The Recall Desk

FDA (Devices) recall action 61480

Terumo Cardiovascular Systems Corporation recalled 2 products on 2012-07-18

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2012-07-18
Critical
0
Units
1,625

Why these products were recalled

Terumo Advanced Perfusion System 1 battery charge is being depleted faster than expected during periods of system shutdown. The System 1 may falsely indicate a longer battery life than actual battery time remaining if not subjected to regular use, such as would occur with a backup system. This may result in unexpected system shutdown. Prolonged battery depletion may cause permanent battery damage.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2012-07-18

    Terumo Advanced Perfusion System 1 Recalled for Battery Depletion Issues

    Terumo is recalling approximately 993 Advanced Perfusion System 1 units because the battery may deplete faster than expected during shutdown, potentially causing unexpected system failure.

    Product
    System 1 Base, 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional…
    Category
    Distribution
    41 states
  • HighFDA (Devices)··2012-07-18

    Terumo Advanced Perfusion System 1 Recalled for Battery Depletion Issues

    Terumo is recalling approximately 632 System 1 Base units because the backup battery depletes faster than expected, potentially causing unexpected system shutdowns.

    Product
    System 1 Base, 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who…
    Category
    Distribution
    41 states