The Recall Desk

FDA (Devices) recall action 68763

Medical Positioning Inc recalled 4 products on 2014-08-20

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2014-08-20
Critical
0
Units
40

Why these products were recalled

Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • HighFDA (Devices)··2014-08-20

    Medical Positioning Rapid Response HUT Table Recall for Malfunction Risk

    Medical Positioning Inc is recalling 20 Rapid Response HUT tables because a defective component may cause the table to drop into the Trendelenburg position too rapidly.

    Product
    Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1017. Product Usage: Used to conduct syncope examinations.
    Category
    Distribution
    18 states
  • HighFDA (Devices)··2014-08-20

    Medical Positioning Rapid Response HUT Table Recall for Malfunction Risk

    Medical Positioning Inc. is recalling 17 Rapid Response HUT tables because a defective component may cause the table to drop into the Trendelenburg position more rapidly than normal under excessive weight.

    Product
    Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217. Product Usage: Used to conduct syncope examinations.
    Category
    Distribution
    18 states
  • SevereFDA (Devices)··2014-08-20

    Medical Positioning Rapid Response HUT Table Recalled for Malfunction Risk

    Medical Positioning Inc is recalling two Model # 1019 HUT tables due to a component defect that may cause the table to drop into the Trendelenburg position more rapidly than normal.

    Product
    Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1019. Product Usage: Used to conduct syncope examinations.
    Category
    Distribution
    18 states
  • HighFDA (Devices)··2014-08-20

    Medical Positioning Rapid Response HUT Table Recalled for Malfunction Risk

    Medical Positioning Inc. is recalling Model # 1219 HUT tables because a defective component may cause the table to drop rapidly into the Trendelenburg position under excessive weight.

    Product
    Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1219. Product Usage: Used to conduct syncope examinations.
    Category
    Distribution
    18 states