Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

38026–38050 of 40489

  • ModerateFDA (Devices)·Z-1784-2013·2013-07-31

    Instrumentation Laboratory Recalls HemosIL RecombiPlasTin 2G Coagulation Reagent

    Instrumentation Laboratory is recalling HemosIL RecombiPlasTin 2G reagent kits due to out-of-specification quality issues. The product is used for evaluating the extrinsic coagulation pathway.

    Product
    HemosIL RecombiPlasTin 2G, Part Number 0020002950 (8 ml size). The product is used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Anticoagulant Therapy.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1803-2013·2013-07-31

    Becton Dickinson Recalls Brucella Laked Blood Agar Due to Reduced Vancomycin

    Becton Dickinson is recalling specific lots of Brucella Laked Blood Agar because the media may contain reduced levels of Vancomycin.

    Product
    BBL Brucella Laked Blood Agar with Kanamycin and Vancomycin (KV), Catalog # 297840, packaged in cartons of 20 plates, labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 USA 800-638-8663 www.bd.com/ds*** Brucella Laked Blood with KV is used for…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1796-2013·2013-07-31

    Haemonetics Recalls TEG 5000 Analyzers Due to Manual Errors

    Haemonetics is recalling 1,590 TEG 5000 Thromboelastograph Coagulation Analyzers because the user manual incorrectly included Research Use Only parameters.

    Product
    TEG 5000 Thromboelastograph Coagulation Analyzer Product Usage: Multipurpose System for In Vitro Coagulation Studies
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1782-2013·2013-07-31

    Siemens ARTIS zee and zeego X-ray Systems Recalled for Software Issue

    Siemens is recalling 29 ARTIS zee and zeego x-ray systems due to a software issue affecting rotational 3D-acquisition programs with real-time subtraction.

    Product
    ARTIS zee and zeego x-ray, angiographic system with Software Artis VC20x/VC21A that have Patch 7 and a 3D and DSA license installed. Angiographic x-ray system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1771-2013·2013-07-31

    Spacelabs ARKON Anesthesia Workstation Recalled for Software Defect

    Del Mar Reynolds Medical is recalling nine Spacelabs ARKON Anesthesia Workstations due to a software defect that may cause false communication error alerts.

    Product
    Spacelabs ARKON Anesthesia Workstation. Spacelabs ARKON Anesthesia Workstation may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the…
    Category
    Medical Device
    Distribution
    1 state
  • ModerateFDA (Devices)·Z-1806-2013·2013-07-31

    Aesculap ProSpace PEEK Spinal Implant Recalled for Incorrect Etching

    Aesculap is recalling nine ProSpace PEEK spinal implants because they are incorrectly etched with an 8-degree lordosis instead of the labeled 5-degree lordosis.

    Product
    ProSpace Peek Implant 5 degree x 8.5 x 22 mm The device is intended for vertebral body replacement or intervertebral body fusion to aid in the surgical correction or stabilization of the spine. The Aesculap PEEK Spinal Implant System is indicated for use in the thoracolumbar…
    Category
    Medical Device
    Distribution
    2 states
  • LowFDA (Devices)·Z-1778-2013·2013-07-31

    National Biological Dermalight 80 UVB Phototherapy Units Recalled for Incorrect Manual

    National Biological Corp is recalling Dermalight 80 UVB phototherapy units because they may have been packaged with the manual for a different UVA-083 device.

    Product
    Dermalight 80 Model Number UVB-080, one unit per box. The Dermalight 80 is a Phototherapy light unit that provides ultraviolet radiation therapy for diagnosed skin disorders such as Psoriasis and Vitiligo.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • LowFDA (Devices)·Z-1810-2013·2013-07-31

    Medtronic Preimplantation Test Kit Recalled for Incorrect Expiration Dates

    Medtronic Neurosurgery is recalling Preimplantation Test Kits because the outer carton expiration dates are later than the dates on the individual sterile components.

    Product
    Medtronic Preimplantation Test Kit, Catalog No. 21047. The kit is indicated for use with Medtronic CSF-Flow Control Valves and Delta Valves, when a simple preimplantation test is desired to verify that the valve conforms to labeled product specifications.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • LowFDA (Devices)·Z-1779-2013·2013-07-31

    National Biological Dermalight 80 UVB Phototherapy Units Recalled for Incorrect Manual

    National Biological Corp is recalling six Dermalight 80 UVB phototherapy units because they may have been packaged with the manual for a different UVA-080 device.

    Product
    Dermalight 80 Model Number UVB-083, one unit per box. The Dermalight 80 is a Phototherapy light unit that provides ultraviolet radiation therapy for diagnosed skin disorders such as Psoriasis and Vitiligo.
    Category
    Medical Device
    Distribution
    4 states
  • CriticalFDA (Devices)·Z-1691-2013·2013-07-24

    Codman MedStream Infusion Pump Recalled for Miscalibrated Fill Level Sensor

    Codman & Shurtleff is recalling MedStream Programmable Infusion Pumps because a miscalibrated fill level sensor may affect dosing levels.

    Product
    MedStream Programmable Infusion Pump , Implantable for the intrathecal delivery of Baclofen or Morphine Product Code: 91-4201US (40ml pump) 91-4201 (40ml pump) Intended for the intrathecal delivery of Baclofen or Morphine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • CriticalFDA (Devices)·Z-1690-2013·2013-07-24

    Codman MedStream Infusion Pump Recalled for Miscalibrated Fill Level Sensor

    Codman & Shurtleff is recalling MedStream Programmable Infusion Pumps because a miscalibrated fill level sensor may affect dosing levels.

    Product
    MedStream Programmable Infusion Pump- Implantable for the intrathecal delivery of Baclofen or Morphine Product Code:91-4200 (20ml pump) US 91-4200. Intended for the intrathecal delivery of Baclofen or Morphine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1738-2013·2013-07-24

    St. Jude Medical Recalls Eon Spinal Cord Stimulator Charging Systems

    St. Jude Medical is recalling Eon and Eon Mini charging systems due to reports of skin burns and discomfort at the implant site during use.

    Product
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1753-2013·2013-07-24

    Liebel-Flarsheim Hydra Vision DR 60/80 System Recalled for High Dosage

    Liebel-Flarsheim is recalling 12 Hydra Vision DR 60/80 systems because a service call found the maximum fluoroscopy dose was three times the safety limit.

    Product
    Hydra Vision DR 60/80 System, Models 700539 and 700540, Liebel-Flarshiem Company. The System is used in radiologic imaging procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1739-2013·2013-07-24

    St. Jude Medical Recalls Eon Mini Spinal Cord Stimulator Chargers

    St. Jude Medical is recalling Eon Mini chargers due to reports of skin burns and discomfort at the implant site during charging.

    Product
    Eon Mini - 3721 (Charger). Product Usage: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1736-2013·2013-07-24

    Integra PANTA Arthrodesis Nail System Recalled for Screw Misalignment

    Integra LifeSciences is recalling PANTA Arthrodesis Nail Systems due to screw misalignment that may require revision surgery.

    Product
    Integra PANTA Arthrodesis Nail System. Intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1733-2013·2013-07-24

    Terumo Recalls CDI 500 Blood Parameter Monitoring System Due to Software Error

    Terumo is recalling 3,972 CDI 500 Blood Parameter Monitoring Systems because a software algorithm inaccuracy was discovered during internal testing.

    Product
    CDI Blood Parameter Monitoring System 500 with Arterial blood parameter module. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1746-2013·2013-07-24

    Staar Surgical Recalls Intraocular Lenses Due to Glass Particles

    Staar Surgical is recalling nanoFLEX and Afinity intraocular lenses because glass particles may be present in the storage vials.

    Product
    nanoFLEX CC4204A Intraocular Lens Correct aphakia in persons 60 years of age or older in whom a cataractous lens has been removed by cataract extraction. The IOL is to be implanted in the posterior chamber and in the capsular bag through a tearfree capsulorhexis.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1747-2013·2013-07-24

    Staar Surgical Recalls Intraocular Lenses Due to Glass Particles

    Staar Surgical is recalling specific Afinity and nanoFLEX intraocular lenses because glass particles may be present in the storage vials.

    Product
    Afinity CQ2015A Intraocular Lens The visual correction of aphakia in persons 60 years of age or older in whom a cataractous lens has been removed by cataract extraction.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1737-2013·2013-07-24

    SpineFrontier Recalls KRD1 PedFuse Rods Due to Fatigue Testing Failures

    SpineFrontier, Inc. is recalling KRD1 PedFuse Rods because one rod in each of two test constructs failed to meet fatigue requirements during dynamic compression testing.

    Product
    KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicle screw tulip and they are locked into place with a set screw or locking cap. The rod may be…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1740-2013·2013-07-24

    Stryker TPS Universal Driver Recalled for Unintended Activation Risk

    Stryker is recalling TPS Universal Drivers due to incorrect programming that may cause unintended activation, posing a risk of injury to patients or users.

    Product
    Stryker TPS Universal Driver REF 5100- 99 Rx , The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental orthopedic…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1729-2013·2013-07-24

    KooJoo Trading Company Recalls Unsterilized Safety Lancets

    KooJoo Trading Company is recalling specific lots of Odyssey safety lancets because they were not sterilized within the established validation dose range.

    Product
    Safety Lancets - Odyssey, Button Activated; Tri- Bevel Tip for Virtually Painless Sampling; Needle automatically retracts, No lancing device required. 26G - Single Use Only, Made In Korea Distributed by Hummingbird Medical, Savannah, GA 100 count Boxes; 40 boxes per…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1749-2013·2013-07-24

    Edwards Lifesciences Recalls Suction Wands Due to Plastic Particulate

    Edwards Lifesciences is recalling specific models of rigid suction wands because plastic particulate was found inside the sterile pouch and within the wand.

    Product
    Edwards Lifesciences Suction Wand, models S099 and S0998B Suction wands are intended for hand held use and pericardial sumps are intended for stationary use in removing excess fluid from the surgical field.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1741-2013·2013-07-24

    Stryker CORE Universal Driver Recalled for Unintended Activation Risk

    Stryker is recalling 11 CORE Universal Drivers because incorrect programming parameters may cause unintended activation, posing a potential injury risk to patients and users.

    Product
    CORE Universal Driver REF 5400- 99 Rx Only This drill may also be used with theTotal Performance System (TPS").The Stryker Total Performance System (TPS) is intended for usein the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1750-2013·2013-07-24

    Siemens AXIOM Artis zee/zeego Systems Recalled for Generator Power Supply Defect

    Siemens is recalling 23 AXIOM Artis zee/zeego angiographic x-ray systems because a tolerance problem in the generator power supply may cause the generator to fail.

    Product
    Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-ray system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1751-2013·2013-07-24

    Hospira Recalls Symbiq IV Pump Sets Due to Air-in-Line Alarm Risk

    Hospira is recalling Symbiq IV pump sets because a defective ball valve may fail to seal, causing air-in-line alarms during fluid administration.

    Product
    LifeShield; Sterile; Rx only; List No. 19666-28; SYMBIQ PUMP SET, CONVERTIBLE PIN, PIGGYBACK, 2 CLAVES PORTS AND OPTION LOK, NON-DEHP; NOMINAL LENGTH: 97 IN (249 CM); PRIMING VOLUME: 14 ML; For the administration of I.V. fluids; Made in Costa Rica; Hospira, Inc., Lake Forest, IL…
    Category
    Medical Device
    Distribution
    Distributed nationwide