FDA (Devices) recall action 73099

Siemens Healthcare Diagnostics, Inc. recalled 2 products on 2016-03-02

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2016-03-02
Critical
0
Units
3,575

Why these products were recalled

False short clotting times for PT on BCS and BCS XP with Dade Innovin can occur. This has also an effect for false low PT INR values. These false short clotting times were observed only with turbid, hemolytic and/or icteric samples from intensive care patients. The corresponding reaction curves of these samples showed a pre-peak, which can lead to falsely short clotting time. The result was not flagged by the existing check algorithms.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2016-03-02

    Siemens BCS Blood Coagulation Analyzer Recall for False Results

    Siemens is recalling BCS Automated Blood Coagulation Analyzer Systems because they may produce falsely short clotting times and low PT INR values when testing specific types of patient samples.

    Product
    BCS Automated Blood Coagulation Analyzer System, Multipurpose system for in vitro coagulation studies, Catalog Numbers: 10454728 (BCS RECONDITIONED), 10454729 (BCS RECONDITIONED), 10454742 (BEHRING COAGULATION SYSTEM), 10459303 (BCS RECONDITIONED), 10460659 (BCS INSTRUMENT), 1046
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2016-03-02

    Siemens BCS XP Analyzer Recalled for False Clotting Results

    Siemens Healthcare Diagnostics is recalling BCS XP Automated Blood Coagulation Analyzer Systems because they may produce falsely short clotting times and low INR values for specific patient samples.

    Product
    BCS XP Automated Blood Coagulation Analyzer System, Multipurpose system for in vitro coagulation studies, Catalog Numbers: 10459330 (BCS XP SYSTEM, COMPLETE), 10462449 (BCS XP ANALYZER SYSTEM), 10461894 (BCS XP), 10470625 (BCS XP REFURBISHED); IVD --- Device Listing D060801
    Category
    Distribution
    0 states