FDA (Devices) recall action 71084

Philips Ultrasound, Inc. recalled 2 products on 2015-05-13

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2015-05-13
Critical
0
Units
—

Why these products were recalled

When Epiq 7 Ultrasound System, WITH the Pediatric Cardiology option, is set up to Metric and weight and/or height is entered, a unit conversion error may result in the incorrect calculation of the patients Body Surface Area (BSA).

When Epiq 5 Ultrasound System, WITH the Pediatric Cardiology option, is set up to Metric and weight and/or height is entered, a unit conversion error may result in the incorrect calculation of the patients Body Surface Area (BSA).

The agency’s own wording, quoted from the enforcement report.

1–2 of 2

  • HighFDA (Devices)··2015-05-13

    Philips EPIQ 7 Ultrasound System Recalled for Pediatric BSA Calculation Error

    Philips is recalling EPIQ 7 Ultrasound Systems with the Pediatric Cardiology option because a unit conversion error may cause incorrect Body Surface Area calculations.

    Product
    EPIQ 7 Ultrasound System with Pediatric Cardiology option, Model: EPIQ 7G, EPIC 7C, EPIQ 7W. Catalog Number: 795200 / 795201. Part Number: 989605386721 With: 453561726491 (1.0) 453561728121 (1.0.1) 453561736781 (1.1) 453561750021 (1.1.1) 453561753631 (1.1.2) 453561772251…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-05-13

    Philips EPIQ 5 Ultrasound Systems Recalled for Pediatric BSA Calculation Error

    Philips is recalling EPIQ 5 Ultrasound Systems with the Pediatric Cardiology option because a unit conversion error may cause incorrect Body Surface Area calculations.

    Product
    EPIQ 5 Ultrasound System with Pediatric Cardiology option, Model: EPIQ 5G, EPIC 5C, EPIQ 5W. Catalog Number: 795204 / 795205. Part Number: 989605408541 With 453561736761 (1.1) 453561750041 (1.1.1) 453561753651 (1.1.2) 453561772231 (1.2) 453561772611 (1.2.1) 453561786571…
    Category
    Distribution
    0 states