The Recall Desk

FDA (Devices) recall action 65205

Surgical Instrument Service And Savings, Inc. recalled 22 products on 2013-07-10

The FDA (Devices) publishes a recall like this one as 22 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 22 pages.

Products
22
Announced
2013-07-10
Critical
0
Units
2,270

Why these products were recalled

The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–22 of 22

  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service Recalls Re-New II Endocut Scissor Tips

    Surgical Instrument Service and Savings, Inc. is recalling 10 Re-New II Endocut Scissor Tips because the sterile barrier seal is compromised, allowing the product to fall out of the pouch.

    Product
    - Re-New II Endocut Scissor Tip, 16.51mm 5.0mm Ref. 3142 Lot 122295, 122308, 122455, 122875 Manual surgical instrument for general use.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service Recalls Arthroscopic Electrodes Due to Seal Compromise

    Surgical Instrument Service and Savings, Inc. is recalling 447 units of arthroscopic electrodes and wands because the sterile barrier seal may be compromised, allowing product to fall out of the pouch.

    Product
    -90 Degree L-Hook Tip Arthroscopic Electrode, for use with 7205440 Switchpen, (Blue) 170mm. Ref. 7205441 Lot 122502 -Acromioplasty Electrode (Grey) 6.5" x 165mm. Ref. C5012A Lot 122288 -CAPSure 30 ArthroWand, CW, 30 Degree Shaft Angle, (Grey) 3.0mm. Ref. AC 1830-01 Lot…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls Relieva Flex Sinus Guide Catheter

    Surgical Instrument Service And Savings, Inc. is recalling Relieva Flex Sinus Guide Catheter and Luma Sentry 20 Tip Angle products because the sterile barrier seal may be compromised.

    Product
    - Relieva Flex Sinus Guide Catheter (Green) Tip Shape: F-70 Ref. GC070RF Lot 122429 - Relieva Luma Sentry 20 Tip Angle, 0.035 Diameter 100cm Ref. SIS100B Lot 122429 Manual surgical instrument for general use
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    TPS Micro Reciprocating Rasp Recalled Due to Compromised Sterile Seal

    Surgical Instrument Service And Savings, Inc. is recalling four units of TPS Micro Reciprocating Small Tear Cross Cut Rasps because the sterile barrier seal may be compromised.

    Product
    TPS Micro Reciprocating Small Tear Cross Cut Rasp 5.0mm x 11mm Ref. 5100-37-113 Lot 122518, 122447. Orthopedic manual surgical instrument
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls Scorpion Needles Due to Seal Compromise

    Surgical Instrument Service And Savings, Inc. is recalling specific lots of Scorpion Needles and Suture Passing Wires because the sterile barrier seal may be compromised, allowing product to fall out of the pouch.

    Product
    -Scorpion Needle, For Use With Scorpion Suture Passer, (Blue). Ref. AR-13990N Lot 123263 -SureFire Scorpion Needle (purple). Ref. AR-13991N Lot 123263 and 122279 -Suture Passing Wire. Ref. AR-1255-18 Lot 123190 Manual surgical instrument for general use.
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Barron Radial Vacuum Trephine Set Recalled Due to Compromised Sterile Seal

    Surgical Instrument Service and Savings, Inc. is recalling Barron Radial Vacuum Trephine Sets because the sterile seal may be compromised, allowing product to fall out of the pouch.

    Product
    Barron Radial Vacuum Trephine (1) Complete Trephine Set 9.0mm Dia. Ref. K20-2062 Lot 122336 Opthalmic
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Suture Recall Due to Compromised Sterile Barrier Seal

    Surgical Instrument Service and Savings, Inc. is recalling GII QuickAnchor Plus sutures because the sterile barrier seal may be compromised, allowing product to fall out of the pouch.

    Product
    GII QuickAnchor Plus, ETHIBOND Polyester Suture, 36", double-Armed w/CP 2 Needles 2. Ref. 212034 Lot 123266 General and Plastic Surgery
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls Phacofragmentation System Tips

    Surgical Instrument Service And Savings, Inc. is recalling specific phacofragmentation system tips because the sterile barrier seal became compromised, allowing the product to fall out of the pouch.

    Product
    - Turbosonics Mini-Flared ABS Tip, 45 Degree OZil 12 Degree, (Purplish/ Pink) 0.9mm Ref. 8065751177 Lot 123215 - TurboSonics¿ Mini-Flared ABS¿ Tip, 20 Degree Tip Angle, 45¿ Kelman, (Purplish Pink/Silver) 0.9mm Ref. 8065750853 Lot 122590, 123372 - TurboSonics¿ Tapered ABS¿…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Recall Due to Compromised Sterile Seal

    Surgical Instrument Service And Savings is recalling Large Tear Cross Cut Rasps because the sterile seal may be compromised, allowing the product to fall out of the pouch.

    Product
    Large Tear Cross Cut Rasp 14mm x 7mm Ref. 5100-37-114 Lot 122347 Manual surgical instrument for general use
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls Carpal Tunnel Release Blades

    Surgical Instrument Service And Savings, Inc. is recalling specific carpal tunnel release blades and kits because the sterile barrier seal was compromised, potentially allowing product to fall out of the pouch.

    Product
    -Instratek Carpal Tunnel Release Blades, (1) Triangle Blade, (1) Hook Blade. REF R2056 - LOT 122279 -Instratek Carpal Tunnel Release Hook Blade. REF 2055 Lot 122669 -Linvatec CTS Relief Kit. REF 9971, lot 122504 -Biomet Indiana Tome Blade for Carpal Tunnel Release…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service Recalls ENDOPATH Trocars Due to Compromised Sterile Seals

    Surgical Instrument Service and Savings, Inc. is recalling 259 ENDOPATH trocars and accessories because compromised sterile seals may allow product to fall out of the pouch.

    Product
    - ENDOPATH Resposable Bladeless Trocar, Non-Handled, w/Housing, 100mm, (Green) 10/11mm Ref. 511NA Lot 123215, 122647 - ENDOPATH Resposable Bladeless Trocar, Non-Handled, w/Housing, 100mm, (Orange) 5.0mm Ref. 35LNA Lot 123215, 122647 - ENDOPATH Resposable Trocar w/Dilating…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls Orthopedic Drill Bits

    Surgical Instrument Service And Savings, Inc. is recalling 10 orthopedic drill bits because the sterile barrier seal was compromised, allowing product to fall out of the pouch.

    Product
    - Cannulated Drill Bit for Mini TightRope Ref. AR-8911DC Lot 122511 - Drill Bit, Straight Shank 2.7mm x 12.7cm Ref. 2318-027-00 Lot 122322 - Drill Bit, w/Depth Mark, Quick Coupling (Red) 2.0mm x 140mm Ref. 323.062 Lot 122347 - Fast Cutting Drill Bit 2.0mm Ref. FDB 2.0…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls 15 Suture Instruments

    Surgical Instrument Service And Savings, Inc. is recalling 15 suture instruments because compromised seals may allow the sterile barrier to fail.

    Product
    - ConMed Linvatec Spectrum II Suture Hook Crescent, Medium (Teal) 4.0 x 20.0mm. REF C6386 Lot 122592. - Arthrex SutureLasso SD, 25 Degree Tight Curve Left, Wire Loop, (Green) 1.8mm. REF AR-4068-25TL Lot 122518 and Lot 123190. -- Arthrex SutureLasso SD, 25 Degree Tight…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls 907 Arthroscopic Instruments

    Surgical Instrument Service And Savings, Inc. is recalling 907 arthroscopic instruments because compromised sterile seals may allow product to fall out of the pouch.

    Product
    - Abrader Bur, Round (Aqua) 4.0mm Ref. 7205324 Lot 123270 - AcromioBlaster Burr (Brick Red) 5.5mm Ref. 7205669 Lot 122590, 122454, 122465, 122434, 122665 - AcromioBlaster Burr (Sage Green) 4.0mm Ref. 7205668 Lot 122590, 123270 - Acromionizer Bur, Barrel, (Mauve) 4.0mm…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls Suture Graspers Due to Seal Compromise

    Surgical Instrument Service And Savings, Inc. is recalling two lots of IDEAL Suture Graspers because the sterile barrier seal may be compromised, allowing the product to fall out of the pouch.

    Product
    - IDEAL 30 Suture Grasper (Green) 30 Degree Ref. 251721 Lot 122511 - IDEAL 60 Suture Grasper (Pink) 60 Degree Ref. 251723 Lot 122631 Passer
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls Reprocessed Medical Devices

    Surgical Instrument Service And Savings, Inc. is recalling 34 reprocessed medical devices because the sterile barrier seal became compromised, allowing product to fall out of the pouch.

    Product
    - ACCU-PASS Suture Shuttle, 70 Degree, Upbend, (Orange) Ref. 72200419 Lot 122504 - Carpal Tunnel Release Blade Assembly, (White) Ref. 81010 Lot 123263, 122406, 122482, 123043, 122518, 122511 - ENDO-SCRUB 2, (1) Sheath, 0 Degree, For K. Storz, Medtronic Sharpsite AC…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls Compression Sleeves Due to Seal Failure

    Surgical Instrument Service And Savings, Inc. is recalling 229 units of compression sleeves and garments because the sterile barrier seal may be compromised, allowing product to fall out of the pouch.

    Product
    - Calf Garment Standard Ref. L501-M Lot 122549, 122985 - Calf Garment, DVT10 Standard Ref. DVT10 Lot 122570, 12546 - Calf Garment, DVT20 Large Ref. DVT20 Lot 122549, 122570 - SCD Express Compression Sleeve, Knee Length, (White) Large Ref. 9789 Lot 122574 - SCD…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls 13 Medical Devices Due to Seal Compromise

    Surgical Instrument Service And Savings, Inc. is recalling 13 medical devices because compromised seals may allow products to fall out of sterile pouches.

    Product
    -ENDOPATH ETS-Flex Articulating Endoscopic Linear Cutter, (Blue) w/o Staples 35mm. REF ATB35 Lot 122300. -Ligaclip MCA Multiple Clip Applier, 20 Small Clips Small. REF MCS20 Lot 122300. - Ethicon Pistol Grip, Fixed Head Skin Stapler, Multi-Directional Release, (Grey) 35W.…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Harmonic ACE Curved Shear Recall Due to Compromised Sterile Seal

    Surgical Instrument Service And Savings, Inc. is recalling Harmonic ACE Curved Shears because the sterile seal may be compromised, allowing the product to fall out of the pouch.

    Product
    Harmonic ACE Ergonomically-Enhanced Curved Shear, Pistol Grip, Hand Controlled, (White) 5.0mm x 36cm Ref. ACE36E Lot 122295, 122308, 122762 Unclassified
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls 73 Medical Devices with Compromised Sterile Seals

    Surgical Instrument Service And Savings, Inc. is recalling 73 surgical instruments and accessories because their sterile seals may be compromised, allowing product to fall out of the pouch.

    Product
    - Average Medium Blade 9mm x 25mm x 0.38mm Ref. 2296-3-111 Lot 122590, 122447 - Cannulated Drill Bit, Quick Coupling, For 3.5 & 4mm Cannulated Screws, 1.35mm Cannulation 2.7mm x 160mm Ref. 310.67 Lot 122289 - Hall MicroChoice, Oscillating Blade, Fine, Flat, Stainless…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls Arthroscopic Cannulas Due to Seal Defect

    Surgical Instrument Service And Savings, Inc. is recalling 73 arthroscopic cannulas because a compromised seal may allow the product to fall out of the sterile pouch.

    Product
    - CLEAR-TRAC COMPLETE, Threaded Cannula with Obturator, (Grey) 7.0mm x 72mm Ref. 72200905 Lot 122631, 122609 - CLEAR-TRAC Flexible Cannula System Threaded, w/Obturator, (Lime Green) 8.0mm x 72mm Ref. 72200425 Lot 122631 - CLEAR-TRAC Flexible Cannula, Threaded, w/Obturator…
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2013-07-10

    Surgical Instrument Service And Savings Recalls 30 Surgical Instrument Motors and Accessories

    Surgical Instrument Service And Savings, Inc. is recalling 30 surgical instrument motors and accessories because the sterile barrier seal may be compromised, allowing product to fall out of the pouch.

    Product
    - Acorn Bur, Long, For Use With TPS Universal Drill, Long "AM" Attachment Head: 7.5mm Dia. x 9.5mm Length Ref. 5130-30-75 Lot 122665 - Cannulated Drill Bit, for 3.0mm Cannulated Screws, 1.2mm Cannulation, Quick Coupling 2.0mm x 150mm Ref. 310.221 Lot 122289 - Drill Bit…
    Category
    Distribution
    0 states