The Recall Desk

FDA (Devices) recall action 66895

Asante Solutions, Inc. recalled 8 products on 2014-01-08

The FDA (Devices) publishes a recall like this one as 8 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 8 pages.

Products
8
Announced
2014-01-08
Critical
0
Units
6,977

Why these products were recalled

Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–8 of 8

  • HighFDA (Devices)··2014-01-08

    Asante Conset Infusion Set Recalled for Tubing Disconnects

    Asante Solutions is recalling Conset Infusion Sets because tubing may disconnect from the connector, causing insulin leakage and reduced delivery.

    Product
    Asante Conset Infusion Set with 110cm tubing & 6mm cannula packaged in 5-packs, singles,or inside a Combination Kit (FG-5246-4) Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2014-01-08

    Asante Comfort Infusion Set Recalled for Tubing Disconnects

    Asante Solutions is recalling 352 Comfort Infusion Sets because the tubing may disconnect from the connector, causing insulin leakage and reduced delivery.

    Product
    Asante Comfort Infusion Set with 110cm tubing & 17mm cannula packaged in 5-packs, singles, or inside a Combination Kit (FG-5147-4). Manufactured for: Asante Solutions Sunnyvale, CA; Manufactured by: Unomedical, Lejre, Denmark
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2014-01-08

    Asante Comfort Infusion Set Recalled for Tubing Disconnection and Insulin Leakage

    Asante Solutions is recalling 246 Comfort Infusion Sets because the tubing may disconnect from the connector, causing insulin leakage and reduced delivery to patients.

    Product
    Asante Comfort Infusion Set with 60cm tubing & 17mm cannula packaged in 5-packs, singles,or inside a Combination Kit (FG-5127-4). Manufactured for: Asante Solutions Sunnyvale, CA; Manufactured by: Unomedical Lejre, Denmark
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2014-01-08

    Asante Comfort Infusion Set Recalled Due to Tubing Disconnects

    Asante Solutions is recalling 834 Comfort Infusion Sets because tubing may disconnect from the connector, causing insulin leakage and reduced delivery.

    Product
    Asante Comfort Infusion Set with 60cm tubing & 13mm cannula packaged in 5-packs, singles,or inside a Combination Kit (FG-5123-4) Manufactured for: Asante Solutions Sunnyvale, CA; Manufactured by: Unomedical Lejre, Denmark
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2014-01-08

    Asante Conset Infusion Set Recalled for Tubing Disconnection and Insulin Leakage

    Asante Solutions is recalling Conset Infusion Sets because the tubing may disconnect from the connector, causing insulin leakage and delivering less insulin than intended.

    Product
    Asante Conset Infusion Set with 60cm tubing & 9mm cannula packaged in 5-packs,singles, or inside a Combination Kit (FG-5229-4); Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2014-01-08

    Asante Conset Infusion Set Recalled for Tubing Disconnects

    Asante Solutions is recalling 1,085 Conset Infusion Sets because tubing may disconnect from the connector, causing insulin leakage and under-dosing.

    Product
    Asante Conset Infusion Set with tubing & 9mm cannula packaged in 5-packs,singles,or inside a Combination Kit (FG-5249-4) Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2014-01-08

    Asante Conset Infusion Set Recalled for Tubing Disconnects

    Asante Solutions is recalling Conset Infusion Sets because tubing may disconnect from the connector, causing insulin leakage and under-dosing.

    Product
    Asante Conset Infusion Set with 60cm tubing & 6mm cannula packaged in 5-packs,singles,or inside a Combination Kit (FG-5226-4) Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2014-01-08

    Asante Comfort Infusion Set Recalled for Tubing Disconnection

    Asante Solutions is recalling 635 Comfort Infusion Sets because the tubing may disconnect from the connector, causing insulin leakage and reduced delivery.

    Product
    Asante Comfort Infusion Set with 110cm tubing & 13mm cannula packaged in 5-packs,singles, or inside a Combination Kit (FG-5143-4). Manufactured for: Asante Solutions Sunnyvale, CA; Manufactured by: Unomedical Lejre, Denmark
    Category
    Distribution
    0 states