Intuitive Surgical, Inc. recalled 2 products on 2016-01-20

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2016-01-20
Critical
0
Units
5,248

Why these products were recalled

This correction to the da Vinci Xi I&A User Manual communicates proper use of the da Vinci Xi 8 mm Endoscopes in combination with the da Vinci 8 mm and 12 mm cannulae, cannula reducers, and monopolar energy instruments. It informs customers of scenarios that may result in unintended patient harm due to capacitive coupling from monopolar energy sources.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2016-01-20

    Intuitive Surgical Recalls da Vinci Xi 8mm Endoscopes for Safety Manual Correction

    Intuitive Surgical is correcting the da Vinci Xi I&A User Manual to prevent unintended patient harm from capacitive coupling when using 8mm endoscopes with specific instruments.

    Product
    0 deg, 8mm Endoscope, IS4000: Model numbers: 470026-34, 470026-40, 470026-41. General and Plastic Surgery: The da Vinci Xi endoscope produces high definition (HD), visible light (VIS). The HD video is processed by the system electronics in the Vision Cart and displayed on…
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2016-01-20

    Intuitive Surgical Recalls da Vinci Xi Endoscopes for Capacitive Coupling Risk

    Intuitive Surgical is correcting the da Vinci Xi I&A User Manual to warn about unintended patient harm from capacitive coupling when using 8 mm endoscopes with specific instruments.

    Product
    30 deg, 8mm Endoscope, IS4000: Model numbers: 470027-34, 470027-40, 470027-41. General and Plastic Surgery: The da Vinci Xi endoscope produces high definition (HD), visible light (VIS). The HD video is processed by the system electronics in the Vision Cart and displayed on…
    Category
    Distribution
    1 state