462

Why these products were recalled

An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+) is used, the ventilator may adjust the ventilation on the inaccurate flow measurement and decrease the amount of ventilation delivered.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2015-07-08

    Hamilton Medical Infant Flow Sensor Recalled for Inaccurate Ventilation Measurements

    Hamilton Medical is recalling Infant Flow Sensors because they may provide inaccurate flow measurements, causing ventilators to decrease ventilation in specific modes.

    Product
    Hamilton Medical Infant Flow Sensor, single use, (1.88m), Part number 155500. The infant flow sensor is an accessory for the Hamilton Medical Ventilators. Product Usage: The infant flow sensor is a device to measure patient air flow and pressures within the Hamilton Medical V
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-07-08

    Hamilton Medical Infant Flow Sensor Recalled for Inaccurate Ventilation Measurements

    Hamilton Medical is recalling Infant Flow Sensors because inaccurate flow measurements may cause ventilators to decrease ventilation delivery.

    Product
    Hamilton Medical Infant Flow Sensor, single use, (1.6m), Part number 260177. The infant flow sensor is an accessory for the Hamilton Medical Ventilators. Product Usage: The infant flow sensor is a device to measure patient air flow and pressures within the Hamilton Medical Ve
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-07-08

    Hamilton Medical Infant Flow Sensor Recalled for Inaccurate Ventilation Measurements

    Hamilton Medical is recalling Infant Flow Sensors because inaccurate flow measurements may cause ventilators to decrease ventilation delivery.

    Product
    Hamilton Medical Infant Flow Sensor, single use, (3.1m), Part number 260179. The infant flow sensor is an accessory for the Hamilton Medical Ventilators. Product Usage: The infant flow sensor is a device to measure patient air flow and pressures within the Hamilton Medical Ve