedure Trays with Defective Light Gloves

    Resource Optimization & Innovation LLC is recalling 180 custom procedure trays because a light glove component can split, breaking the sterile field.

    Product
    Regard, Item Number: 800548002, Sterile, OR0637B - Extremity - Medical Center Santa Rosa
    Category
    Distribution
    2 states