FDA (Devices) recall action 71134

Philips Medical Systems (Cleveland) Inc recalled 35 products on 2015-06-17

The FDA (Devices) publishes a recall like this one as 35 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 35 pages.

Products
35
Announced
2015-06-17
Critical
0
Units
4,590

Why these products were recalled

The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–25 of 35

  • HighFDA (Devices)··2015-06-17

    Philips Recalls Ingenuity PET/MR Uninterruptible Power Supply Accessory Devices

    Philips Medical Systems is recalling nine Ingenuity PET/MR Uninterruptible Power Supply (UPS) accessory devices due to risks of battery acid leakage, overheating, and fume emission.

    Product
    Ingenuity PET/MR (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • SevereFDA (Devices)··2015-06-17

    Philips Medical Systems Recalls Extended Brilliance Workstation UPS Accessories

    Philips Medical Systems is recalling Extended Brilliance Workstation UPS accessory devices due to risks of battery acid leakage, overheating, and fume emission.

    Product
    Extended Brilliance Workstation (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Precedence 16 UPS Accessory Devices Recalled for Battery Acid Leakage

    Philips Medical Systems is recalling 36 Precedence 16 UPS accessory devices used in CT and diagnostic imaging systems due to risks of battery acid leakage, overheating, and fume emission.

    Product
    Precedence 16 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Recalls Gemini 16 PET/CT Uninterruptible Power Supply Accessories

    Philips is recalling 7 Gemini 16 PET/CT Uninterruptible Power Supply units worldwide due to risks of battery acid leakage, overheating, and fumes.

    Product
    Gemini 16 PET/CT System (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Recalls GEMINI TF PET/CT 16 Slice UPS Accessory Devices

    Philips is recalling 69 GEMINI TF PET/CT 16 Slice systems due to UPS accessory devices that may leak battery acid, overheat, or emit fumes.

    Product
    GEMINI TF PET/CT 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Recalls BrightView SPECT UPS Accessory Devices for Safety Hazards

    Philips Medical Systems is recalling 277 BrightView SPECT UPS accessory devices due to risks of battery acid leakage, overheating, and fume emission.

    Product
    BrightView SPECT (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips CardioMD III UPS Accessory Devices Recalled for Battery Acid Leakage

    Philips Medical Systems is recalling 81 CardioMD III UPS accessory devices for CT and diagnostic imaging systems due to risks of battery acid leakage, overheating, and fume emission.

    Product
    CardioMD III (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Recalls CT Scanner UPS Accessories Due to Battery Acid Leakage Risk

    Philips Medical Systems is recalling 788 Uninterruptible Power Supply (UPS) accessories for CT scanners due to risks of battery acid leakage, overheating, and fume emission.

    Product
    Brilliance CT 64 channel with Essence, Technology (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Ingenuity TF UPS Accessory Devices Recalled for Battery Acid Leakage

    Philips Medical Systems is recalling 14 Ingenuity TF UPS accessory devices due to risks of battery acid leakage, overheating, and fume emission.

    Product
    Ingenuity TF (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Medical Systems Recalls MX 16 Slice UPS Accessory Devices

    Philips Medical Systems is recalling 125 MX 16 Slice UPS accessory devices due to risks of battery acid leakage, overheating, and fume emission.

    Product
    MX 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Recalls GEMINI TF Big Bore UPS Accessory Devices for Leakage Risk

    Philips Medical Systems is recalling 39 GEMINI TF Big Bore UPS accessory devices due to risks of battery acid leakage, overheating, and fume emission.

    Product
    GEMINI TF Big Bore (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips iCT SP UPS Accessory Devices Recalled for Battery Acid Leakage Risk

    Philips Medical Systems is recalling 110 iCT SP UPS accessory devices used with CT and diagnostic imaging systems due to risks of battery acid leakage, overheating, and fume emission.

    Product
    iCT SP (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • SevereFDA (Devices)··2015-06-17

    Philips Ingenuity Flex UPS Accessory Devices Recalled for Battery Acid Leakage

    Philips Medical Systems is recalling 23 Ingenuity Flex UPS accessory devices used with CT and diagnostic imaging systems due to risks of battery acid leakage, overheating, and fume emission.

    Product
    Ingenuity Flex (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • SevereFDA (Devices)··2015-06-17

    Philips Precedence 6 UPS Accessory Devices Recalled for Battery Acid Leakage

    Philips Medical Systems is recalling Precedence 6 UPS accessory devices for CT and diagnostic imaging systems due to risks of battery acid leakage, overheating, and fume emission.

    Product
    Precedence 6 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips CT Scanner UPS Accessory Recall for Battery Acid Leakage

    Philips is recalling 25 Brilliance CT 6 Slice Air UPS accessories due to risks of battery acid leakage, overheating, and fume emission.

    Product
    Brilliance CT 6 Slice Air (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Ingenuity CT Uninterruptible Power Supply Recall for Battery Acid Leakage

    Philips is recalling 156 Ingenuity CT Uninterruptible Power Supply (UPS) accessory devices due to a risk of battery acid leakage, overheating, and fume emission.

    Product
    Ingenuity CT (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips CT Scanner UPS Accessory Recall for Battery Acid Leakage

    Philips Medical Systems is recalling 847 UPS accessory devices for CT scanners due to risks of battery acid leakage, overheating, and fume emission.

    Product
    Brilliance CT 16 Slice Air (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Recalls Brilliance CT 40 Uninterruptible Power Supply Accessory Devices

    Philips Medical Systems is recalling 17 Brilliance CT 40 Uninterruptible Power Supply (UPS) accessory devices due to risks of battery acid leakage, overheating, and fume emission.

    Product
    Brilliance CT 40 - Slices (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • SevereFDA (Devices)··2015-06-17

    Philips Gemini GXL 16 Slice CT Systems Recalled for UPS Battery Hazards

    Philips is recalling 15 Gemini GXL 16 Slice CT systems worldwide because their UPS accessory devices may leak battery acid, overheat, or emit fumes.

    Product
    Gemini GXL 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Gemini LXL 16 Slice CT Systems Recalled for UPS Defects

    Philips is recalling 10 Gemini LXL 16 Slice CT systems worldwide due to UPS accessory defects that may cause battery acid leakage, overheating, or fumes.

    Product
    Gemini LXL 16 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips MX 16 Slice Phase II CT Systems Recalled for UPS Battery Hazards

    Philips is recalling 11 MX 16 Slice Phase II CT systems because their UPS accessory devices may leak battery acid, overheat, or emit fumes.

    Product
    MX 16 Slice Phase II (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips CT Scanner UPS Accessory Recall for Battery Acid Leakage Risk

    Philips Medical Systems is recalling two Diamond Select Brilliance CT 16 systems due to a risk of battery acid leakage, overheating, and fumes from the UPS accessory.

    Product
    Diamond Select Brilliance CT 16 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Recalls GEMINI TF PET/CT 64 Slice UPS Accessory Devices

    Philips Medical Systems is recalling 71 GEMINI TF PET/CT 64 Slice UPS accessory devices due to risks of battery acid leakage, overheating, and fume emission.

    Product
    GEMINI TF PET/CT 64 Slice (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Recalls BrightView XCT UPS Accessory Devices for Battery Acid Leakage

    Philips Medical Systems is recalling 337 BrightView XCT UPS accessory devices due to risks of battery acid leakage, overheating, and fume emission.

    Product
    BrightView XCT (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states
  • HighFDA (Devices)··2015-06-17

    Philips Medical Systems Recalls Ingenuity Core 128 UPS Accessory Devices

    Philips Medical Systems is recalling 288 Ingenuity Core 128 UPS accessory devices due to risks of battery acid leakage, overheating, and fume emission.

    Product
    Ingenuity Core 128 (Uninterruptible Power Supply (UPS) accessory devices) Computed Tomography X-ray systems and Diagnostic Imaging systems.
    Category
    Distribution
    3 states