The Recall Desk

FDA (Devices) recall action 71907

Transonic Systems Inc recalled 10 products on 2015-09-09

The FDA (Devices) publishes a recall like this one as 10 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 10 pages.

Products
10
Announced
2015-09-09
Critical
0
Units
60

Why these products were recalled

Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–10 of 10

  • ModerateFDA (Devices)··2015-09-09

    Transonic Flow Probe Recalled Due to Software Defect

    Transonic Systems Inc is recalling 36 Transonic Flow Probes due to programming errors that cause incorrect meter readings.

    Product
    Transonic Flow Probe, 3 mm, Catalog Number HQC3FMC -- Product Usage: to measure flow intra-operatively.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-09-09

    Transonic Flow Probe Recalled Due to Software Defect in EPROM

    Transonic Systems Inc is recalling a 6 mm Transonic Flow Probe due to a software defect causing incorrect meter readings.

    Product
    Transonic Flow Probe, 6 mm, Catalog Number HQC6FMV -- Product Usage: to measure flow intra-operatively.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-09-09

    Transonic Flow Probe Recalled Due to Software Defect in EPROM

    Transonic Systems Inc is recalling 2 mm Flow Probes due to programming errors causing incorrect meter readings.

    Product
    Transonic Flow Probe, 2 mm, Catalog Number HQC2FMC Product Usage: to measure flow intra-operatively.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-09-09

    Transonic Flow Probe Recalled Due to Software Defect

    Transonic Systems Inc is recalling 3 mm Flow Probes due to a software defect that causes flow meters to display incorrect status messages.

    Product
    Transonic Flow Probe, 3 mm, Catalog Number HQC3MP -- Product Usage: to measure flow intra-operatively.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-09-09

    Transonic Flow Probe Recalled Due to Software Defect in EPROM

    Transonic Systems Inc is recalling 2 mm Flow Probes due to programming errors causing incorrect meter readings.

    Product
    Transonic Flow Probe, 2 mm, Catalog Number HQC2MP -- Product Usage: to measure flow intra-operatively.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-09-09

    Transonic Flow Probe Recalled Due to Software Defect

    Transonic Systems Inc is recalling 6 mm Flow Probes due to programming errors causing incorrect meter readings.

    Product
    Transonic Flow Probe, 6 mm, Catalog Number HQD6FMV -- Product Usage: to measure flow intra-operatively.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-09-09

    Transonic Flow Probe Recall Due to Software Defect in EPROM

    Transonic Systems Inc is recalling 2 mm Flow Probes due to programming errors causing incorrect meter readings.

    Product
    Transonic Flow Probe, 2 mm, Catalog Number HQD2FMC -- Product Usage: to measure flow intra-operatively.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-09-09

    Transonic Flow Probe Recalled Due to Software Defect

    Transonic Systems Inc is recalling 4 mm Flow Probes due to software defects causing incorrect meter readings.

    Product
    Transonic Flow Probe, 4 mm, Catalog Number HQD4FMC -- Product Usage: to measure flow intra-operatively.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-09-09

    Transonic Flow Probe Recalled Due to Software Defect

    Transonic Systems Inc is recalling 3 mm Flow Probes due to software defects causing incorrect readings on flow meters.

    Product
    Transonic Flow Probe, 3 mm, Catalog Number HQD3FMC -- Product Usage: to measure flow intra-operatively.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-09-09

    Transonic Flow Probe Recalled Due to Software Defect

    Transonic Systems Inc is recalling 8 mm Flow Probes due to a software defect that causes flow meters to display incorrect error messages.

    Product
    Transonic Flow Probe, 8 mm, Catalog Number HQC8FMV -- Product Usage: to measure flow intra-operatively.
    Category
    Distribution
    0 states