The Recall Desk

Category

Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

14101–14125 of 40151

  • HighFDA (Devices)·Z-2116-2021·2021-07-28

    Hip Implant Acetabular Liners Recalled for Premature Wear Risk

    Exactech Connexion GXL acetabular polyethylene liners used in orthopedic hip implants are recalled due to risk of edge-loading and premature wear in certain configurations. The recall affects 89,050 units distributed nationwide.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-28-XX, Acumatch GXL 0 Degree Liner, 28mm; Novation GXL Liner, Neutral, 28mm; and Novation GXL Neutral Liner, G0, 28mm ID. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2096-2021·2021-07-28

    Mediana Vital Signs Monitor May Fail to Power Up or Respond

    Mediana AVSM3 SNF vital signs monitors may fail to power up or have unresponsive button controls, potentially preventing patient vital sign monitoring. The company is recalling 618 units distributed in Wisconsin.

    Product
    AVSM3 SNF, Vital Signs Monitor The monitor is intended to be used to monitor noninvasive blood pressure (NIBP), functional arterial oxygen saturation (SpO2), pulse rate (PR), temperature (Temp) for adult, pediatric and neonatal patients in all areas of a hospital and…
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2097-2021·2021-07-28

    Medical Device Recall: Illuminated Laser Probe with Incorrect Expiration Date

    Dutch Ophthalmic USA is recalling 13 boxes of Illuminated Stepped Laser Probe Curved devices due to a barcode labeling error. The barcode displays 2025-10-27 when the correct expiration date is 2024-10-01.

    Product
    Illuminated Stepped Laser Probe Curved-for use in combination with 23G vitrectomy system Product Code: 7510.ALC
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2126-2021·2021-07-28

    Exactech GXL Hip Implant Liners Recalled for Premature Wear Risk

    Exactech is recalling its Connexion GXL acetabular polyethylene liners used in hip implants because they may experience edge-loading and premature wear in certain patient configurations and surgical positioning. A total of 89,050 units were distributed nationwide.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-28-XX, Novation GXL Liner, +5mm Lateralized, 28mm; and Novation GXL Liner, +5mm Lateralized, G0, 22mm ID. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2132-2021·2021-07-28

    Exactech Hip Implant Liners Recalled Due to Premature Wear Risk

    Exactech is recalling Connexion GXL hip implant liners due to edge-loading and premature wear risk in certain patient configurations. The recall affects 89,050 units distributed nationwide.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-32-XX, Novation GXL Liner, 10 Deg Face, 32mm. Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-36-XX, Acumatch…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2124-2021·2021-07-28

    Hip implant liners recalled for edge-loading and premature wear

    Exactech Connexion GXL hip implant liners are recalled due to risk of edge-loading and premature wear in certain configurations. About 89,050 units distributed nationwide are affected.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 134-28-XX, Acumatch GXL, Ext Cov Liner, 28mm. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2123-2021·2021-07-28

    Exactech Connexion GXL Hip Implant Liners Recalled for Premature Wear Risk

    Exactech recalls approximately 89,050 Connexion GXL acetabular polyethylene hip implant liners due to risk of edge-loading and premature wear in certain implant configurations.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-40-XX, Novation GXL Liner, Lipped Ant, 40mm; Novation GXL Liner, Lipped, 40mm. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2128-2021·2021-07-28

    Hip Implant Liners Recalled Due to Premature Wear Risk

    Exactech Connexion GXL acetabular polyethylene liners used in hip replacement systems are recalled due to risk of edge-loading and premature wear in certain patients. The 89,050 affected units were distributed nationwide.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-36-XX, Novation GXL Liner, +5mm Lateralized, 36mm. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2086-2021·2021-07-28

    Medical dialysis concentrates recalled for freezing temperature exposure

    Fresenius Medical Care recalls Citrasate and NaturaLyte hemodialysis acid concentrates due to exposure to freezing temperatures during refrigerated truck transport. The affected products were distributed exclusively to Texas.

    Product
    Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are each intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine according to a physician s prescription. Both concentrates are formulated for…
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2130-2021·2021-07-28

    Exactech Connexion GXL hip implant liners risk edge-loading and wear

    Exactech Connexion GXL acetabular polyethylene liners used in hip replacements may experience edge-loading and premature wear in certain patients due to implant configuration or surgical positioning. Approximately 89,050 units were distributed nationwide.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-22-XX, Novation GXL 10 deg Liner, G00, 22mm ID. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2113-2021·2021-07-28

    Exactech Connexion GXL hip implant liners recalled due to premature wear risk

    Exactech is recalling 89,050 Connexion GXL hip implant liners distributed nationwide due to a risk of edge-loading and premature wear in certain patient implant configurations.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 104-32-XX, MCS +5GXL Liner 5/15 DEG and MCS GXL Liner 5/15 DEG. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2099-2021·2021-07-28

    Medical Device Calibration Standard Assigned Incorrectly in NEFA CAL Kit

    Randox Laboratories recalls NEFA CAL IVD calibration standard kits (Batch 544642) due to incorrectly assigned calibration values that could cause measurement errors up to 8% in quality control and patient test results.

    Product
    NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2115-2021·2021-07-28

    Exactech Hip Implant Liners Recalled Due to Premature Wear Risk

    Exactech is recalling 89,050 Connexion GXL acetabular polyethylene hip implant liners due to risk of premature prosthesis wear in certain patient configurations and surgical positioning scenarios.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-22-XX, Novation GXL Neutral Liner, G00, 22mm ID. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2107-2021·2021-07-28

    Suture Anchor Products Recalled Due to Improper Sterilization

    Arthrex is recalling 3,985 BioComposite SwiveLock C suture anchors due to improper sterilization from seal defects. The affected lots (12758314 and 13022357) were distributed nationwide.

    Product
    Suture Anchor, BioComposite SwiveLock C, 5.5 mm x 19.1 mm, Closed Eyelet, REF AR-2323BCC
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2119-2021·2021-07-28

    Exactech Connexion GXL Hip Implant Liners Risk Premature Wear

    Exactech hip implant liners may experience premature wear and edge-loading in certain patient configurations, according to this Class II recall of 89,050 units distributed nationwide.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-40-XX, Novation GXL Liner, Neutral, 40mm. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2082-2021·2021-07-28

    Hemodialysis concentrate recalled for freezing exposure during transport

    Fresenius Medical Care recalls Granuflo and Naturalyte hemodialysis concentrate affected by freezing temperatures during refrigerated truck transport. Freezing may compromise product sterility and safety.

    Product
    (1) 08-4078-BB BIBAG 650 GRAMS/BAG 16 BAGS/CASE (2) 08-4080-BB BIBAG 900 grams/bag 12 bags/case For patients with acute and chronic end stage renal disease going under PD in healthcare or home setting GRANUFLO & NATURALYTE: Concentrate is formulated to be used in…
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2129-2021·2021-07-28

    Exactech Hip Implant Liners Recalled for Premature Wear Risk

    Exactech Connexion GXL hip implant liners are recalled due to risk of edge-loading and premature wear in certain patient configurations. No injuries have been reported.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-40-XX, Novation GXL Liner, +5mm Lateralized, 40mm. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2106-2021·2021-07-28

    Cell Marque MSH6 Antibody Recalled for Distribution Without FDA Clearance

    Cell Marque Corporation is recalling Cell Marque Antibodies, MSH6 (44), Mouse Monoclonal Antibody because the product was distributed as an in-vitro diagnostic device without FDA clearance.

    Product
    Cell Marque Antibodies, MSH6 (44), Mouse Monoclonal Antibody, REF: 287M-10, IVD, CE, GTIN: 00841683104513
    Category
    Medical Device
    Distribution
    18 states
  • HighFDA (Devices)·Z-2118-2021·2021-07-28

    Hip implant polyethylene liners recalled due to premature wear risk

    Exactech Connexion GXL acetabular liners used in hip implants are recalled due to risk of edge-loading and premature wear in patients with specific implant configurations and surgical positioning.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-36-XX, Acumatch GXL 0 Degree Liner, 36mm; and Novation GXL Liner, Neutral, 36mm. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2090-2021·2021-07-28

    Fresenius Sterile Stay Safe Cap Recalled for Freezing Temperature Exposure

    Fresenius recalls 39 cases of Sterile Stay Safe Caps (lot numbers B1ZH191, B1ZI092) distributed to Texas due to exposure to freezing temperatures during refrigerated truck malfunction, which may have compromised product sterility.

    Product
    050-95012 Sterile Stay Safe Cap The Indications for Use/Intended Use for the Fresenius Sterile Stay Safe¿ Cap is equivalent to that for the Fresenius Stay Safe Cap and is as follows: Indications for Use Fresenius Sterile Stay Safe¿ Cap is intended to be used for closure of the…
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2093-2021·2021-07-28

    Laboratory Information System SoftGenomics Recalled for Displaying Incorrect Results

    SoftGenomics version 4.1.15.6, a laboratory information system used in clinical laboratories, has been recalled due to reports that the software displays incorrect results and PDF reports do not match the hospital information system.

    Product
    SoftGenomics version 4.1.15.6 SoftGenomics is a laboratory workflow and information management system to be used in a clinical molecular or cytogenetics laboratory or medical research laboratory setting by knowledgeable and trained users. SoftGenomics is used to record and…
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-2111-2021·2021-07-28

    Ultrasound Imaging Software Recalled for Inaccurate Measurement Estimates

    Caption Health recalled ultrasound imaging software version 1.2.0 due to a bug that includes unintended video frames in its analysis, potentially resulting in incorrect automated ejection fraction and image quality score estimates. The recall affects approximately 15 systems distributed across five U.S. states.

    Product
    Version: 1.2.0 (1.2.0+85ela410) Caption Health, Caption AI, Ultrasound Imaging System Application Software (GMDN: 40873), UDI: (01)00860003586303(10)1.2.0)
    Category
    Medical Device
    Distribution
    5 states
  • HighFDA (Devices)·Z-2131-2021·2021-07-28

    Exactech Hip Implant Liners Recalled Due to Premature Wear Risk

    Exactech is recalling certain acetabular polyethylene hip implant liners due to risk of edge-loading and premature wear in specific implant configurations and surgical positions.

    Product
    Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-28-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 28mm; Novation GXL 10 Deg Liner, G0, 28mm; and Novation GXL Liner, 10 Deg Face, 28mm. Orthopedic hip implant component.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2089-2021·2021-07-28

    MTS Set dialysis accessory recalled for freezing temperature exposure during transport

    Fresenius is recalling MTS Sets distributed to Texas due to exposure to freezing temperatures from refrigerated truck malfunctions. This exposure may compromise the sterility of the device.

    Product
    The MTS Set is a single-use, sterile (EO), non-pyrogenic, optional accessory that provides additional connections/disconnections when used with a stay"safe compatible cycler set during acute and chronic PD treatment. The MTS Set utilizes the same PIN engagement technology used…
    Category
    Medical Device
    Distribution
    1 state