The Recall Desk

FDA (Devices) recall action 65356

B. Braun Medical, Inc. recalled 2 products on 2013-12-04

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2013-12-04
Critical
0
Units
1,750,079

Why these products were recalled

Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2013-12-04

    B. Braun addEASE Binary Connector Recalled for Possible Solution Leakage

    B. Braun Medical is recalling addEASE Binary Connectors because a misaligned needle may cause solution to leak from the PAB bag.

    Product
    20mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC2000 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2013-12-04

    B. Braun Recalls addEASE Binary Connectors Due to Solution Leakage

    B. Braun Medical is recalling 4,000 addEASE Binary Connectors because a misaligned needle may cause solution to leak from the bag stopper.

    Product
    13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.
    Category
    Distribution
    0 states