The Recall Desk

FDA (Devices) recall action 76165

PerkinElmer Life and Analytical Sciences, Wallac, OY recalled 2 products on 2017-02-08

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2017-02-08
Critical
0
Units
—

Why these products were recalled

The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the defective axles. Ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2017-02-08

    PerkinElmer Spare Disk Remover Module Recall Due to Corrosion Risk

    PerkinElmer is recalling Spare Disk Remover Modules because defective axles may corrode in humid conditions, potentially contaminating test plates.

    Product
    Spare Disk Remover Module; Product Number: 61001820 Intended for in vitro quantitative and qualitative determination of analytes in body fluids.
    Category
    Distribution
    4 states
  • HighFDA (Devices)··2017-02-08

    PerkinElmer Genetic Screening Processor Recalled for Corroding Axles

    PerkinElmer is recalling 24 Genetic Screening Processor instruments because defective ball bushing axles may corrode in humid conditions, risking contamination of sample wells.

    Product
    Genetic Screening Processor (GSP); PerkinElmer Part Number= 2021-0010 Intended for in vitro quantitative and qualitative determination of analytes in body fluids.
    Category
    Distribution
    4 states