The Recall Desk

FDA (Devices) recall action 67871

Med-Mizer, Inc. recalled 3 products on 2014-05-14

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2014-05-14
Critical
0
Units
3,172

Why these products were recalled

All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on th

All configurations of the SS Retractabed, Clinical Contour, made prior to June 7, 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on t

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • SevereFDA (Devices)··2014-05-14

    Med-Mizer Retractabed Recall for Insufficient Welding on Foot Cross Tube

    Med-Mizer is recalling 191 Retractabed Clinical Contour adjustable beds manufactured before June 7, 2008, due to insufficient welding that could cause bed collapse.

    Product
    RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2 motors); 763 (3 motors); 763 A ( 3 motors, autocontour) 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2014-05-14

    Med-Mizer RetractaBed Recall for Insufficient Welding on Foot Cross Tube

    Med-Mizer is recalling RetractaBed Clinical Contour adjustable beds due to insufficient welding on the foot cross tube, which may cause bed collapse or rollers to detach.

    Product
    RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 799, 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2014-05-14

    Med-Mizer RetractaBed Recall Due to Insufficient Welding on Foot Cross Tube

    Med-Mizer is recalling 2,980 RetractaBed Clinical Contour adjustable beds due to insufficient welding on the foot cross tube, which may cause bed collapse.

    Product
    RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 801, 802, 803, 803A: 80 " bed with 1, 2, or 3, motors and autocontour 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
    Category
    Distribution
    0 states