The Recall Desk

FDA (Devices) recall action 61731

Terumo Cardiovascular Systems Corporation recalled 20 products on 2012-07-04

The FDA (Devices) publishes a recall like this one as 20 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 20 pages.

Products
20
Announced
2012-07-04
Critical
0
Units
1,062,780

Why these products were recalled

During production of the Sarns Antegrade Cardioplegia Cannula, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of a foreign substance on the inner surface of some cannulae tips. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface. The substance was likely deposited during the molding process, but the exact composition of the

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–20 of 20

  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannulas Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance was found on the inner surface of some tips that could be dislodged.

    Product
    X-COATED FX25RW H/L PACK Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannulas Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance found on the inner surface of some tips can be dislodged during use.

    Product
    Sarns Antegrade Cardioplegia Cannula: ROOT INFUSTION VENT CATHETER WITH 12 GAUGE BLUE NEEDLE RECTANGULAR FLANGE, 6" (15 CM) LONG Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Cardioplegia Cannula Kits Due to Foreign Substance

    Terumo is recalling Sarns Antegrade Cardioplegia Cannula kits because a foreign substance was found on the inner surface of some tips that could dislodge.

    Product
    CPB CATHETER KIT-DR VANTRIGHT Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannula Kits Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannula kits because a foreign substance on the cannula tips can be dislodged during use.

    Product
    CPB CATHETER KIT-DR BARTLES Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannula Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulae because a foreign substance was found on the inner surface of some tips that could be dislodged.

    Product
    CUSTOM CPB CATHETER KIT-DR. Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannulas Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance was found on the inner surface of some tips that can be dislodged.

    Product
    SARNS ANTEGRADE: CARDIOPLEGIA CANNULA: ROOT INFUSION VENT/CATHETER WITH 14- GAUGE WHITE NEEDLE, RECTANGULAR FLANGE, 6" (15 CM) LONG Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver…
    Category
    Distribution
    29 states
  • SevereFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannulas Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance was found on the inner surface of some cannula tips that can be dislodged.

    Product
    X-COATED SMALL PATIENT PACK Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Cardioplegia Cannula Kits Due to Foreign Substance

    Terumo is recalling Sarns Cardioplegia Cannula Kits because a foreign substance was found on the inner surface of some cannula tips that could be dislodged.

    Product
    CPB CATHETER KIT-DR GERHARDT Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannula Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance was found on the inner surface of some tips that could dislodge.

    Product
    X-COATED TRANSPLANT PACK Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannulas Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance was found on the inner surface of some tips that could be dislodged.

    Product
    SARNS ANTEGRADE CARDIOPLEGIA CANNULA: ROOT INFUSION CANNULA WITH 14-GAUGE WHITE NEEDLE,RECTANGULAR FLANGE AND SUTURE COLLAR, 6" (15CM) LONG Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Cardioplegia Cannulas Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance was found on the inner surface of some tips that can be dislodged.

    Product
    MAYO MN XCOATED 3/8X1/2 AV Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannula Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulae because a foreign substance was found on the inner surface of some tips that could be dislodged.

    Product
    XC HEART/LUNG PACK W/SX25RX Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannulas Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance was found on the inner surface of some tips that can be dislodged.

    Product
    X-COATED H/L PK W/FX15RW40 Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannulas Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance was found on the inner surface of some tips that could be dislodged.

    Product
    SARNS ANTEGRADE CARDIOPLEGIA CANNULA: ROOT INFUSION CANNULA WITH 12-GAUGE BLUE NEEDLE,RECTANGULAR FLANGE AND SUTURE COLLAR, 6" (15CM) LONG Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannula Kits Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannula kits because a foreign substance was found on the inner surface of some cannula tips that could be dislodged.

    Product
    CPB CATHETER KIT- DR OWEN Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • SevereFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannula Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance on the inner surface of some tips can be dislodged.

    Product
    X-COATED SMALL PATIENT PACK Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Cardioplegia Cannulas Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance on the inner surface of the tips can be dislodged during use.

    Product
    INDIANA/OHIO HEART Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during…
    Category
    Distribution
    29 states
  • SevereFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Antegrade Cardioplegia Cannula Kits Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannula kits because a foreign substance was found on the inner surface of some cannula tips that could dislodge.

    Product
    CPB CATHETER KIT-DR VANTRIGHT Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Cardioplegia Cannulas Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance was found on the inner surface of some tips that can be dislodged.

    Product
    CPB CATHETER KIT- DR HENDRICKSO Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta…
    Category
    Distribution
    29 states
  • HighFDA (Devices)··2012-07-04

    Terumo Recalls Sarns Cardioplegia Cannulas Due to Foreign Substance

    Terumo CVS is recalling Sarns Antegrade Cardioplegia Cannulas because a foreign substance on the inner surface may dislodge during use.

    Product
    FX PACK Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during cardiopulmonary…
    Category
    Distribution
    29 states