FDA (Devices) recall action 64128

Philips Healthcare Inc. recalled 2 products on 2013-06-26

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2013-06-26
Critical
0
Units
182

Why these products were recalled

An inspection shows several IATDs (Installation Acceptance Test could not be located or different templates have been used

An inspection of IATD (Installation Acceptance Test Document)IA SID (Source Image Distance) records shows table height measurements out of specification with the consequence of potential system mis-alignment

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • LowFDA (Devices)··2013-06-26

    Philips BuckyDiagnost Radiographic X-ray System Recalled for Documentation Issues

    Philips Healthcare Inc. is recalling 66 BuckyDiagnost Radiographic X-ray systems because installation acceptance test documentation was missing or inconsistent.

    Product
    Philips BuckyDiagnost Radiographic, a Diagnostic X-ray system. Catalog Numbers: 704031, 704032, 704035, 704060, 704062 This system is used for making x-ray exposures for diagnostics.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2013-06-26

    Philips BuckyDiagnost Radiographic Systems Recalled for Table Height Defects

    Philips Healthcare is recalling 116 BuckyDiagnost radiographic systems because table height measurements were found out of specification, potentially causing system mis-alignment.

    Product
    Philips BuckyDiagnost Radiographic. Catalog Numbers: 704031, 704032, 704035, 704060, 704062 This system is used for making x-ray exposures for diagnostics