The Recall Desk

FDA (Devices) recall action 66506

Medtronic Neuromodulation recalled 2 products on 2013-11-20

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2013-11-20
Critical
0
Units
126,797

Why these products were recalled

Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2013-11-20

    Medtronic RestoreUltra and RestoreSensor Neurostimulators Recalled for Loss of Stimulation

    Medtronic is recalling 64,163 RestoreUltra and RestoreSensor neurostimulators worldwide due to a risk of loss of stimulation when switching program groups.

    Product
    Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2013-11-20

    Medtronic Activa Neurostimulators Recalled for Loss of Stimulation

    Medtronic is recalling specific Activa rechargeable neurostimulators because they may lose stimulation when switching programs. Therapy can be restored using the Physician Recharge mode.

    Product
    Medtronic Activa PC, model 37601, and Activa RC, model 37612, Activa SC, models 37602 and 37603. Multi-program Rechargeable Neurostimulator for Deep Brain Stimulation.
    Category
    Distribution
    1 state