The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2015-12-02
Critical
0
Units
2,274

Why these products were recalled

Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed

Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • SevereFDA (Devices)··2015-12-02

    NxStage Hemodialysis System Recalled for Ultrafiltration Volume Software Error

    NxStage Medical is recalling 7 units of the System One S Cycler hemodialysis system due to a software error that may cause the pump to continue removing fluid after the target volume is reached.

    Product
    NxStage System One S Cycler -High Permeability Hemodialysis System Model no. NX1000-4
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-12-02

    NxStage System One S Cycler Recalled for Ultrafiltration Software Error

    NxStage Medical is recalling 2,134 System One S Cycler units due to a software error that may cause the pump to continue removing fluid after the target volume is reached.

    Product
    NxStage System One S Cycler - Model no. NX1000-3. For home hemodialysis.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-12-02

    NxStage System One S Cycler Recalled for Ultrafiltration Software Error

    NxStage Medical is recalling 133 System One S Cycler units due to a software error that may cause the pump to continue removing fluid after the target volume is reached.

    Product
    NxStage System One S Cycler - Model no. NX1000-3-A (OUS). For home hemodialysis.