Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

8526–8550 of 40095

  • HighFDA (Devices)·Z-0587-2024·2023-12-27

    Exactech Vantage Ankle System Recalled Due to Vacuum Loss

    Exactech is recalling two catalog numbers of Vantage Ankle System devices due to loss of vacuum in the inner-most vacuum bag. Three complaints have been received relating to four affected devices.

    Product
    Exactech Vantage Ankle System, Catalog Numbers: a) 350-21-04, b) 350-22-03
    Category
    Medical Device
    Distribution
    16 states
  • HighFDA (Devices)·Z-0589-2024·2023-12-27

    Acrobat-i Vacuum Positioner System Recalled for Positioner Arm Locking Failure

    Maquet Cardiovascular is recalling the Acrobat-i Vacuum Positioner System because the positioner arm may fail to tighten or lock during normal use, which could affect surgical positioning.

    Product
    Acrobat-i Vacuum Positioner System, Model Number XP-5000Z
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0578-2024·2023-12-27

    Hip Implant Liner Recalled Due to Loss of Vacuum in Inner Bag

    Exactech has recalled certain Alteon Hip XLE Liner hip implant components due to vacuum loss in the inner vacuum bag. The company has received 3 complaints involving 4 devices related to this issue.

    Product
    Exactech Alteon Hip XLE Liner, Catalog Numbers: a) 01-030-40-0636, b) 01-030-40-0640, c) 01-030-42-0536
    Category
    Medical Device
    Distribution
    16 states
  • HighFDA (Devices)·Z-0565-2024·2023-12-27

    CardioQuip Modular Cooler-Heater Device Recalled for Potential Bacterial Contamination

    CardioQuip is recalling the Modular Cooler-Heater (Model MCH-1000i) due to potential bacterial contamination that could cause patient infection. Four units were distributed in California, Texas, Pennsylvania, Washington, Ohio, and Nebraska.

    Product
    CardioQuip Modular Cooler-Heater, Model: MCH-1000(i), Power: 115V-60Hz, 16A
    Category
    Medical Device
    Distribution
    6 states
  • HighFDA (Devices)·Z-0606-2024·2023-12-27

    RAPIDPoint 500 Measurement Cartridges Recalled for Sodium Sensor Bias

    Siemens is recalling RAPIDPoint 500 Systems Measurement Cartridges because the sodium sensor can develop a negative measurement bias that could delay diagnosis of serious electrolyte disorders or cause treatment errors.

    Product
    RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 100 -Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10844813
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0580-2024·2023-12-27

    Exactech Equinoxe Shoulder System recalled for loss of vacuum

    Exactech is recalling specific models of its Equinoxe Shoulder System implants due to loss of vacuum in the inner-most vacuum bag. Four devices have been affected according to three complaints received.

    Product
    Exactech Equinoxe Shoulder System, Catalog Numbers: a) 314-23-02, b) 314-23-03, c) 314-23-13, d) 314-23-14, e) 314-23-15, f) 314-24-22, g) 314-24-23, h) 314-24-24, i) 314-24-32, j) 314-24-33, k) 314-24-34, l) 314-24-35
    Category
    Medical Device
    Distribution
    16 states
  • HighFDA (Devices)·Z-0571-2024·2023-12-27

    Injection Needle Recall Due to Cleaning Process Concerns

    Karl Storz Endoscopy is recalling injection needles used in general and visceral surgery because the manual cleaning process cannot be assured effective, which may expose patients to a higher risk of infection.

    Product
    Injection Needle, REF: 25207
    Category
    Medical Device
    Distribution
    38 states
  • HighFDA (Devices)·Z-0585-2024·2023-12-27

    Exactech Truliant Knee System Recalled Due to Vacuum Bag Failure

    Exactech recalls certain Truliant Knee System units due to loss of vacuum in the inner vacuum bag. The affected units were distributed across multiple U.S. states and international locations.

    Product
    Exactech Truliant Knee System, Catalog Numbers: a) 02-012-35-3509, b) 02-022-35-1509, c) 02-022-35-1513, d) 02-022-35-3010, e) 02-022-35-3509, f) 02-022-35-3510, g) 02-022-35-3511, h) 02-022-35-4511, i) 02-022-44-3512, j) 02-022-47-2509, k) 02-022-51-4519
    Category
    Medical Device
    Distribution
    16 states
  • HighFDA (Devices)·Z-0572-2024·2023-12-27

    Injection needles and cannulas recalled for cleaning process efficacy issues

    Karl Storz Endoscopy is recalling injection needles and cannulas because the manual cleaning process may not effectively remove contaminants, potentially exposing patients to infection risk.

    Product
    Injection Needle, 3 Fr., REF: 27030N; Injection Cannula, 6 Fr., REF: 27182A
    Category
    Medical Device
    Distribution
    38 states
  • HighFDA (Devices)·Z-0576-2024·2023-12-27

    Philips Inclusive CT Systems: Software Errors in Image Display and Interpretation

    Philips North America is recalling 330 Philips Inclusive CT imaging systems due to software issues causing incorrect image display, errors in image interpretation, and patient exposure to incorrect images and system functions during clinical use.

    Product
    The device is a whole-body CT x-ray system. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body taken at different angles and planes. The system also includes signal analysis and display equipment, patient and equipment suppor
    Category
    Medical Device
    Distribution
    25 states
  • ModerateFDA (Devices)·Z-0603-2024·2023-12-27

    ZEUS ELISA Parvovirus B19 Test Kits Recalled for Unapproved Modification

    Zeus Scientific recalls ZEUS ELISA Parvovirus B19 IgM test kits due to an unapproved modification to the test conjugate that was implemented without FDA review and approval.

    Product
    ZEUS ELISA Parvovirus B19 IgM Test System, Product Number SM9Z7701G
    Category
    Medical Device
    Distribution
    7 states
  • ModerateFDA (Devices)·Z-0564-2024·2023-12-27

    Vero Biotech GENOSYL DS cassette recalled for manufacturing flow defect

    Vero Biotech is recalling GENOSYL DS Generation 3 cassettes that failed to meet manufacturing flow specifications. The recall affects 2 units distributed in Arizona and Florida.

    Product
    VERO BIOTECH GENOSYL DS (Delivery System) Generation 3 Cassette, Model Number 602722-01; generates and delivers nitric oxide (NO) for inhalation at the point of use
    Category
    Medical Device
    Distribution
    2 states
  • ModerateFDA (Devices)·Z-0595-2024·2023-12-27

    CARDIOHELP-i Heart-Lung Support System not properly tested for leakage current

    The CARDIOHELP-i Heart-Lung Support System was not properly tested to measure leakage current, an electrical safety concern. The FDA is recalling 1,594 units distributed worldwide.

    Product
    CARDIOHELP-i Heart Lung Support System: Blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support for periods appropriate to cardiopulmonary bypass (up to six hours). Product REF Numb
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0602-2024·2023-12-27

    ZEUS ELISA Parvovirus B19 Test Kits Recalled for Unapproved Modification

    Zeus Scientific is recalling 296 ZEUS ELISA Parvovirus B19 IgG Test System kits distributed in seven US states because an unapproved modification was made to the test reagent without FDA approval.

    Product
    ZEUS ELISA Parvovirus B19 IgG Test System, Product Number SM9Z7701M
    Category
    Medical Device
    Distribution
    7 states
  • ModerateFDA (Devices)·Z-0560-2024·2023-12-27

    Roche Cobas Interface Module Firmware May Cause Delayed Sample Processing

    Inpeco is recalling firmware versions for the Roche Cobas 8000 and PRO Interface Module that may cause delayed sample tube processing in laboratory automation systems.

    Product
    Roche Cobas 8000 and Cobas PRO Interface Module (CO8 IM)- IVD FlexLab Automation System -a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. Hardware versions: FLX-247-0
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0568-2024·2023-12-27

    ACMI Roller Bar Electrodes Recall Due to Incorrect Pouch Labeling

    Olympus Corporation of the Americas is recalling ACMI Roller Bar Electrodes due to incorrect pouch labels. Pouches are mislabeled as Right Angle Cutting Loop Electrodes but contain the correct Roller Bar product.

    Product
    ACMI Roller Bar Electrodes bearing incorrect pouch label of ACMI USA Elite System and USA Series Right Angle Cutting Loop Electrode Stabilized, 0.012 Wire 24 Fr (8mm), Model: REF MLE 24-012. Product packaged in boxes of 6 pouches. Box of 6 has correct box label of ACMI USA Elite
    Category
    Medical Device
    Distribution
    13 states
  • ModerateFDA (Devices)·Z-0586-2024·2023-12-27

    Exactech knee replacement inserts recalled for vacuum seal loss

    Exactech is recalling Truliant and Activit-E knee replacement inserts for loss of vacuum in storage packaging. The defect was identified in 3 complaints involving 4 devices, with no reported injuries.

    Product
    Exactech Truliant, Activit-E Knee Insert, Catalog Numbers: a) 02-023-02-0029, b) 02-023-02-0032, c) 02-023-02-0035, d) 02-023-02-0038, e) 02-024-35-1512, f) 02-024-35-1515, g) 02-024-35-1519, h) 02-024-35-2010, i) 02-024-35-2013, j) 02-024-35-2015, k) 02-02
    Category
    Medical Device
    Distribution
    16 states
  • SevereFDA (Devices)·Z-0505-2024·2023-12-20

    Philips Panorama MR System Structural Failure Risk from Helium Gas Pressure

    Philips Panorama 1.0T MR Systems (Models 781250, 781350) may experience structural failure during helium quench events, releasing pressurized gas and debris that could cause helium exposure, asphyxia, and injury to patients and operators.

    Product
    Panorama 1.0T HFO, Magnetic Resonance (MR) System; Model Numbers 781250 and 781350. Philips Magnetic Resonance (MR) systems are Medical Electrical Systems for use as a diagnostic device.
    Category
    Medical Device
    Distribution
    34 states
  • SevereFDA (Devices)·Z-0548-2024·2023-12-20

    CereLink ICP Extension Cable Recalled for Out of Range Readings

    Integra LifeSciences is recalling CereLink ICP Extension Cables due to a design defect causing out of range readings. The Class I recall affects 4,941 units distributed nationwide and internationally.

    Product
    CereLink ICP Extension Cable (Extension Cable), Model No. 826845 Used with Codman CereLink Intracranial Pressure (ICP) Monitor, Model No. 826820/826820P. Cable is also provided with monitor as part of the same SKU.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0528-2024·2023-12-20

    Sterile PVP Medical Solution Recalled for Non-Sterile Contents

    Medline Industries is recalling 700 units of Sterile PVP solution because the labeling claims sterility but the contents are non-sterile. The affected products were distributed nationwide in eight states.

    Product
    STERILE PVP, REF DYNDA1649
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0555-2024·2023-12-20

    CARDIOHELP-i Heart-Lung Support System: Incorrectly Documented Factory Settings

    The CARDIOHELP-i Heart-Lung Support System is recalled because factory settings were incorrectly described in the manual. Healthcare providers should verify device settings with the manufacturer.

    Product
    CARDIOHELP-i Heart Lung Support System- A blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support for periods appropriate to cardiopulmonary bypass Part Number: 701072780, (CARDIO
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0533-2024·2023-12-20

    FDA Recalls Medline Leg Bags for Undeclared Latex

    Medline leg bags contain undeclared latex, creating a risk for users with latex allergies. The company is recalling 28,942 devices distributed in the US and Panama.

    Product
    Medline Leg bag, REF DYND12574
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0524-2024·2023-12-20

    Medical Device Recall: PHYSICA HPS Tibial Liner Incorrect Labeling

    Limacorporate is recalling 33 PHYSICA HPS Tibial Liner units distributed in the U.S. due to incorrect product labeling. No adverse events have been reported.

    Product
    PHYSICA HPS Tibial Liner #6 H10 REF 6539.54.610
    Category
    Medical Device
    Distribution
    5 states
  • HighFDA (Devices)·Z-0529-2024·2023-12-20

    Medline PVP Solution Recalled Due to Mislabeled Sterility

    Medline PVP Solution (REF DYNDA1858) is being recalled because it is labeled as sterile but contains non-sterile contents. 1,950 units are affected.

    Product
    PVP SOLUTION SING STRL, REF DYNDA1858
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0557-2024·2023-12-20

    Olympus tracheal intubation fiberscope recalled due to suction flow specification issue

    Olympus tracheal intubation fiberscopes are being recalled because the manufacturer lacks adequate data confirming the suction flow rate meets device specifications. No injuries have been reported.

    Product
    TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-GP, Model Number: LF-GP
    Category
    Medical Device
    Distribution
    Distributed nationwide