The Recall Desk

Category

Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

22451–22475 of 40339

  • HighFDA (Devices)·Z-3211-2018·2018-09-26

    Accuray Recalls CyberKnife Calibration Film Due to Dimensional Inaccuracy

    Accuray is recalling 259 units of Ashland Box Film Ball Cube II calibration film because manufacturing defects may cause up to 0.5 mm of positional inaccuracy in CyberKnife systems.

    Product
    Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2762-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensor Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 1,025 Temperature Sensor Catheters because defective sensors may display lower body temperatures than actual patient readings.

    Product
    Temperature Sensor Catheter 18FR, Catalogue Number 102201101863BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3192-2018·2018-09-26

    Invivo Xper Flex Cardio Physiomonitoring System Recalled for Firmware Error

    Invivo Corporation is recalling 38 Xper Flex Cardio Physiomonitoring Systems distributed in the Netherlands due to incorrect firmware that may cause inaccurate ECG and heart rate displays.

    Product
    Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2781-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone, Ltd. is recalling 4,372 temperature sensing Foley catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Temperature sensing catheter 14FR, Catalogue Number 102203101463DO Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3186-2018·2018-09-26

    Polymedco PATHFAST NTproBNP Cartridges Recalled for Measurement Bias

    Polymedco is recalling 272 PATHFAST NTproBNP cartridges due to a manufacturing defect that may cause measurement values to be 12% to 20% higher than actual.

    Product
    PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole…
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-3209-2018·2018-09-26

    Exactech Recalls Optetrak Logic CC Stem Extension Due to Thread Defects

    Exactech is recalling one unit of the Optetrak Logic CC Stem Extension because it may have non-conforming internal threads.

    Product
    Optetrak Logic CC Stem Extension, 22mm x 160mm The OPTETRAK Logic CC Total Knee System is indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic…
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2991-2018·2018-09-26

    Mobius AIRO Mobile CT System Recalled for Software Fault

    Mobius Imaging is recalling 147 AIRO Mobile CT Systems because a software fault prevents the tube current modulation feature from working.

    Product
    AIRO Mobile CT System Model # MobiCT-32
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2755-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensor Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 11,488 temperature sensor Foley catheters because defective sensors may display lower body temperatures than actual.

    Product
    Temperature sensor Foley catheter 8FR, Catalogue Number 102205100880AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2776-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone, Ltd. is recalling 100 units of temperature sensing Foley catheters because defective sensors may display lower body temperatures than actual.

    Product
    Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101080MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3191-2018·2018-09-26

    Swissray Medical Recalls ddR Formula B X-ray Systems for Bucky Movement Risk

    Swissray Medical is recalling 27 ddR Formula B X-ray systems because the bucky component may move unexpectedly, posing a risk of injury to patients.

    Product
    ddR Formula B X-ray System, Model ddR Formula B X-ray system used for imaging
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2780-2018·2018-09-26

    Degania Silicone Recalls Thermistor Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling Thermistor Foley catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Thermistor Foley catheter, nonsterile, Catalogue Number 102201101463MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2775-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 500 sterile Foley catheters with defective temperature sensors that may display inaccurate readings.

    Product
    Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2784-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone, Ltd. is recalling 1,000 temperature sensing Foley catheters because defective sensors may display lower body temperatures than actual.

    Product
    Temperature sensing catheter 14FR, Catalogue Number 102201101463NU Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2773-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 500 temperature-sensing Foley catheters because defective sensors may display lower body temperatures than actual.

    Product
    Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101480MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2757-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensor Catheters Due to Defective Sensors

    Degania Silicone, Ltd. is recalling 1,588 Temperature Sensor Catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Temperature Sensor Catheter 8FR, Catalogue Number 102201100863BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2777-2018·2018-09-26

    Degania Silicone Recalls Thermistor Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 2,784 Thermistor Foley catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Thermistor Foley catheter, nonsterile, Catalogue Number 102201101863MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2764-2018·2018-09-26

    Degania Foley Catheters Recalled for Defective Temperature Sensors

    Degania Silicone is recalling 47,520 Foley catheters with temperature sensors because defective sensors may display lower body temperatures than actual.

    Product
    Foley Catheter with temperature sensor 400TM 12FR, Catalogue Number 102201101280TY Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2760-2018·2018-09-26

    Degania Temperature Sensor Catheters Recalled for Defective Sensors

    Degania Silicone is recalling 16,350 temperature sensor catheters because defective sensors may display lower body temperatures than actual.

    Product
    Temperature Sensor Catheter 14FR, Catalogue Number 102201101463BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3183-2018·2018-09-26

    Med-Mizer Recalls PR1000 Pivot Rehab Beds Due to Cable Pinching Risk

    Med-Mizer is recalling 230 PR1000 Pivot Rehab Beds because wires may become pinched during extreme left pivoting, requiring patient transfer until repaired.

    Product
    PR1000 Pivot Rehab Bed AC powered adjustable bed.
    Category
    Medical Device
    Distribution
    28 states
  • HighFDA (Devices)·Z-2759-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensor Catheters Due to Defective Sensors

    Degania Silicone is recalling 1,028 temperature sensor catheters because defective sensors may display lower body temperatures than actual.

    Product
    Temperature Sensor Catheter 12FR, Catalogue Number 102201101263BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3181-2018·2018-09-26

    Stryker Recalls IsoFlex LAL Support Surfaces Due to Fire Barrier Defect

    Stryker is recalling 243 IsoFlex LAL support surfaces because they were assembled with a non-conforming fire barrier. The units were distributed in five US states.

    Product
    IsoFlex LAL Support Surface, Model Number 2860 Product Usage: : The IsoFlex LAL support surface assists in the prevention and treatment of all pressure injuries (including stages 1,2,3,4, unstageable injury, and deep tissue injury) and is recommended to be implemented in…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2758-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensor Foley Catheters Due to Defective Sensors

    Degania Silicone is recalling 475 Temperature Sensor Foley Catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Temperature Sensor Catheter 10FR, Catalogue Number 102201101063BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2785-2018·2018-09-26

    Degania Silicone Recalls Temperature Sensing Foley Catheters Due to Defective Sensors

    Degania Silicone, Ltd. is recalling 1,000 temperature sensing Foley catheters because defective sensors may display lower body temperatures than the patient's actual temperature.

    Product
    Temperature sensing catheter 16FR, Catalogue Number 102201101663NU Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder…
    Category
    Medical Device
    Distribution
    Distributed nationwide

Looking for a different category? Browse all recalls.