The Recall Desk

FDA (Devices) recall action 67477

Smith & Nephew Inc. recalled 6 products on 2014-03-19

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2014-03-19
Critical
0
Units
2,423,739

Why these products were recalled

Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–6 of 6

  • HighFDA (Devices)··2014-03-19

    Smith & Nephew RENASYS-F XL Dressing Kits Recalled for Sealing Defects

    Smith & Nephew is recalling RENASYS-F XL with Soft Port dressing kits due to inadequate sealing that may prevent fluid evacuation and fail to trigger alarms.

    Product
    Smith & Nephew RENASYS-F XL with Soft Port REF #: 66800797 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2014-03-19

    Smith & Nephew RENASYS-G Gauze Dressing Kit Recall for Sealing Issues

    Smith & Nephew is recalling RENASYS-G Sterile Gauze Dressing Kits due to inadequate Soft Port sealing that may prevent fluid evacuation and fail to trigger alarms.

    Product
    Smith & Nephew RENASYS-G Sterile, Gauze Dressing Kit with Soft Port REF #'s: 66800961 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2014-03-19

    Smith & Nephew RENASYS-AB Abdominal Dressing Kit Recall for Sealing Defects

    Smith & Nephew is recalling RENASYS-AB Abdominal Dressing Kits because inadequate sealing may prevent fluid evacuation and fail to trigger pump alarms.

    Product
    Smith & Nephew RENASYS-AB Abdominal Dressing Kit with Soft Port REF #'s: 66800980 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2014-03-19

    Smith & Nephew RENASYS Soft Port Dressing Kits Recalled for Sealing Defects

    Smith & Nephew is recalling RENASYS Soft Port Dressing Kits because inadequate sealing may prevent wound fluid evacuation and fail to trigger pump alarms.

    Product
    Smith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2014-03-19

    Smith & Nephew Recalls RENASYS Gauze with Soft Port Kits

    Smith & Nephew is recalling RENASYS Gauze with Soft Port Kits because inadequate sealing can cause fluid to remain under the dressing and prevent the pump alarm from activating.

    Product
    Smith & Nephew RENASYS-Gauze with Soft Port Kit REF #'s: 66800933, 66800934, 66800935, and 66800936 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2014-03-19

    Smith & Nephew RENASYS-F Foam Dressing Kit Recall for Sealing Issues

    Smith & Nephew is recalling RENASYS-F Foam Dressing Kits with Soft Port due to inadequate sealing that may prevent fluid evacuation and fail to trigger alarms.

    Product
    Smith & Nephew RENASYS-F Foam Dressing Kit with Soft Port REF #'s: 66800794, 66800795, and 66800796 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
    Category
    Distribution
    31 states