Alere San Diego, Inc. recalled 3 products on 2015-01-14

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2015-01-14
Critical
0
Units
463,934

Why these products were recalled

In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • SevereFDA (Devices)··2015-01-14

    Alere INRatio PT/INR Test Strips Recalled for Inaccurate Results

    Alere San Diego, Inc. is recalling INRatio PT/INR Test Strips because the system may provide INR results significantly lower than reference laboratory methods.

    Product
    Alere INRatio PT/INR Test Strips, Alere INRatio PT/INR System Professional - 0100071 Alere INRatio PT/INR Test Strips, Box of 12 - 0100139 Alere INRatio PT/INR Test Strips, Box of 48 The test strips are packaged in individually pouched test strips in a labeled box with 12…
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2015-01-14

    Alere INRatio Monitors Recalled for Potential Inaccurate INR Results

    Alere San Diego, Inc. is recalling INRatio PT/INR Monitor systems because they may provide INR results significantly lower than reference laboratory methods.

    Product
    INRatio Monitors - 0100004 Alere INRatio PT/INR System Professional - 0100007 INRatio Prothrombin Time (PT) Monitoring System - 0100137 Alere INRatio Replacement Monitor The test systems contain individually packaged components needed to use the product, such as a monitor…
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2015-01-14

    Alere INRatio 2 Monitors Recalled for Potential Inaccurate INR Results

    Alere San Diego, Inc. is recalling INRatio 2 PT/INR monitors because they may provide results significantly lower than reference laboratory methods.

    Product
    INRatio 2 Monitors - 0200431 Alere INRatio 2 PT/INR Professional Testing System - 0200432 Alere INRatio 2 PT/INR Home Monitoring System - 0200433 Alere INRatio 2 PT/INR Home Monitoring System - 0200457 Alere INRatio 2 Replacement Monitor - 55112 Alere INRatio 2 Replacement…