Boston Scientific Corporation recalled 2 products on 2014-02-19

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2014-02-19
Critical
0
Units
70

Why these products were recalled

Boston Scientific is initiating this Recall Removal of one lot of the Coyote" OVER-THE-WIRE PTA Balloon Dilatation Catheter and one lot of the Coyote" MONORAIL" PTA Balloon Dilatation Catheter. Boston Scientific has become aware that the balloons of the devices from these lots may not deflate as expected during a procedure. To date, Boston Scientific has received three complaints for this issue. T

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2014-02-19

    Boston Scientific Recalls Coyote Balloon Dilation Catheters Due to Deflation Issues

    Boston Scientific is recalling specific lots of Coyote balloon dilation catheters because the balloons may not deflate as expected during procedures.

    Product
    Boston Scientific, 2.5mm Coyote" OVER-THE-WIRE PTA Balloon Dilation Catheter, UPN# H74939186251210, Catalog # 39186-25121.
    Category
    Distribution
    11 states
  • HighFDA (Devices)··2014-02-19

    Boston Scientific Recalls Coyote PTA Balloon Dilatation Catheters

    Boston Scientific is recalling Coyote PTA Balloon Dilatation Catheters because the balloons may not deflate as expected during procedures.

    Product
    Boston Scientific, Coyote" MONORAIL" PTA Balloon Dilatation Catheter, UPN # H74939185201510, Catalogue # 39185-20151.
    Category
    Distribution
    11 states