The Recall Desk

FDA (Devices) recall action 70924

GE Healthcare recalled 3 products on 2015-05-06

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2015-05-06
Critical
0
Units
18

Why these products were recalled

An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2015-05-06

    GE Healthcare Recalls Optima XR200 amx Mobile X-Ray Units

    GE Healthcare is recalling six Optima XR200 amx mobile X-ray units in the US due to potential mAs accuracy issues that may affect exposure times.

    Product
    GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR200 amx
    Category
    Distribution
    5 states
  • ModerateFDA (Devices)··2015-05-06

    GE Healthcare Recalls Six Optima XR220 amx Mobile X-Ray Units

    GE Healthcare is recalling six Optima XR220 amx mobile X-ray units in the US due to potential mAs accuracy inaccuracies.

    Product
    GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR220 amx
    Category
    Distribution
    5 states
  • HighFDA (Devices)··2015-05-06

    GE Healthcare Recalls Brivo XR285amx Mobile X-Ray Units

    GE Healthcare is recalling six Brivo XR285amx mobile X-ray units in the US due to potential inaccuracies in mAs accuracy checks.

    Product
    GE Healthcare Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx
    Category
    Distribution
    5 states