The Recall Desk

Category

Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

2376–2400 of 39987

  • HighFDA (Devices)·Z-0715-2026·2025-12-03

    Vista set US9923F recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 74,652 Vista Basic sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Vista set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. VISTA BASIC PUMP IV SET W/ 15 DROP/ML;
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0651-2026·2025-12-03

    Pump administration sets recalled over piggyback backflow and priming failure

    B. Braun Medical is recalling 135,000 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET 3 CARESITE ULTRA ZER
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0604-2026·2025-12-03

    ADDitIV sets recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 100 ADDitIV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. V1782F UNIV IV SET W/0.2 MICRON FILTER; C
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0632-2026·2025-12-03

    Short extension sets recalled over piggyback backflow and occlusion

    B. Braun Medical is recalling 13,600 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXT SET W/ B/C VLV, ULTRAPORT, 57 I
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0665-2026·2025-12-03

    IV set 490489 recalled over piggyback backflow and inability to prime

    B. Braun Medical is recalling 3,120 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15DR, 92 IN W/
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0647-2026·2025-12-03

    Single lot IV sets recalled over backflow and priming occlusion

    B. Braun Medical is recalling 48 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN 15D, 2 B/C VLV, 3
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0668-2026·2025-12-03

    IV set 490504 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 36,936 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0630-2026·2025-12-03

    IV extension sets recalled over backflow between piggyback and primary containers

    B. Braun Medical is recalling 43,452 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. EXTENSION SET, 18 IN; Catalog Numb
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0606-2026·2025-12-03

    Anesthesia set 490189 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 32,436 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML, 108
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0582-2026·2025-12-03

    Demonstration only gel recalled because it may be placed in patients

    Straumann USA is recalling 61 units of Emdogain because product intended for demonstration purposes only could be placed in patients.

    Product
    Emdogain; Article Numbers: (1) 440.230V10, (2) 475.132V10;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0614-2026·2025-12-03

    Anesthesia set 490468 recalled over backflow and inability to prime

    B. Braun Medical is recalling 58,752 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 117 IN.; Catalog N
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0710-2026·2025-12-03

    Rate flow sets recalled over backflow and inability to prime

    B. Braun Medical is recalling 150,664 RATE FLOW sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    RATE FLOW set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. RATEFLOW ADMIN 15 DROP W/2 CARESITE
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0684-2026·2025-12-03

    Anesthesia set 490472 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 10,536 units anesthesia iv sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADM 15D 2 CARESITE, SAFE
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0587-2026·2025-12-03

    Q400 massage tables recalled over heat components that can overheat

    Pivotal Health Solutions is recalling 34 Quantum Intersegmental Tables, model Q400, because damage to the heat function components can lead to overheating and smoldering.

    Product
    Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0669-2026·2025-12-03

    High volume IV sets recalled over piggyback backflow and occlusion

    B. Braun Medical is recalling 190,625 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DROP W/ CAR
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0603-2026·2025-12-03

    Primary gravity sets recalled over backflow from secondary containers

    B. Braun Medical is recalling 43,900 ADDitIV primary gravity sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. Primary Gravity IV Set, 2 Non-Needle-Fre
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0626-2026·2025-12-03

    Caresite set 354207 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 381,850 Caresite primary gravity sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Primary Gravity Filtered Set with 2 CARE
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0588-2026·2025-12-03

    Heated massage tables recalled over heat components that can smolder

    Pivotal Health Solutions is recalling 464 Quantum Intersegmental Tables because damage to the heat function components can lead to overheating, smoldering and melting of the vinyl cushion.

    Product
    Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0611-2026·2025-12-03

    Anesthesia infusion sets recalled over backflow into the primary bag

    B. Braun Medical is recalling 19,392 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 15 DROPS/ML 125 I
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0692-2026·2025-12-03

    Large volume IV sets recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 45,288 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. 130 IN. IV ADMIN SET W/ 9 I
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0619-2026·2025-12-03

    Blood administration sets recalled over backflow and priming failure

    B. Braun Medical is recalling 164,568 blood administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Y-TYPE BLOOD SET, HAND P
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0661-2026·2025-12-03

    IV set 490309 recalled over backflow into the primary bag

    B. Braun Medical is recalling 259,392 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 2 CARESITE,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0685-2026·2025-12-03

    IV set 490312 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 226,632 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMN SET W/15 DROP, ULTR
    Category
    Medical Device
    Distribution
    0 states