The Recall Desk

FDA (Devices) recall action 67184

Elekta, Inc. recalled 3 products on 2014-03-12

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2014-03-12
Critical
0
Units
889

Why these products were recalled

During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2014-03-12

    Elekta Agility Software Recall for Calibration Input Errors

    Elekta is recalling Agility software version 3.0 because incorrect values can be entered during calibration procedures for medical linear accelerators.

    Product
    Agility. Version 3.0 of the Integrity interface and control software for the Elekta range of medical digital linear accelerators which is associated with Agility collimators.
    Category
    Distribution
    35 states
  • HighFDA (Devices)··2014-03-12

    Elekta Agility software recall for potential calibration input errors

    Elekta is recalling Agility software version 3.1 for linear accelerators because incorrect values could be entered during calibration procedures.

    Product
    Agility. Version 3.1 of the Integrity interface and control software for the Elekta range of medical digital linear accelerators which is associated with Agility collimators.
    Category
    Distribution
    35 states
  • HighFDA (Devices)··2014-03-12

    Elekta Integrity Software Recalled for Calibration Input Errors

    Elekta, Inc. is recalling Integrity Version 1.1 software for linear accelerators because incorrect values can be entered during calibration procedures.

    Product
    Integrity. Version 1.1 of the Integrity interface and control software for the Elekta range of medical digital linear accelerators.
    Category
    Distribution
    35 states