The Recall Desk

Category

Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

34501–34525 of 40489

  • ModerateFDA (Devices)·Z-0465-2015·2014-12-03

    Remel Recalls Legionella Pneumo Groups 2-14 Latex Test Kits

    Remel Inc is recalling specific lots of the Oxoid Legionella Pneumo Groups 2-14 Latex Test because a reagent may produce false negative results.

    Product
    Oxoid Legionella Pneumo Groups 2-14 Latex Test, DR0802M, containing 50 tests per box. The box contains a variety of components, one of which is Oxoid L. pneumophila Group 2-14 Latex Reagent, Lot 1301912. The product is a latex agglutination test for the identification of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0424-2015·2014-12-03

    Zimmer Recalls Trabecular Metal Tibial Impactor Replacement Pads

    Zimmer, Inc. is recalling 4,961 Trabecular Metal Tibial Impactor Replacement Pads due to complaints of pad fracture. The recall provides clarifying instructions for assembly and disassembly.

    Product
    Trabecular Metal Tibial Impactor Replacement Pads. Product Usage: TM Tibial Impactor pads are used in conjunction with the TM Tibial Impactor to impact the tibial tray onto the tibia.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0475-2015·2014-12-03

    Dharma Cavity Varnish Recalled for Lack of FDA Clearance

    Dharma Research, Inc. is recalling Dharma Cavity Varnish with Fluoride because the product was marketed without a cleared 510(k).

    Product
    Dharma Cavity Varnish with Fluoride
    Category
    Medical Device
    Distribution
    5 states
  • ModerateFDA (Devices)·Z-0431-2015·2014-12-03

    Intuitive Surgical Recalls 8mm Micro Bipolar Forceps Due to Reprocessing Deviations

    Intuitive Surgical is recalling 509 units of 8mm Micro Bipolar Forceps because deviations in reprocessing steps can cause surface degradation and mechanical wear.

    Product
    8MM,MICRO BIPOLAR FORCEPS; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-0433-2015·2014-12-03

    Intuitive Surgical Recalls 8mm Maryland Bipolar Forceps Due to Reprocessing Deviations

    Intuitive Surgical is recalling 8mm Maryland Bipolar Forceps because deviations in reprocessing steps can cause surface degradation and accelerated mechanical wear.

    Product
    8MM,MARYLAND BIPOLAR FORCEPS; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-0437-2015·2014-12-03

    Intuitive Surgical Recalls 8mm Permanent Cautery Spatulas for Housing Degradation

    Intuitive Surgical is recalling 509 units of 8mm Permanent Cautery Spatulas because deviations in reprocessing steps can cause surface degradation of the housing and accelerate mechanical wear.

    Product
    8MM PERMANENT CAUTERY SPATULA,; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-0423-2015·2014-12-03

    Zimmer Recalls Trabecular Metal Tibial Impactor Pads Due to Fracture Complaints

    Zimmer, Inc. is recalling Trabecular Metal Tibial Impactor Pads to provide clarifying instructions on assembly and disassembly following complaints of pad fracture.

    Product
    Trabecular Metal Tibial Impactor. Product Usage: TM Tibial Impactor pads are used in conjunction with the TM Tibial Impactor to impact the tibial tray onto the tibia.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0430-2015·2014-12-03

    Intuitive Surgical Recalls 8mm Long Tip Forceps Due to Reprocessing Deviations

    Intuitive Surgical is recalling 509 units of 8mm Long Tip Forceps because deviations in reprocessing steps can cause surface degradation and accelerated mechanical wear.

    Product
    8MM,LONG TIP FORCEPS; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-0436-2015·2014-12-03

    Intuitive Surgical Recalls 8mm Permanent Cautery Hooks Due to Reprocessing Deviations

    Intuitive Surgical is recalling 509 8mm Permanent Cautery Hooks because deviations in reprocessing steps can cause surface degradation and accelerated mechanical wear.

    Product
    8MM PERMANENT CAUTERY HOOK,; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-2145-2014·2014-12-03

    Market-Tiers recalls laparoscopic morcellator instructions with missing safety pages

    Market-Tiers Inc. is recalling 100 units of the blue endo MOREsolution Electromechanical Laparoscopic Morcellator System because the Instructions for Use omitted critical safety pages.

    Product
    Instructions for Use for the blue endo MOREsolution Electromechanical Laparoscopic Morcellator System, Model 55-000011, consisting of the control unit, hand-grip, flexible shaft, trocars/obturators, sleeves, valves, and blades.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0450-2015·2014-12-03

    Philips BrightView Ultrasound Systems Recalled for Software Issues

    Philips Medical Systems is recalling 85 BrightView ultrasound systems worldwide due to software issues. The voluntary recall was initiated in November 2014.

    Product
    BrightView product code: 882478 BrightView X product code: 882480 BrightView XCT product code: 882482 and 882454
    Category
    Medical Device
    Distribution
    17 states
  • ModerateFDA (Devices)·Z-0438-2015·2014-12-03

    Intuitive Surgical Recalls 8mm Fenestrated Bipolar Forceps Due to Reprocessing Deviations

    Intuitive Surgical is recalling 509 units of 8mm Fenestrated Bipolar Forceps because deviations in reprocessing steps can cause surface degradation and accelerated mechanical wear.

    Product
    8MM FENESTRATED BIPOLAR FORCEPS,; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-0448-2015·2014-12-03

    Intuitive Surgical Recalls 8mm Tip-Up Fenestrated Graspers for Reprocessing Deviations

    Intuitive Surgical is recalling 509 units of 8mm Tip-Up Fenestrated Graspers because deviations in reprocessing steps can cause surface degradation and mechanical wear.

    Product
    8MM, TIP-UP FENESTRATED GRASPER; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-0466-2015·2014-12-03

    Remel Recalls Oxoid Legionella Latex Test Due to False Negative Results

    Remel Inc is recalling Oxoid Legionella Latex Test kits because a reagent may produce false negative results. The recall covers 59 boxes distributed nationwide.

    Product
    Oxoid Legionella Latex Test, DR0800M, containing 50 tests per box. The box contains a variety of components, one of which is Oxoid L. pneumophila Latex Reagent, DR0802, Lot 1301912. The responsible firm on the label is Oxoid Ltd.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0445-2015·2014-12-03

    Intuitive Surgical Recalls 8mm Small Grasping Retractor for Reprocessing Deviations

    Intuitive Surgical is recalling 509 units of the 8mm Small Grasping Retractor because deviations in reprocessing steps can cause surface degradation and mechanical wear.

    Product
    8MM, SMALL GRASPING RETRACTOR ; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-0420-2015·2014-12-03

    Siemens Hemoglobin A1c Reagent Cartridges Recalled for Positive Bias

    Siemens is recalling certain lots of Dimension HB1C Flex reagent cartridges because they show a positive bias in hemoglobin A1c measurements.

    Product
    Siemens Dimension HB1C Flex reagent cartridge; Dimension Hemoglobin A1c (HB1C) Flex reagent cartridge on the Dimension clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood.…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0483-2015·2014-12-03

    Beevers Cannulaide Respirator Recalled for Improper Pouch Seal

    Beevers Manufacturing recalled Cannulaide respirators due to improper pouch seals that could affect device shelf life.

    Product
    Cannulaide, Model 101, 102, 103. Respirator. The Cannulaide product is composed of die-cut hydrocolloid-coated polyurethane film with an integral hook strip (i.e., Velcro) to allow for the use of a loop material to help secure a nasal interface. The Cannulaide is…
    Category
    Medical Device
    Distribution
    7 states
  • ModerateFDA (Devices)·Z-0449-2015·2014-12-03

    Remel TB Potassium Permanganate Stain Recalled for Performance Issues

    Remel Inc is recalling TB Potassium Permanganate stain due to potential cloudiness or discoloration that may affect mycobacterial diagnosis.

    Product
    Remel TB Potassium Permanganate (.5% Aqueous) 40192, packaged in 250 ml bottles, 5/pkg. A stain for use in qualitative procedures as a counterstain in the fluorescent microscopic detection of mycobacteria.
    Category
    Medical Device
    Distribution
    16 states
  • ModerateFDA (Devices)·Z-0444-2015·2014-12-03

    Intuitive Surgical Recalls 8mm MEGA SUTURECUT ND Instruments

    Intuitive Surgical is recalling 509 units of 8mm MEGA SUTURECUT ND instruments due to potential surface degradation and accelerated mechanical wear.

    Product
    8MM, MEGA SUTURECUT ND; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
    Category
    Medical Device
    Distribution
    8 states
  • LowFDA (Devices)·Z-0469-2015·2014-12-03

    Novadaq LUNA Chest Recalled for Sterile Labeling Error

    Novadaq Technologies is recalling 828 LUNA Chest units because non-sterile drapes were inadvertently mislabeled as sterile.

    Product
    LUNA Chest, Model No. LU4006
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • LowFDA (Devices)·Z-0463-2015·2014-12-03

    Abbott m2000sp Software Incompatibility Recall for Nucleic Acid Testing

    Abbott Molecular is recalling m2000sp CD-ROMs because some Application Specifications are incompatible with software versions 6.0 and 7.0, causing error code 9000.

    Product
    Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • LowFDA (Devices)·Z-0470-2015·2014-12-03

    Novadaq Non-Sterile Drape Recalled for Incorrect Sterile Labeling

    Novadaq Technologies is recalling 324 non-sterile LUNA drapes that were inadvertently mislabeled with an additional contents sterile label.

    Product
    Non-sterile drape
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0220-2015·2014-11-26

    Biomet Spine Recalls Durango Anchored ALIF Plate Due to Set Screw Risk

    Biomet Spine is voluntarily recalling 120 Durango Anchored ALIF Plate units because the set screw component may become dislodged from the plate assembly.

    Product
    LANX, REF, PCR8XXX-XXX, LOT, RXONLY. Packaged within aluminum sterilization case. Product Usage: The Durango implant is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) at one or…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0223-2015·2014-11-26

    Nidek Excimer Laser System Recalled Due to Fuse Holder Contact Failure

    Nidek Inc is recalling 14 EC-5000 CXIII Excimer Laser Corneal Surgery Systems because a contact failure in fuse holders could cause serious adverse events for patients.

    Product
    Nidek Excimer Laser Corneal Surgery System, model EC-5000 CXIII. Opthalmic Device. The system is composed as a laser generator which produces an excimer laser radiation of wavelength 193nm, a beam delivery unit, an optical system for observation, a gas system and a computer for…
    Category
    Medical Device
    Distribution
    10 states
  • HighFDA (Devices)·Z-1578-2014·2014-11-26

    Medtronic Recalls Sterile Mouse Bags Due to Sterility Control Issues

    Medtronic Navigation is recalling sterile mouse bags because returned devices may not have been properly examined for damage that could compromise sterility.

    Product
    BAG 9732315 MOUSE STERILE 15 PK
    Category
    Medical Device
    Distribution
    Distributed nationwide

Looking for a different category? Browse all recalls.