The Recall Desk

FDA (Devices) recall action 71534

American Medical Systems Innovation Center - Silicon Valley recalled 32 products on 2015-07-29

The FDA (Devices) publishes a recall like this one as 32 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 32 pages.

Products
32
Announced
2015-07-29
Critical
0
Units
398,040

Why these products were recalled

validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex'" Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products

validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products

validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sure flex'" Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

26–32 of 32

  • ModerateFDA (Devices)··2015-07-29

    FDA Recalls Fiber Stripper Tools Due to Cleaning Validation Issues

    American Medical Systems is recalling 16,710 Fiber Stripper tools because cleaning and sterilization validation data do not meet FDA guidance for reusable products.

    Product
    Tool, Stripper W/BLD, 44S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-07-29

    FDA Recalls Sureflex 273 Lithotripsy Fibers Due to Cleaning Validation Issues

    American Medical Systems is recalling Sureflex 273 Lithotripsy Fibers because cleaning and sterilization validation data do not meet FDA guidance for reusable products.

    Product
    Sureflex 273 Lithotripsy Fiber, 5x, Blue; Sure F!ex laser fibers are intended for use in laser-based surgical applications, including, but not limited to endoscopic, laparoscopic and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-07-29

    FDA Recalls Fiber Stripper Tools Due to Sterilization Validation Issues

    American Medical Systems is recalling 16,710 Fiber Stripper tools because cleaning and sterilization validation data do not meet FDA guidance for reusable medical devices.

    Product
    Fiber Stripper, 273¿m, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-07-29

    FDA Recalls Fiber Stripper Tools Due to Sterilization Validation Issues

    American Medical Systems is recalling Fiber Stripper tools because cleaning and sterilization validation data do not meet FDA guidance for reusable products.

    Product
    Fiber Stripper, 150¿m, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-07-29

    FDA Recalls Sureflex 365 Lithotripsy Fibers Due to Sterilization Validation Issues

    American Medical Systems is recalling 16,710 Sureflex 365 Lithotripsy Fibers because cleaning and sterilization validation data do not meet FDA guidance for reusable products.

    Product
    Sureflex 365 Lithotripsy Fiber, 5x, Blue; Sure F!ex laser fibers are intended for use in laser-based surgical applications, including, but not limited to endoscopic, laparoscopic and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-07-29

    FDA Recalls Disposable Fiber Strippers Due to Sterilization Validation Issues

    American Medical Systems is recalling 16,710 disposable fiber strippers because cleaning and sterilization validation data do not meet FDA guidance for reusable products.

    Product
    Stripper, 0.4mm Disposable EndoStat Fiber, The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-07-29

    FDA Recalls American Medical Systems Fiber Strippers for Validation Issues

    American Medical Systems is recalling 16,710 disposable fiber strippers because cleaning and sterilization validation data do not meet new FDA guidance for reusable products.

    Product
    Stripper, 0.3mm Disposable EndoStat Fiber, The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes…
    Category
    Distribution
    0 states