The Recall Desk

FDA (Devices) recall action 71904

Resource Optimization & Innovation Llc recalled 77 products on 2015-10-14

The FDA (Devices) publishes a recall like this one as 77 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 77 pages.

Products
77
Announced
2015-10-14
Critical
0
Units
38,056

Why these products were recalled

The kits contain an equipment glove component which may contain splits or holes compromising the sterility.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

26–50 of 77

  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Sterile Total Knee Surgical Kits

    Resource Optimization & Innovation LLC is recalling 200 sterile total knee surgical kits because the equipment glove component may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800553, Sterile, OR0642 - Total Knee - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a surgical…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls 872 Sterile Surgical Kits

    Resource Optimization & Innovation LLC is recalling 872 sterile surgical kits because a glove component may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800558, Sterile, OR0647 - Extremity - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a surgical…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Sterile Surgical Kits Due to Sterility Concerns

    Resource Optimization & Innovation LLC is recalling 1,233 sterile surgical kits because the equipment glove component may contain splits or holes that compromise sterility.

    Product
    regard Item Number: 880326, Sterile, GS0757 - Major Basin - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Sterile Open Heart Surgical Kits

    Resource Optimization & Innovation LLC is recalling 108 sterile open heart surgical kits because an equipment glove component may contain splits or holes that compromise sterility.

    Product
    regard Item Number: 800569, Sterile, CV0659 - Open Heart A & B - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • SevereFDA (Devices)··2015-10-14

    FDA Recalls Pacemaker Surgical Kits Due to Sterility Compromise

    Resource Optimization & Innovation LLC is recalling 40 pacemaker surgical kits because equipment gloves may have splits or holes that compromise sterility.

    Product
    regard Item Number: 880328, Sterile, HT0759 - Pacemaker Pk - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    FDA Recalls 164 Neuro Surgical Kits Due to Sterility Compromise

    Resource Optimization & Innovation LLC is recalling 164 sterile neuro surgical kits because equipment glove components may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800617, Sterile, NU0768 - Neuro Pk - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a surgical…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Sterile C-Section Surgical Kits

    Resource Optimization & Innovation LLC is recalling 144 sterile C-section surgical kits because equipment glove components may contain splits or holes that compromise sterility.

    Product
    regard Item Number: 800524, Sterile, LD0608 - C Section - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a surgical…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation LLC Recalls Sterile Surgical Kits

    Resource Optimization & Innovation LLC is recalling 1,144 sterile surgical kits because a glove component may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800533, Sterile, GS0606 - Abdominal Pk - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO Custom procedure tray contains surgical instruments and accessories intended for use during a surgical…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls 37 Sterile Open Heart Surgical Kits

    Resource Optimization & Innovation LLC is recalling 37 sterile open heart surgical kits because a glove component may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800569001, Sterile, CV0659A - Open Heart A&B - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Robotic Hysterectomy Surgical Kits

    Resource Optimization & Innovation LLC is recalling 30 sterile robotic hysterectomy surgical kits because the equipment glove component may have splits or holes that compromise sterility.

    Product
    regard Item Number 800567001, Sterile, GY0656A - Robotic Hysterectomy - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    FDA Recalls 88 Sterile Lap Chole Surgical Kits Due to Sterility Concerns

    Resource Optimization & Innovation LLC is recalling 88 sterile surgical kits because equipment glove components may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800560001, Sterile, GS0649A - Lap Chole - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Sterile Laparoscopy Surgical Kits

    Resource Optimization & Innovation LLC is recalling 400 sterile laparoscopy surgical kits because the equipment glove component may contain splits or holes that compromise sterility.

    Product
    regard Item Number: 800417, Sterile, GY0486 - Laparoscopy Pk - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Sterile Surgical Kits Due to Sterility Concerns

    Resource Optimization & Innovation LLC is recalling 24 sterile surgical kits because a component may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800597, Sterile, GY0688 - LAVH - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a surgical…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Sterile C-Section Surgical Kits

    Resource Optimization & Innovation LLC is recalling 26 sterile C-section surgical kits because the equipment glove component may contain splits or holes that compromise sterility.

    Product
    regard Item Number: 800091, Sterile, LD0692 - C Section - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a surgical…
    Category
    Distribution
    3 states
  • SevereFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Sterile Open Heart Surgical Kits

    Resource Optimization & Innovation LLC is recalling 199 sterile open heart surgical kits because equipment glove components may contain splits or holes that compromise sterility.

    Product
    regard Item Number: 800410, Sterile, CV0479 - Open Heart Pk - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • SevereFDA (Devices)··2015-10-14

    FDA Recalls Robotic Laparoscopic Surgical Kits Due to Sterility Compromise

    Resource Optimization & Innovation LLC is recalling 144 robotic laparoscopic surgical kits because the equipment glove component may contain splits or holes that compromise sterility.

    Product
    regard Item Number: 800599, Sterile, GS0694 - Robotic Lap - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a surgical…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls 124 Sterile Arthroscopy Surgical Kits

    Resource Optimization & Innovation LLC is recalling 124 sterile arthroscopy surgical kits because the equipment glove component may contain splits or holes that compromise sterility.

    Product
    regard Item Number: 800422, Sterile, OR0491 - Arthroscopy Pk - surgical kit containing cover light handle soft, Rx. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    FDA Recalls Pediatric Surgical Kits Due to Potential Sterility Compromise

    Resource Optimization & Innovation LLC is recalling 746 sterile pediatric surgical kits because equipment glove components may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800554, Sterile, GS0643 - Pediatric Basic - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    FDA Recalls Sterile Surgical Kits Due to Potential Sterility Compromise

    Resource Optimization & Innovation LLC is recalling 296 sterile surgical kits because an equipment glove component may contain splits or holes that compromise sterility.

    Product
    regard Item Number: 800418, Sterile, GY0487 - Major Vaginal Pk - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls 874 Sterile Surgical Kits

    Resource Optimization & Innovation LLC is recalling 874 sterile surgical kits because a glove component may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800414, Sterile, OR0483 - Total Joint Pk - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Sterile Orthopedic Surgical Kits

    Resource Optimization & Innovation LLC is recalling 1,476 sterile orthopedic surgical kits because equipment glove components may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800407, Sterile, OR0476 - Ortho Pk - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROI, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a surgical…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation LLC Recalls Sterile C-Section Surgical Kits

    Resource Optimization & Innovation LLC is recalling 144 sterile C-section surgical kits because equipment glove components may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800535, Sterile, LD0620 - C Section - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a surgical…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Sterile Open Heart Surgical Kits

    Resource Optimization & Innovation LLC is recalling 36 sterile Open Heart surgical kits because equipment glove components may contain splits or holes that compromise sterility.

    Product
    regard Item Number: 880327002, sterile, CV0758B - Open Heart A&B - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • SevereFDA (Devices)··2015-10-14

    FDA Recalls 64 Sterile Peripheral Vascular Surgical Kits Due to Sterility Compromise

    Resource Optimization & Innovation LLC is recalling 64 sterile surgical kits because equipment gloves may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800615, Sterile, CV0741 - Peripheral Vascular - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-10-14

    Resource Optimization & Innovation Recalls Sterile Hip Shoulder Surgical Kits

    Resource Optimization & Innovation LLC is recalling 92 sterile hip shoulder surgical kits because equipment glove components may have splits or holes that compromise sterility.

    Product
    regard Item Number: 800527, Sterile, OR0611 - Hip Shoulder Pk - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a…
    Category
    Distribution
    3 states