FDA (Devices) recall action 66171

Boston Scientific CRM Corp recalled 2 products on 2013-10-23

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2013-10-23
Critical
0
Units
77,000

Why these products were recalled

The performance of a low voltage capacitor in a subset of COGNIS CRT-Ds and TELIGEN ICDs manufactured prior to December 2009, may be compromised over time, causing increased current drain that can lead to premature battery depletion.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2013-10-23

    Boston Scientific COGNIS CRT-D Recall for Premature Battery Depletion

    Boston Scientific is recalling COGNIS CRT-D devices because a capacitor defect may cause premature battery depletion.

    Product
    Boston Scientific COGNIS CRT-D, Cardiac Resynchronization Therapy Defibrillator, models N118, N119. The following models are not available in the US: models N106, N107, P106, P107. Sterilized using ethylene oxide. Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112-5
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2013-10-23

    Boston Scientific TELIGEN ICDs Recalled for Premature Battery Depletion Risk

    Boston Scientific is recalling TELIGEN implantable cardiac defibrillators because a capacitor defect may cause premature battery depletion.

    Product
    Boston Scientific TELIGEN, Implantable Cardiac Defibrillator models E102, E110, the following models are not available in the US: models F102, F110, F111. Sterilized using ethylene oxide. Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Indicated for
    Category
    Distribution
    1 state