The Recall Desk

FDA (Devices) recall action 69234

Ortho-Clinical Diagnostics recalled 2 products on 2014-10-22

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2014-10-22
Critical
0
Units
—

Why these products were recalled

Ortho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the VITROS 4600 Chemistry Systems and VITROS Integrated Systems. Internal testing confirmed that when using calibrator barcode labels supplied with VITROS Chemistry Products Calibrator Kits 1, 2, 3, 4, 6, or 9, an unexpected assay calibration may occur if assay targets are unassigned (i.e., hidden).

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–2 of 2

  • ModerateFDA (Devices)··2014-10-22

    Ortho VITROS 5600 Integrated System Recalled for Calibration Anomaly

    Ortho-Clinical Diagnostics is recalling VITROS 5600 Integrated Systems due to a software anomaly that may cause unexpected assay calibration when using specific calibrator kits.

    Product
    VITROS 5600 Integrated System, System Product Code 6802413. For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents…
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2014-10-22

    Ortho-Clinical Diagnostics Recalls VITROS 4600 Chemistry Systems Due to Software Anomaly

    Ortho-Clinical Diagnostics is recalling VITROS 4600 Chemistry Systems because a software anomaly may cause unexpected assay calibration when using specific calibrator barcode labels.

    Product
    VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), System Product Code 6802445. The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products…
    Category
    Distribution
    1 state