The Recall Desk

Category

Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

38776–38800 of 40489

  • HighFDA (Devices)·Z-0979-2013·2013-04-03

    Cardinal Health Recalls Needle Counters Due to Packaging Defects

    Cardinal Health is recalling specific lots of sterile needle counters because small holes in the packaging may compromise sterility.

    Product
    Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Adhesive Area (Cat. No. 4FA40SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0998-2013·2013-04-03

    Advanced Sterilization Products Recalls STERRAD 200 System Carriage Units

    Advanced Sterilization Products is recalling STERRAD 200 System Carriages because they may become dislodged under excessive load.

    Product
    STERRAD 200 System Carriage, P/N 10208. Used as a tool to expedite loading and unloading processed items from a STERRAD 200 sterilizer.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0991-2013·2013-04-03

    Biomerieux VITEK 2 Compact Computer System Recall for Patient Data Error

    Biomerieux Inc. is recalling VITEK 2 Compact computer systems because a software issue could link test results to the wrong patient.

    Product
    VITEK 2 Compact computer system when equipped with the following Hewlett-Packard PC models: (1) dc7100 8-port, Cat. #W0441; (2) dc7100 8-port (refurbished), Cat. #W0441R; (3) dc7700 8-port, Cat. #W0447; (4) dc7700 8-port (refurbished), Cat. #W0447R; (5) dc7800, Cat.…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1017-2013·2013-04-03

    Medtronic Archer Super Stiff Guidewire Recalled for Coating Separation Risk

    Medtronic is recalling Archer Super Stiff Guidewires because the PTFE coating may separate from the wire during use.

    Product
    Medtronic Archer Super Stiff Guidewire 0.035 inch (0.89mm), Model ARCSJ200W. Intended to facilitate catheter placement and exchange during procedures in the aorta.
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-0990-2013·2013-04-03

    Biomerieux VITEK 2 XL Computer System Recalled for Patient Data Linking Errors

    Biomerieux Inc. is recalling VITEK 2 XL computer systems equipped with specific Hewlett-Packard PCs due to a potential risk of linking test results to the wrong patient.

    Product
    VITEK 2 XL computer system when equipped with the following Hewlett-Packard PC models: (1) rp5700, Cat. #W0452; (2) rp5700 (refurbished), Cat. #W0452R; (3) rp5800 XPE, Cat. 413642; (4) rp5800 WES7, Cat. #413862; (5) dc7100 8-port, Cat. #W0441; (6) dc7100 8-port…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0996-2013·2013-04-03

    Smiths Medical Medfusion Syringe Pumps Recalled for Seal Migration Risk

    Smiths Medical is recalling six Medfusion Model 3500 syringe pumps in New Hampshire because a seal defect could cause the plunger rod seal to migrate during specific infusion conditions.

    Product
    Medfusion Model 3500 Syringe Pump, Model #3500-415 and 3500-500. Intended for use in critical care areas for the administration of fluids requiring precisely controlled infusion rates.
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-1019-2013·2013-04-03

    Medtronic Archer Super Stiff Guidewires Recalled for Potential Kinking and Breaking

    Medtronic is recalling specific Archer Super Stiff Guidewires due to a manufacturing defect that may cause the device to kink or break during use.

    Product
    Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W. Sterile using ethylene oxide. Product Usage: The Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewires are intended to facilitate catheter placement and exchange…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0995-2013·2013-04-03

    Kimberly-Clark Recalls KIMGUARD Container Filters Due to Sterility Concerns

    Kimberly-Clark is recalling KIMGUARD Container Filters because thin areas in some units may compromise the sterile barrier.

    Product
    KIMGUARD Container Filters, 7.5in (19cm), Round and 9in. X 9in. (23cm x 23cm), Square. Intended to be used to enclose another medical device that is to be sterilized by a health care provider.
    Category
    Medical Device
    Distribution
    36 states
  • HighFDA (Devices)·Z-1015-2013·2013-04-03

    CareFusion Recalls Pyxis Anesthesia System Model 2000 Due to Drawer Inaccessibility

    CareFusion is recalling 2,637 Pyxis Anesthesia System Model 2000 units because system downtime or lock-ups can make medication drawers inaccessible.

    Product
    Pyxis Anesthesia System Model 2000 (PAS 2000) Product Usage: The PAS 2000 is an anesthesia medication dispensing device that provides medications and inventory management for the anesthesiologist at the point of care within the operating room.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0989-2013·2013-04-03

    Biomerieux VITEK 2 60 Computer System Recalled for Data Linking Error

    Biomerieux Inc is recalling VITEK 2 60 computer systems due to a potential error that could link test results to the wrong patient.

    Product
    VITEK 2 60 computer system when equipped with the following Hewlett-Packard PC models: (1) rp5700, Cat. #W0452; (2) rp5700 (refurbished), Cat. #W0452R; (3) rp5800 XPE, Cat. 413642; (4) rp5800 WES7, Cat. #413862; (5) dc7100 8-port, Cat. #W0441; (6) dc7100 8-port…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1002-2013·2013-04-03

    Vital Images Recalls Vitrea Software Due to Measurement Errors

    Vital Images is recalling Vitrea software versions prior to 6.3.3 due to a potential error in measurements generated when images are rotated during reconstruction.

    Product
    Vitrea Enterprise Suite, Vitrea, VitreaAdvanced, VitreaCore (VitalConnect) and Vitrea fX prior to version 6.3.3, except 6.1.6, 6.2.3 and 6.3 The software is a medical diagnostic system that allows the processing, review, analysis, communication, and media interchange of…
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-1021-2013·2013-04-03

    Microbiologics Recalls KWIK-QC Slides Due to Improper Fixation

    Microbiologics Inc is voluntarily recalling KWIK-QC Acid Fast and Mycobacterium slides because control organisms are not properly fixated, posing a contamination risk.

    Product
    Microbiologics KWIK-QC SLIDES ( KWIK-QC Acid Fast Stain Slides and KWIK-QC Mycobacterium Slides). Slide preparations serving as quality control challenges to demonstrate satisfactory performance of staining reagents.
    Category
    Medical Device
    Distribution
    7 states
  • HighFDA (Devices)·Z-0997-2013·2013-04-03

    Siemens Artis Angiographic X-Ray Systems Recalled for Image Registration Issues

    Siemens is recalling 81 Artis angiographic x-ray systems because a software issue may cause imprecise image registration during specific procedures.

    Product
    Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0977-2013·2013-04-03

    Mindray BS-200 Chemistry Analyzer Recalled for Inaccurate Clinical Results

    Mindray is recalling three BS-200 Chemistry Analyzers because clinical chemistry results may be inaccurate. The units were distributed in North Carolina and Texas.

    Product
    BS-200 Chemistry Analyzer Designed for clinical laboratory use.
    Category
    Medical Device
    Distribution
    2 states
  • ModerateFDA (Devices)·Z-1013-2013·2013-04-03

    bioMerieux VITEK 2 Streptococcus Card Recalled for Incorrect Package Insert

    Biomerieux Inc is recalling VITEK 2 AST-ST01 Streptococcus Susceptibility Cards because the package insert did not list the correct limitations of use and performance requirements.

    Product
    bioMerieux VITEK 2 AST-ST01 Streptococcus Susceptibility Card, REF 410 028. For use with VITEK 2 systems to determine the susceptibility of Streptococcus to antimicrobial agents.
    Category
    Medical Device
    Distribution
    1 state
  • ModerateFDA (Devices)·Z-0980-2013·2013-04-03

    Cardinal Health Recalls Sterile Needle Counters Due to Packaging Defects

    Cardinal Health is recalling sterile needle counters with blade disarmers because small holes in the packaging may compromise sterility.

    Product
    Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 4FS30SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Removable Numbered Foam Block…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1016-2013·2013-04-03

    Synthes Dens Instrument Set Recalled Due to Lack of Indicated Screws

    Synthes USA HQ, Inc. is recalling 148 Dens Instrument Sets because there are no screws specifically indicated for the procedure.

    Product
    Synthes Dens Instrument Set (Dens Graphic Case), Part Number 687.030 (also known as the Anterior Column Retractor Set) The Synthes Dens instrument Set has been designed to facilitate internal fixation of the odontoid process for the reduction of Type II and shallow Type III…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0976-2013·2013-04-03

    Becton Dickinson Recalls BD MAX PCR Cartridges Due to Sealing Defects

    Becton Dickinson & Co. is recalling BD MAX PCR Cartridges because they may not seal properly, potentially causing false or invalid test results.

    Product
    BD MAX PCR Cartridges Catalog #437519, box of 24 cartridges labeled in part:***BD MAX PCR Cartridges, GeneOhm Sciences Canada, Inc. 2555 boul du Parc-Technologique, Quebec, Canada G1P 4S5***BD Max Group B Streptococcus (GBS) Assay is an in vitro diagnostic (IVD) test for…
    Category
    Medical Device
    Distribution
    11 states
  • ModerateFDA (Devices)·Z-0994-2013·2013-04-03

    Toshiba Infinix-i X-Ray System Recalled for Software Image Error

    Toshiba is recalling 95 Infinix-i X-Ray systems because a software error may display incorrect images or cause image loss during fluoroscopy.

    Product
    Toshiba Infinix-i X-Ray System. A radiographic and fluoroscopic x-ray system for use in diagnostic and angiographic procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1023-2013·2013-04-03

    Biosense Webster Recalls 20-Pole Eco Cable Due to Calibration Error

    Biosense Webster is recalling 54 units of the 20-Pole Eco Cable because incorrect calibration prevents the connected catheter from being recognized by the system.

    Product
    20-Pole Eco Cable, Catalog #EM-5050-060 Product Usage: The 20-Pole Eco Cable is intended to be used as an interface cable between the LASSO NAV Eco Catheter and CARTO 3 System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0975-2013·2013-04-03

    Orthohelix Intraosseous Fixation System Drill Bits Recalled for Manufacturing Defect

    Orthohelix Surgical Designs Inc is recalling specific lots of Intraosseous Fixation System drill bits because a burr prevents them from fitting over the guide wire.

    Product
    Intraosseous Fixation System (IFS) 1.7 mm, 2.0 mm & 2.5 mm Cannulated Drill Bit. Intended to stabilize and aid in fixation of fractures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1022-2013·2013-04-03

    Genetic Testing Institute Recalls PF4 Enhanced Assay Kits Due to Invalid Results

    Genetic Testing Institute is recalling 386 lots of LIFECODES PF4 Enhanced assay kits because inconsistent values can lead to invalid test results.

    Product
    Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45 GTI PF4 Enhanced is designed as a solid phase enzyme-linked immunosorbent assay (ELISA).
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-227-2013·2013-04-03

    CareFusion Blood Culture Prep Kits Recalled for Misleading Sterility Labeling

    CareFusion is recalling Blood Culture Prep Kit II units because the product is not sterile, making the labeling misleading regarding sterility claims.

    Product
    Blood Culture Prep Kit II containing 1 FREPP (70% isopropyl alcohol), Net 1.1 mL and 1 SEPP (iodine tincture 2% USP), Net 0.67 mL per kit, packaged in 20-count kits per carton, Cat. No 260307, NDC 54365-200-99, CareFusion, Leawood, KS 66211.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-228-2013·2013-04-03

    CareFusion Recalls Donor Prep Kits Due to Misleading Sterility Labeling

    CareFusion 213, LLC is recalling 46,500 Donor Prep Kits because the labeling is misleading regarding sterility claims. The product is not sterile.

    Product
    Donor Prep Kit containing 1 FREPP (70% isopropyl alcohol), Net 1.1 mL and 1 SEPP (iodine tincture 2% USP), Net 0.67 mL per kit, packaged in 20-count kits per carton, Cat. No 260311, NDC 54365-201-99, CareFusion, Leawood, KS 66211.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0986-2013·2013-04-03

    Cardinal Health Recalls Needle Counters Due to Packaging Sterility Concerns

    Cardinal Health is recalling specific lots of sterile needle counters because small holes in the packaging may compromise sterility.

    Product
    Needle Counters with Blade Disarmer; 12.2 x 10.5 x 3.5 cm; 60 Count/100 Capacity; Sterile; Removable Numbered Foam Block; Double Numbered Magnets (Cat. No. 4FM100SSA) Product Usage: Disposal system for counting of used surgical blades and needles.
    Category
    Medical Device
    Distribution
    Distributed nationwide

Looking for a different category? Browse all recalls.