The Recall Desk

FDA (Devices) recall action 69838

Stryker Endoscopy recalled 207 products on 2014-12-24

The FDA (Devices) publishes a recall like this one as 207 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 207 pages.

Products
207
Announced
2014-12-24
Critical
0
Units
37,017,465

Why these products were recalled

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Stryker Endoscopy is recalling Laparoscopic manual instruments and accessories because the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods.

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

126–150 of 207

  • ModerateFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of atraumatic forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, ATRAUMATIC FORCEPS, DOUBLE ACTION, P/N 0250080071. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) por
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, SCHERTL GRASPER, FENESTRATED, 20CM, P/N 0250282006 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequen
    Category
    Distribution
    2 states
  • LowFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Graspers Due to Sterilization Documentation Issues

    Stryker Endoscopy is recalling 180,573 units of laparoscopic graspers because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, BLUNT WITH SPOON, GRASPER, 29CM, P/N 0250282055 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Tissue Grasper Forceps Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of tissue grasper forceps because the instructions for use do not support two specific sterilization methods.

    Product
    PKG, GRASPER FORCEPS, TISSUE, P/N 0250080284. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which a
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Maryland Dissecting Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Maryland Dissecting Forceps units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, MARYLAND DISSECTING FORCEPS, CURVED LEFT, 45CM, P/N 00250080578 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Laparoscopic Graspers Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, RETENTION GRASPER, 33CM, P/N 0250080697 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) por
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Instruments Due to Sterilization Labeling Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM X 20CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW, P/N 0250282104 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio fre
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Laparoscopic Manual Instruments Recalled for Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, PARK FORCEPS, P/N 0250080733 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which a
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Metzenbaum Scissors Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling Metzenbaum Scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, METZENBAUM SCISSORS, CURVED, LONG, P/N 0250080025. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) por
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, PADDLE BABCOCK, 45CM, P/N 0250080727 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Laparoscopic Forceps Recalled for Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 laparoscopic forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, DISSECTING & GRASPER FORCEPS, SERRATED, 45CM, P/N 0250080579 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have rad
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Park Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of Park Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, 45CM, PARK FORCEPS, P/N 0250080745 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports wh
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BIOPSY FORCEPS, PUNCH, P/N 0250080290. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which all
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Biopsy Punch Instruments Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 biopsy punch instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, BIOPSY PUNCH, 29CM, P/N 0250282060 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports wh
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Laparoscopic Forceps Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of bipolar forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080361. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports wh
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Metzenbaum Scissors Recalled Due to Sterilization Instructions Error

    Stryker Endoscopy is recalling Metzenbaum Scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, METZENBAUM SCISSORS, CURVED LONG, P/N 0250080265. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) port
    Category
    Distribution
    2 states
  • LowFDA (Devices)··2014-12-24

    Stryker Laparoscopic Manual Instruments Recalled for Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSULATED SHAFT 45CM, FOR BIPOLAR FORCEPS, P/N 0250080370. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequen
    Category
    Distribution
    2 states
  • LowFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls ATRAUMATIC FORCEPS Due to Sterilization Documentation Issues

    Stryker Endoscopy is recalling 180,573 units of ATRAUMATIC FORCEPS because the provided instructions do not support specific sterilization methods.

    Product
    PKG, ATRAUMATIC FORCEPS, SINGLE ACTION, P/N 0250080310. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) port
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Maryland Forceps Recalled Due to Sterilization Parameter Errors

    Stryker Endoscopy is recalling Maryland Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, MARYLAND FORCEPS, CURVED, LEFT, P/N 0250080042. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Laparoscopic Needleholder Recall Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 laparoscopic needleholders because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, STRAIGHT NEEDLEHOLDER, 29CM, P/N 0250282061 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, FUNDUS GRASPER, 29CPKG, 3MM INSERT, ALLIGATOR GRASPER, 29CM, P/N 0250282051 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instr
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Baby Metzenbaum Scissors Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Baby Metzenbaum Scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BABY METZENBAUM SCISSORS, CURVED, P/N 0250080028. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) port
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Laparoscopic Forceps Recalled for Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 laparoscopic dissecting forceps units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM X 45CM INSERT, DISSECTING FORCEPS, P/N 0250080762 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) p
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Fundus Grasper Recall Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Fundus Grasper instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, FUNDUS GRASPER, 20CM, P/N 0250282003 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports
    Category
    Distribution
    2 states
  • ModerateFDA (Devices)··2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Instruments Due to Sterilization Labeling Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, CANNULA 8MM AUTOMATIC VALVE STOP COCK, W/O TROCAR, P/N 0250080172 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio freq
    Category
    Distribution
    2 states