The Recall Desk

FDA (Devices) recall action 63993

Hospira Inc. recalled 2 products on 2016-11-09

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2016-11-09
Critical
0
Units
35,596

Why these products were recalled

Hospira has received reports of Malfunction S205 Backup Battery Failure alarms in Symbiq v3.13 pumps. This issue to an incorrect installation of components on some of the Power Supply Controller (PSC) Boards distributed prior to September 2012. On pumps where incorrect installation of components has occurred, the backup battery power discharges and subsequently, the battery cannot be charged. Because of the battery depletion, or lack of charge, a visual and audible warning alarm, S205, is displayed. An S205 alarm is intended to notify the customer that the coin call battery, powering the backu…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2016-11-09

    Hospira Symbiq Infusion Pumps Recalled for Backup Battery Failure

    Hospira is recalling Symbiq Two Channel Infusers due to a backup battery failure caused by incorrect component installation on some power supply boards.

    Product
    Symbiq Two Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids; Product List Number 16027
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2016-11-09

    Hospira Recalls Symbiq Infusion Pumps Due to Backup Battery Failure

    Hospira is recalling 20,311 Symbiq One Channel Infusion Pumps due to a backup battery failure caused by incorrect component installation on some power supply boards.

    Product
    Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids; Product List Number 16026
    Category
    Distribution
    0 states