Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

30601–30625 of 40489

  • LowFDA (Devices)·Z-1390-2016·2016-04-20

    BD Retic-Count Reagent System Recalled for Incorrect Storage Labeling

    Becton, Dickinson and Company is recalling one lot of BD Retic-Count Reticulocyte Reagent System because the carton label listed incorrect storage conditions.

    Product
    BD-Retic-Count Reticulocyte Reagent System; Hematology: Used for in vitro diagnostics
    Category
    Medical Device
    Distribution
    10 states
  • CriticalFDA (Devices)·Z-1275-2016·2016-04-13

    FDA Recalls Guardian II Hemostasis Valves Due to Air Embolism Risk

    Vascular Solutions is recalling Guardian II Hemostasis Valves because the low-pressure seal may fail, potentially allowing air into the device and causing an air embolism.

    Product
    Guardian II Hemostasis Valve, models 8210 and 8211. The following models are only available Outside the US: FH101, FH101-T, FH101-25, and FH101-50. Product Usage: The Guardian hemostasis valves are designed to be used as conduits when interventional devices with…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1272-2016·2016-04-13

    Focus Diagnostics Recalls Simplexa HSV 1&2 Direct Amplification Discs

    Focus Diagnostics is recalling Simplexa HSV 1&2 Direct Amplification Discs because they may cause false negative or positive results and error codes.

    Product
    Simplexa HSV 1&2 Direct (MOL2150) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model Number: MOL2150, MOL1451, MOL1452, MOL1455
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1274-2016·2016-04-13

    Focus Diagnostics Recalls Simplexa Group A Strep Direct Amplification Discs

    Focus Diagnostics is recalling Simplexa Group A Strep Direct Amplification Discs because they may cause false test results or error codes.

    Product
    Simplexa Group A Strep Direct (MOL2850) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model Number: MOL2850, MOL1451, MOL1452, MOL1455
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1327-2016·2016-04-13

    Synthes Recalls Titanium Low Profile Neuro Screws Due to Thread Defect

    Synthes (USA) Products LLC is recalling specific lots of Titanium Low Profile Neuro Screws because the thread height is undersized, potentially compromising the device's structural integrity.

    Product
    Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1323-2016·2016-04-13

    MicroAire French Rasp Aspiration Cannula Recalled for Bending or Breaking

    MicroAire Surgical Instruments is recalling 37 units of the PAL-R4011XL French Rasp aspiration cannula because they bent or broke during surgery.

    Product
    MicroAire PAL-R4011XL - 40cm x 4 mm Del Vecchio French Rasp aspiration cannula
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1315-2016·2016-04-13

    ScottCare TeleSentry Battery Recall Due to Explosion Risk During Recharging

    ScottCare is recalling 9,067 TeleSentry batteries after a complaint that one exploded while recharging. The batteries are used in cardiac monitoring devices.

    Product
    TeleSentry Battery, 3.7 Li-Ion Polymer Battery, CAUTION Do not expose to high temperature. Do not disassemble. Dispose of properly. Use specified charger only. 60pC/140pF The ScottCare Corporation. Intended for diagnostic evaluation of patients who experience transient…
    Category
    Medical Device
    Distribution
    24 states
  • HighFDA (Devices)·Z-1322-2016·2016-04-13

    Therakos CELLEX Photopheresis System Recalled for Tubing Leaks

    Therakos Inc is recalling 729 units of the CELLEX Photopheresis System, Lot D342, due to an increased trend in tubing leaks within the pump tubing organizer.

    Product
    The THERAKOS CELLEX Photopheresis System is a photoimmune therapy. Catalog CLXUSA. Manufacturing lot D342 UDI: 20705030200003 The system designed to separate the patient's blood into various fractions and collect the white blood cell (WBC) fraction. The WBC fraction is…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1343-2016·2016-04-13

    AngioDynamics Recalls NAMIC Convenience Kits Due to Needle Lumen Defect

    AngioDynamics is recalling 136 NAMIC Convenience Kits because the needle inner lumen is too small to accept the intended 0.038-inch guidewire.

    Product
    NAMIC CONVENIENCE KIT SELF REGIONAL HEALTHCARE PV LEG PACK, REF Catalog No. 651919311, UPN Product No. H7496519193111 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1324-2016·2016-04-13

    Fresenius Hemoflow F3 Dialyzer Recalled for Potential Header Leak

    Fresenius Medical Care is recalling 1,650 Hemoflow F3 Dialyzers due to a potential header leak caused by improper torque.

    Product
    Fresenius Hemoflow F3 Dialyzer Part Number: 0520165A Indicated for Acute and Chronic Hemodialysis
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1317-2016·2016-04-13

    Philips Recalls eValueMed Infant Heel Warmers and Transport Mattresses

    Philips is recalling eValueMed infant heel warmers and transport mattresses because they lack expiration dates and warnings about overheating risks.

    Product
    eValueMed Infant Transport Mattress 301-1015, Distributed by Tri-anima A disposable heel warmer for use when circulation needs to be stimulated in the infant heel in order for blood sampling to occur
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1344-2016·2016-04-13

    AngioDynamics Recalls Renown Regional Medical Center Convenience Kits

    AngioDynamics is recalling 100 Renown Regional Medical Center Convenience Kits because the needle inner lumen is too small to accept the intended guidewire.

    Product
    NAMIC CONVENIENCE KIT RENOWN REGIONAL MEDICAL CENTER, REF Catalog No. 65230029, UPN Product No. H965652300290, Product is packed as a box of 30 pieces, box label contains UPN Product No. H965652300291 Product Usage - This needle is intended to be placed over the guidewire to…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1348-2016·2016-04-13

    Ortho-Clinical VITROS 5600 Integrated System Recalled for Wiring Error

    Ortho-Clinical Diagnostics is recalling VITROS 5600 Integrated Systems because a wiring error may leave a module energized when the power cord is unplugged.

    Product
    VITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & 10758750007110. Intended for use in the in vitro measurement of a variety of analytes of clinical interest.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1346-2016·2016-04-13

    AngioDynamics Recalls NAMIC Convenience Kits Due to Needle Lumen Defect

    AngioDynamics is recalling 45 NAMIC Convenience Kits because the needle inner lumen may be too small to accept a 0.038-inch guidewire.

    Product
    NAMIC CONVENIENCE KIT MERCY GENERAL HOSPITAL NEURO KIT, REF Catalog No. 601319810, UPN Product No. H7496013198101 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1316-2016·2016-04-13

    Recall of eValueMed Infant Heel Warmers and Transport Mattresses

    Philips is recalling eValueMed Infant Heel Warmers and Transport Mattresses because they lack expiration dates and warnings about overheating risks after shelf life.

    Product
    evaluemed Infant Heel Warmer 301-1223; Distributed by Tri-anim A disposable heel warmer for use when circulation needs to be stimulated in the infant heel in order for blood sampling to occur
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1326-2016·2016-04-13

    Synthes Recalls Titanium Low Profile Neuro Screws Due to Thread Defect

    Synthes (USA) Products LLC is recalling specific lots of Titanium Low Profile Neuro Screws because the thread height is undersized, which may affect the device's structural integrity.

    Product
    Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1347-2016·2016-04-13

    GE Medical Systems Recalls SIGNA Pioneer MR Systems Due to Ferrous Material Risk

    GE Medical Systems is recalling 16 SIGNA Pioneer MR systems because ferrous material in the Table Control Box could cause unexpected attraction to the magnetic field during servicing, posing a risk of serious injury to personnel.

    Product
    SIGNA Pioneer MR Systems Product Usage: The SIGNA Pioneer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1273-2016·2016-04-13

    Focus Diagnostics Recalls Simplexa Direct Amplification Discs for Diagnostic Errors

    Focus Diagnostics is recalling Simplexa Direct Amplification Discs because they may cause false test results or error codes.

    Product
    Simplexa Flu A/B & RSV Direct (MOL2650) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model Number: MOL2650, MOL1451, MOL1452, MOL1455
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1270-2016·2016-04-13

    Laserex Systems Recalls Ellex Medical Fixed Eye Safety Filters

    Laserex Systems Inc. is recalling 74 Ellex Medical Fixed Eye Safety Filters due to a design error in the optic adaptor that may compromise eye protection.

    Product
    Ellex Medical Fixed eye Safety filters Model Integre S LP5532. The filter is located in the binocular mount of the laser delivery head. The filter prevents transmission of treatment laser light to physician eyes while the aiming beam to remain visible. .
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1328-2016·2016-04-13

    MicroAire PAL-R4011XL Cannula Recalled Due to Breakage During Surgery

    MicroAire is recalling PAL-R4011XL cannulas because three reports indicated the devices bent or broke during surgical procedures.

    Product
    MicroAire PAL¿ Cannula Product Usage - Power Aspiration Device
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1345-2016·2016-04-13

    AngioDynamics Recalls NAMIC Convenience Kits Due to Needle Lumen Defect

    AngioDynamics is recalling 8 NAMIC Convenience Kits because the needle inner lumen is too small to accept the intended guidewire.

    Product
    NAMIC CONVENIENCE KIT HOLYOKE HOSPITAL SPECIAL PROCEDURE KIT, REF Catalog No. 60080094, UPN Product No. H749600800941 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1331-2016·2016-04-13

    Applied Medical Resources Recalls Epix Universal Clip Applier Kits

    Applied Medical Resources is recalling specific Epix Universal Clip Applier kits due to a potential loading mechanism nonconformance.

    Product
    STD KIT GK213, AME Standard Kit 15 Product Usage: The Epix disposable laparoscopic clip applier is indicated for ligation of tubular structures or vessels in laparoscopic and general surgical procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1320-2016·2016-04-13

    Cardinal Health Recalls Gomco Style Circumcision Clamp Trays Due to Labeling Error

    Cardinal Health is recalling Gomco Style Circumcision Clamp Trays because two lots were labeled as containing 1.1cm clamps but actually contained 1.3cm clamps.

    Product
    Gomco Style Circumcision Clamp Trays is an instrument used to compress the foreskin of the penis during circumcision of a male infant.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1349-2016·2016-04-13

    Siemens PACS Software Recall for Incorrect Distance Measurements

    Siemens Medical Solutions is recalling 49 PACS systems because they may display incorrect distance measurements when used with certain modalities and syngo Imaging.

    Product
    A Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images, including digital mammography images.
    Category
    Medical Device
    Distribution
    Distributed nationwide