FDA (Devices) recall action 66306

Carestream Health Inc. recalled 50 products on 2013-11-20

The FDA (Devices) publishes a recall like this one as 50 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 50 pages.

Products
50
Announced
2013-11-20
Critical
0
Units
—

Why these products were recalled

Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a software problem.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–25 of 50

  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; MX WEB SVR W/3 MX-O LIC < 75K EXAMS The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Defect

    Carestream Health is recalling Vue PACS software because a defect may prevent some CT and MR images from loading into the viewer under specific conditions.

    Product
    Carestream Vue PACS; UNLIMITED READING LIC ADD 10K EX/YR The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Defect

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; UPG MX-O SVR 15K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software versions 11.3.2.0, 11.3.2.4, and 11.4 because specific multi-slice images may fail to load in the viewer.

    Product
    Carestream Vue PACS; UNLIMITED READING LIC < 25K EX/YR The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • HighFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; MX-O SA WORKSTATION SW LICENSE The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software versions 11.3.2.0, 11.3.2.4, and 11.4 because a defect may prevent some CT and MR images from loading in the viewer.

    Product
    Carestream Vue PACS; UPG MX-RAD SVR 75K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • HighFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; UPG MX-O SVR 25K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; MX WEB SVR W/3 MX-RAD LIC<15K EXAMS The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Defect

    Carestream Health is recalling Vue PACS software because a defect may prevent some CT and MR images from loading in the viewer under specific conditions.

    Product
    Carestream Vue PACS; UPG MX-O SVR 25K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; UNLIMITED READING LIC < 50K EX/YR The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; MX WEB SW/ 1 MX-O CONCUR ACCESS LIC The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; UPG MX-RAD SVR 7.5K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software versions 11.3.2.0, 11.3.2.4, and 11.4 because specific multi-slice images may fail to load in the viewer.

    Product
    Carestream Vue PACS; MX WEB SVR W/3 MX-RAD LIC<75K EX/YR The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software versions 11.3.2.0, 11.3.2.4, and 11.4 because specific multi-slice images may fail to load in the viewer.

    Product
    Carestream Vue PACS; Unlimited Reading LIC<15K EX/YR The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; MX WEB SVR W/3 MX-RAD LIC<25K EXAMS The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Defects

    Carestream Health is recalling Vue PACS software because specific multi-slice CT and MR images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; MX WEB SW/5 MX-RAD CONCUR USER LIC The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Defect

    Carestream Health is recalling Vue PACS software because a defect may prevent some CT and MR images from loading into the viewer.

    Product
    Carestream Vue PACS; MX WEB SVR W/3 MX-O LIC< 7.5K EXAMS The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Defect

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; Virtual PWR Reading Basic SW, 8 Pack The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software due to a defect that may prevent some CT and MR images from loading in the viewer.

    Product
    Carestream Vue PACS; UPG MX-O SVR 7.5K EX/YR UNLIMITED The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because specific multi-slice images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; MX WEB SVR W/3 MX-O LIC < 25K EXAMS The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Issues

    Carestream Health is recalling Vue PACS software because a defect may prevent some CT and MR images from loading in the viewer.

    Product
    Carestream Vue PACS; Virtual PWR Reading Basic SW StandAlone The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Defect

    Carestream Health is recalling Vue PACS software because specific multi-slice CT and MR images may fail to load in the viewer application.

    Product
    Carestream Vue PACS; Virtual Access Plus SW, Workgroup The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Defect

    Carestream Health is recalling Vue PACS software versions 11.3.2.0, 11.3.2.4, and 11.4 because specific multi-slice images may fail to load.

    Product
    Carestream Vue PACS; UNLIMITED READING LIC ADD 100K E/YR The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Defect

    Carestream Health is recalling Vue PACS software because a defect may prevent some CT and MR images from loading into the viewer.

    Product
    Carestream Vue PACS; MX WEB SVR W/3 MX-O LIC < 15K EXAMS The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states
  • ModerateFDA (Devices)··2013-11-20

    Carestream Vue PACS Software Recall for Image Loading Defect

    Carestream Health is recalling Vue PACS software because a defect may prevent some CT and MR images from loading into the viewer under specific conditions.

    Product
    Carestream Vue PACS; MX WEB SW/10 MX-RAD CONCUR USER LIC The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute…
    Category
    Distribution
    26 states