FDA (Devices) recall action 63383

Sorin Group USA, Inc. recalled 5 products on 2012-11-14

The FDA (Devices) publishes a recall like this one as 5 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 5 pages.

Products
5
Announced
2012-11-14
Critical
0
Units
47

Why these products were recalled

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determi

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–5 of 5

  • SevereFDA (Devices)··2012-11-14

    Sorin S5 Perfusion System Recalled for Motor Controller Fault

    Sorin Group USA is recalling S5 Perfusion Systems due to a motor controller fault that may cause unintended pump stoppage during surgery.

    Product
    S5 Double Roller Pump 85, Item Number: 10-85-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
    Category
    Distribution
    9 states
  • SevereFDA (Devices)··2012-11-14

    Sorin C5 Perfusion System Recalled for Motor Controller Fault

    Sorin Group USA recalled two Sorin C5 Perfusion Systems due to a motor controller fault that could cause pump stoppage during surgery.

    Product
    Sorin C5 Perfusion System, Item number 58-00-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
    Category
    Distribution
    9 states
  • SevereFDA (Devices)··2012-11-14

    Sorin S5 Perfusion System Recalled for Motor Controller Faults

    Sorin Group USA is recalling S5 Perfusion Systems due to a motor controller fault that could cause unintended pump stoppage during surgery.

    Product
    S5 Mast Roller Pump 150, Item Number: 10-88-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
    Category
    Distribution
    9 states
  • SevereFDA (Devices)··2012-11-14

    Sorin S5 Perfusion System Recalled for Motor Controller Fault

    Sorin Group USA is recalling S5 Perfusion Systems due to a motor controller fault that may cause unintended pump stoppage during surgery.

    Product
    S5 Mast Roller Pump 85, Item Number: 10-88-60 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
    Category
    Distribution
    9 states
  • SevereFDA (Devices)··2012-11-14

    Sorin S5 Perfusion System Recalled for Motor Controller Faults

    Sorin Group USA is recalling S5 Perfusion Systems due to motor controller faults that may cause unintended pump stoppage during surgery.

    Product
    S5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.
    Category
    Distribution
    9 states