The Recall Desk

FDA (Devices) recall action 67883

Ortho-Clinical Diagnostics recalled 2 products on 2014-05-07

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2014-05-07
Critical
0
Units

Why these products were recalled

Ortho Clinical Diagnostics (OCD) confirmed the potential for lower than expected C-Reactive Protein (CRP) results only to be generated when using VITROS Chemistry Products Specialty Diluent (Lot F3168) or VITROS Chemistry Products FS Diluent Pack 3 (Lot 01-3266). The VITROS Instructions for Use states that if the if a C-reactive protein concentration exceeds the systems measuring (reportable or dynamic) range, dilute the sample with VITROS Specialty Diluent or a patient sample containing a low concentration of CRP. An initial threefold dilution is recommended. Internal testing determined that…

Ortho Clinical Diagnostics (OCD) confirmed the potential for lower than expected C-Reactive Protein (CRP) results only to be generated when using VITROS Chemistry Products Specialty Diluent (Lot F3168) or VITROS Chemistry Products FS Diluent Pack 3 (Lot 01-3266). The VITROS Instructions for Use states that if the if a C-reactive protein concentration exceeds the systems measuring (reportable or dynamic) range, dilute the sample with VITROS Specialty Diluent or a patient sample containing a low concentration of CRP. An initial threefold dilution is recommended. Internal testing determined tha…

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–2 of 2

  • ModerateFDA (Devices)··2014-05-07

    Ortho-Clinical Diagnostics Recalls VITROS Specialty Diluent Due to Inaccurate CRP Results

    Ortho-Clinical Diagnostics is recalling VITROS Specialty Diluent because it may cause C-Reactive Protein test results to be approximately 51% lower than expected.

    Product
    VITROS Chemistry Products Specialty Diluent, REF 855 9825, 6 x 3 mL, IVD --- Ortho-Clinical Diagnostics, Inc. Rochester, NY
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2014-05-07

    Ortho-Clinical Diagnostics Recalls VITROS Chemistry Diluent Packs Due to Inaccurate Results

    Ortho-Clinical Diagnostics is recalling VITROS Chemistry Products FS Diluent Pack 3 because internal testing found it may cause C-Reactive Protein test results to be approximately 51% lower than expected when used for dilution.

    Product
    VITROS Chemistry Products FS Diluent Pack 3, REF 680 1754, 3 x 45 mL, (D1)/15 mL (D2), IVD --- Ortho-Clinical Diagnostics, Inc. Rochester, NY, 14626
    Category
    Distribution
    1 state