The Recall Desk

FDA (Devices) recall action 65893

Instrumentation Laboratory Co. recalled 7 products on 2013-09-25

The FDA (Devices) publishes a recall like this one as 7 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 7 pages.

Products
7
Announced
2013-09-25
Critical
0
Units
4,799

Why these products were recalled

Potential carryover issue that can can cause shortened APTT clotting times on the instrument.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–7 of 7

  • HighFDA (Devices)··2013-09-25

    Instrumentation Laboratory ACL TOP 700 CTS Analyzer Recall for Carryover Issue

    Instrumentation Laboratory is recalling 185 ACL TOP 700 CTS analyzers due to a potential carryover issue that may cause shortened APTT clotting times.

    Product
    ACL TOP 700 CTS; PN 0000280050 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems
    Category
    Distribution
    46 states
  • HighFDA (Devices)··2013-09-25

    Instrumentation Laboratory ACL TOP 300 Analyzers Recalled for Carryover Issue

    Instrumentation Laboratory is recalling 662 ACL TOP 300 and ACL TOP 300 CTS analyzers worldwide due to a potential carryover issue that may cause shortened APTT clotting times.

    Product
    ACL TOP 300 and ACL TOP 300 CTS (Japan); PN 0000280060 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or
    Category
    Distribution
    46 states
  • HighFDA (Devices)··2013-09-25

    Instrumentation Laboratory Recalls ACL TOP Analyzers Due to Carryover Issue

    Instrumentation Laboratory is recalling 929 ACL TOP analyzers because a potential carryover issue may cause shortened APTT clotting times.

    Product
    ACL TOP (Base); PN 0000280000 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems p
    Category
    Distribution
    46 states
  • HighFDA (Devices)··2013-09-25

    Instrumentation Laboratory Recalls ACL TOP 700 Analyzers for Carryover Issue

    Instrumentation Laboratory is recalling 311 ACL TOP 700 analyzers due to a potential carryover issue that may cause shortened APTT clotting times.

    Product
    ACL TOP 700; PN 0000280010 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems prov
    Category
    Distribution
    46 states
  • HighFDA (Devices)··2013-09-25

    Instrumentation Laboratory Recalls ACL TOP CTS Analyzers Due to Carryover Issue

    Instrumentation Laboratory is recalling 486 ACL TOP CTS analyzers because a potential carryover issue may cause shortened APTT clotting times.

    Product
    ACL TOP CTS; PN 0000280020 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems prov
    Category
    Distribution
    46 states
  • HighFDA (Devices)··2013-09-25

    Instrumentation Laboratory Recalls ACL TOP 700 LAS Analyzers for Carryover Issue

    Instrumentation Laboratory is recalling 56 ACL TOP 700 LAS analyzers due to a potential carryover issue that can cause shortened APTT clotting times.

    Product
    ACL TOP 700 LAS; PN 0000280030 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems
    Category
    Distribution
    46 states
  • HighFDA (Devices)··2013-09-25

    Instrumentation Laboratory ACL TOP 500 CTS Recall for Carryover Issue

    Instrumentation Laboratory is recalling ACL TOP 500 CTS analyzers due to a potential carryover issue that may cause shortened APTT clotting times.

    Product
    ACL TOP 500 CTS; PN 0000280040 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems
    Category
    Distribution
    46 states