FDA (Devices) recall action 75157

Philips Medical Systems recalled 4 products on 2016-10-12

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2016-10-12
Critical
0
Units
1,302

Why these products were recalled

The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults t…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • ModerateFDA (Devices)··2016-10-12

    Philips Brilliance 64 CT Scanner Recalled for Infant Head Image Quality Issues

    Philips Medical Systems is recalling 185 Brilliance 64 CT scanners because infant head scans in High Resolution mode may show degraded image quality.

    Product
    Brilliance 64 CT Model Number 728231 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
    Category
    Distribution
    46 states
  • ModerateFDA (Devices)··2016-10-12

    Philips Ingenuity CT Scanner Recalled for Infant Head Image Quality Issues

    Philips Medical Systems is recalling 235 Ingenuity CT scanners because infant head scans in High Resolution mode may produce degraded image quality, including non-uniformity and CT number shifts.

    Product
    Ingenuity CT Model Number 728326 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
    Category
    Distribution
    46 states
  • ModerateFDA (Devices)··2016-10-12

    Philips Ingenuity Core 128 CT Scanner Recalled for Image Quality Defects

    Philips Medical Systems is recalling 524 Ingenuity Core 128 CT scanners because high-resolution head scans may produce degraded image quality, including non-uniformity and CT number shifts.

    Product
    Ingenuity Core 128 Model Number 728323 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
    Category
    Distribution
    46 states
  • ModerateFDA (Devices)··2016-10-12

    Philips CT Scanner Recall for Infant Head Image Quality Defects

    Philips is recalling 358 Ingenuity Core CT scanners because infant head scans may show degraded image quality, including non-uniformity and CT number shifts.

    Product
    Ingenuity Core Model Number 728321 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
    Category
    Distribution
    46 states