The Recall Desk

FDA (Devices) recall action 71105

Philips Ultrasound, Inc. recalled 2 products on 2015-05-27

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2015-05-27
Critical
0
Units
—

Why these products were recalled

When EPIQ 5 Ultrasound System, WITHOUT the Pediatric Cardiology option, is set up to "Metric" and weight and/or height is entered, a unit conversion error may result in the incorrect display of their values and incorrect calculation of the patients Body Surface Area (BSA).

When EPIQ 7 Ultrasound System, WITHOUT the Pediatric Cardiology option, is set up to "Metric" and weight and/or height is entered, a unit conversion error may result in the incorrect display of their values and incorrect calculation of the patients Body Surface Area (BSA).

The agency’s own wording, quoted from the enforcement report.

1–2 of 2

  • HighFDA (Devices)··2015-05-27

    Philips EPIQ 5 Ultrasound System Recalled for Body Surface Area Calculation Error

    Philips is recalling EPIQ 5 Ultrasound Systems due to a unit conversion error that may cause incorrect Body Surface Area calculations when using Metric settings.

    Product
    EPIQ 5 Ultrasound System versions 1.3.2 or lower, WITHOUT the Pediatric Cardiology option, Model: EPIQ 5G, EPIC 5C, EPIQ 5W.Catalog Number: 795204 / 795205. Part Number: 989605408541 With 453561736761 (1.1) 453561750041 (1.1.1) 453561753651 (1.1.2) 453561772231 (1.2)…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2015-05-27

    Philips EPIQ 7 Ultrasound System Recalled for Unit Conversion Error

    Philips is recalling EPIQ 7 Ultrasound Systems due to a unit conversion error that may cause incorrect Body Surface Area calculations.

    Product
    EPIQ 7 Ultrasound System versions 1.3.2 or lower, WITHOUT the Pediatric Cardiology option, Model: EPIQ 7G, EPIC 7C, EPIQ 7W. Catalog Number: 795200 / 795201 Part Number: 989605386721 With: 453561726491 (1.0) 453561728121 (1.0.1) 453561736781 (1.1) 453561750021 (1.1.1)…
    Category
    Distribution
    0 states