GE Healthcare, LLC recalled 2 products on 2016-06-15

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2016-06-15
Critical
0
Units
198

Why these products were recalled

The gas venting may not occur properly. A site reported a magnet quenching with subsequent venting of the cryogen gas into the magnet room. Cryogen gas that vents into the magnet room could cause critical deprivation of oxygen to users and patients.

The agency’s own wording, quoted from the enforcement report.