LifeHealth, LLC recalled 2 products on 2015-11-04

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2015-11-04
Critical
0
Units
—

Why these products were recalled

May report values outside the published performance specifications for pH. After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discrepancy is greater as shelf life progresses and can be seen on the aqueous controls before the blood.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2015-11-04

    LifeHealth Recalls BG Cartridges for IRMA TRUPOINT Blood Gas Analyzer

    LifeHealth is recalling BG cartridges for the IRMA TRUPOINT Blood Gas Analyzer because the pH sensor may provide falsely elevated results as the product ages.

    Product
    BG cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 048103. BG Cartridge are single use, disposable, in vitro diagnostics intended for point of care (POC), professional, use with the IRMA TRUpoint¿ Blood Analysis System.
    Category
    Distribution
    26 states
  • HighFDA (Devices)··2015-11-04

    LifeHealth Recalls IRMA TRUPOINT Blood Gas Analyzer CC Cartridges

    LifeHealth, LLC is recalling IRMA TRUPOINT Blood Gas Analyzer CC Cartridges because the pH sensor may provide falsely elevated results as shelf life progresses.

    Product
    CC cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 039903. The CC Cartridge are single use, disposable, in vitro diagnostics intended for point of care (POC), professional, use with the IRMA TRUpoint¿ Blood Analysis System.
    Category
    Distribution
    26 states