Nellcor Puritan Bennett Inc. (dba Covidien LP) recalled 2 products on 2014-01-29

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2014-01-29
Critical
0
Units
78,119

Why these products were recalled

Covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator systems. The Operators Manual for PB840 states a nominal life of one year and the Addendum incorrectly states 2 years.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • LowFDA (Devices)··2014-01-29

    Puritan Bennett Ventilator Manual Addendum Corrected for Oxygen Sensor Life

    Nellcor Puritan Bennett is correcting a labeling error in the PB840 Operator's Manual Addendum that incorrectly stated the oxygen sensor life as two years instead of one.

    Product
    Puritan Bennett 840 Series Ventilator PB840 Operators Manual Addendum PN 066009A 09/02 The intended use of the oxygen sensor is to provide the oxygen concentration of the delivered gas at the patient wye (inspiratory-expiratory interface). The oxygen sensor does not control t
    Category
    Distribution
    4 states
  • ModerateFDA (Devices)··2014-01-29

    Puritan Bennett Ventilator Manual Labeling Correction for Oxygen Sensor Life

    Nellcor Puritan Bennett is correcting operator and service manuals for PB700 and PB840 ventilators to clarify the operational life of the oxygen sensor.

    Product
    Puritan Bennett 700 Series Ventilators (PB740 and PB760) PB700 Series Operators Manual: PN 10066984 & PN G-061874-00 and PB700 Series Service Manual: PN 10070389 The intended use of the oxygen sensor is to provide the oxygen concentration of the delivered gas at the patient
    Category
    Distribution
    4 states