The Recall Desk

FDA (Devices) recall action 73056

Cook Inc. recalled 4 products on 2016-03-02

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2016-03-02
Critical
0
Units
17,827

Why these products were recalled

A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. These catheters are used in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • SevereFDA (Devices)··2016-03-02

    Cook Inc. Recalls Single Lumen Central Venous Catheter Sets

    Cook Inc. is recalling 1,341 single-lumen central venous catheter sets and trays due to a manufacturing defect that may cause catheter tip fracture or separation.

    Product
    Single Lumen Central Venous Catheter Sets and Trays The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2016-03-02

    Cook Inc. Recalls Single Lumen Pressure Monitoring Sets Due to Tip Fracture

    Cook Inc. is recalling 2,280 single-lumen pressure monitoring sets and trays because a manufacturing defect may cause catheter tips to fracture or separate.

    Product
    Single Lumen Pressure Monitoring Sets and Trays The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2016-03-02

    Cook Inc. Recalls Radial Artery Pressure Monitoring Catheter Sets

    Cook Inc. is recalling radial artery pressure monitoring catheter sets because a manufacturing defect may cause the catheter tip to fracture or separate.

    Product
    Radial Artery Pressure Monitoring Catheter Sets and Trays The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.
    Category
    Distribution
    2 states
  • SevereFDA (Devices)··2016-03-02

    Cook Inc. Recalls Femoral Artery Pressure Monitoring Catheters Due to Tip Fracture Risk

    Cook Inc. is recalling specific femoral artery pressure monitoring catheter sets and trays because a manufacturing defect may cause the catheter tip to fracture or separate.

    Product
    Femoral Artery Pressure Monitoring Catheter Sets and Trays The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.
    Category
    Distribution
    2 states