The Recall Desk

Category

Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

28626–28650 of 40397

  • SevereFDA (Devices)·Z-0097-2017·2016-11-02

    Roche Recalls cobas 8000 Analyzers Due to Sharp Edge Injury Reports

    Roche is recalling cobas 8000 Modular Analyzers and e602 Immunoassay Analyzers after four reports of injury caused by a sharp edge on the sample rack tray.

    Product
    cobas 8000 Modular Analyzer Series c8000, Chemistry (Photometric, Discrete) for clinical use Modular e602: Immunoassay Analyzer Product Usage: The VS II system is a stand-alone, computer-controlled instrument designed to aliquot patient sample from a primary tube to a…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0300-2017·2016-11-02

    Ortho-Clinical Diagnostics Recalls enGen Laboratory Automation System Due to Sample Error

    Ortho-Clinical Diagnostics is recalling the enGen Laboratory Automation System because a software error may cause sample identification to be mismatched with assay results.

    Product
    enGen Laboratory Automation System (Product Code ENGEN) configured with Thermo Scientific Centrifuge Module and TCAutomation Software Version 3.6.1 and Below; IVD. --- Thermo Scientific Centrifuge Module has Product Code 952040-EG 6844097. The VITROS 5,1 FS Chemistry System…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0109-2017·2016-11-02

    Roche Recalls cobas p 312 Pre-Analytical System Calculator Modules

    Roche is recalling 9,014 cobas p 312 Pre-Analytical System calculator/data processing modules due to four reported injuries caused by a sharp edge on the sample rack tray.

    Product
    cobas p 312 Pre-Analytical System calculator/data processing module, for clinical use Product Usage: The cobas p471 system is an automatic centrifuge module for centrifuging primary sample tubes, which will then be transported to the output module.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0101-2017·2016-11-02

    Roche Recalls cobas p 701 Post-Analytical Units Due to Sharp Edges

    Roche is recalling 9,014 cobas p 701 Post-Analytical Units because a sharp edge on the sample rack tray caused four reported injuries.

    Product
    cobas p 701 Post-Analytical Unit calculator/data processing module, for clinical use Product Usage: The URISYS 2400 Analyzer is a fully automated urinalysis system intended for in vitro qualitative or semi-quantitative determination of urine analytes, including pH…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0107-2017·2016-11-02

    Roche Recalls cobas p 512 Pre-Analytical Instrument Due to Sharp Edge Injury Reports

    Roche is recalling 9,014 cobas p 512 Pre-Analytical Instruments after four reports of injury caused by a sharp edge on the sample rack tray.

    Product
    cobas p 512 Pre-Analytical Instrument calculator/data processing module, for clinical use Product Usage: The cobas 6800/8800 systems support an automated and integrated workflow to run Polymerase Chain Reaction (PCR) based Nucleic Acid Testing (NAT) to be applied in…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0105-2017·2016-11-02

    Roche Recalls cobas 8000 Sorting System Modules Due to Injury Reports

    Roche is recalling 9,014 cobas 8000 Sorting System calculator/data processing modules after four reports of injury caused by a sharp edge on the sample rack tray.

    Product
    RSD 800 Sorting System calculator/data processing module, for clinical use Product Usage: The cobas 8000 Modular Analyze Series is a fully automated system for Clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0102-2017·2016-11-02

    Roche Recalls cobas p 501 Post-Analytical Units Due to Sharp Edges

    Roche is recalling 9,014 cobas p 501 Post-Analytical Units because a sharp edge on the sample rack tray caused four reported injuries.

    Product
    cobas p 501 Post-Analytical Unit calculator/data processing module, for clinical use Product Usage: The MODULAR PRE-ANALYTICS system is a fully automated system for sample pre-processing, including centrifuging, aliquoting, and barcode labeling, prior to analysis by MODULAR…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0096-2017·2016-11-02

    Roche Recalls cobas 6000 Series Analyzers Due to Sharp Edge Injury Risk

    Roche is recalling cobas 6000 Series analyzers because the sharp edge of the sample rack tray center guide rail caused four reported injuries.

    Product
    cobas 6000 Series system c6000, Chemistry (Photometric, Discrete), for clinical Use. Modular e601: Immunoassay Analyzer Product Usage: The cobas p501/p701 instrument stores primary and secondary sample tubes. The tubes are stored in a controlled, refrigerated sample store.…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0117-2017·2016-11-02

    St Jude Medical Recalls ICD and CRT-D Devices for Battery Short Circuit

    St Jude Medical is recalling implantable cardiac defibrillators and cardiac resynchronization therapy defibrillators due to a battery defect that can cause premature depletion.

    Product
    Quadra Assura, Model No. CD3265-40, CD3265-40Q, CD3365-40C, CD3365-40Q. Quadra Assura MP, Model No. CD3269-40, CD3269-40Q, CD3369-40C. Fortify and Fortify Assura implantable cardioverter defibrillators (ICDS) are intended to provided ventricular anti-tachycardia pacing and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0115-2017·2016-11-02

    St Jude Medical Recalls Implantable Cardiac Defibrillators Due to Battery Defect

    St Jude Medical is recalling specific implantable cardiac defibrillators and cardiac resynchronization therapy defibrillators because a lithium cluster in the battery can cause a short circuit and premature battery depletion.

    Product
    Fortify VR, Model No. CD1231-40, CD1231-40Q. Fortify ST VR, Model No. CD1241-40, CD1241-40Q. Fortify Assura VR, Model No. CD1257-40, CD1257-40Q, CD1357-40C. Fortify Assura ST VR, Model No. CD1263-40, CD1263-40Q, CD1363-40, CD1363-40Q. Fortify DR, Model No. CD2231-40…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0108-2017·2016-11-02

    Roche Recalls cobas p 471 Centrifuge Units Due to Sharp Edge Injury Risk

    Roche is recalling 9,014 cobas p 471 centrifuge units because a sharp edge on the sample rack tray caused four reported injuries.

    Product
    cobas p 471 centrifuge unit calculator/data processing module, for clinical use Product Usage: The cobas p312 pre-analytical system is a computer controlled stand-alone system for sorting and decapping sample tubes. As an option, the system works with an online connection.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0099-2017·2016-11-02

    Roche URISYS 2400 Urine Analyzer Recalled for Sharp Edge Injury Risk

    Roche is recalling URISYS 2400 Urine Analyzers because a sharp edge on the sample rack tray caused four reported injuries.

    Product
    URISYS 2400 Urine Analyzer Automated Urinalysis System Product Usage: The RSD 800 is a computer-controlled stand-alone sorting system for sorting barcoded samples. The system has an online connection to the host.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0098-2017·2016-11-02

    Roche Modular Pre Analytics EVO/Plus Recall for Sharp Edge Injury Risk

    Roche is recalling Modular Pre Analytics EVO/Plus systems because a sharp edge on the sample rack tray caused four reported injuries.

    Product
    Modular Pre Analytics ( EVO/Plus) Product Usage: The RSD 800A is a computer-controlled stand-online system for decapping, aliquoting and sorting of primary tubes and for producing and sorting bar coded secondary tubes.. The system operates with an online connection to the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0100-2017·2016-11-02

    Roche COBAS INTEGRA 800 Analyzer Recalled Due to Sharp Edge Injury Reports

    Roche is recalling COBAS INTEGRA 800 analyzers after four reports of injury caused by the sharp edge of the sample rack tray.

    Product
    Roche COBAS INTEGRA 800 /800 CTS Analyzer Analyzer, Chemistry (Photometric, Discrete for Clinical use Product Usage: The system cobas p612 is a computer-controlled stand-alone system for aliquoting and sorting of barcoded sample tubes. The system operates with an online…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0095-2017·2016-11-02

    Roche Recalls cobas e411 Immunoassay Analyzers Due to Sharp Edge Injury Risk

    Roche is recalling 9,014 cobas e411 Immunoassay Analyzers because the sharp edge of the sample rack tray center guide rail caused four reported injuries.

    Product
    cobas e411 Immunoassay Analyzer e411: Immunoassay Analyzer Product Usage: The cobas p501/p701 instrument stores primary and secondary sample tubes. The tubes are stored in a controlled, refrigerated sample store. Automatic retrieval for add-on testing is possible at any…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0094-2017·2016-11-02

    Roche Modular Analytics Systems Recalled Due to Sharp Sample Rack Edge

    Roche is recalling Modular Analytics Combination Systems because a sharp edge on the sample rack tray caused four reported injuries. The cleaning procedure has been revised to clarify the process.

    Product
    Roche/Hitachi MODULAR Analytics Combination Systems Modular D/P: Clinical Chemistry Analyzer, Chemistry (Photometric, Discrete) for clinical use. Modular E170: Immunoassay Analyzer Product Usage: The Roche COBAS INTEGRA 800/800 CTS Analyzer is fully automated system for…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0110-2017·2016-11-02

    Roche Recalls cobas 6800/8800 Systems Due to Sharp Edge Injury Risk

    Roche is recalling 9,014 cobas 6800/8800 systems because a sharp edge on the sample rack tray caused four reported injuries.

    Product
    cobas 6800/8800 system Hepatitis Viral B DNA Detection Product Usage: The cobas p512 pre-analytical system is a computer controlled fully automatic system for sorting of open and closed barcoded and centrifuged sample tubes. It includes modules for registration and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0106-2017·2016-11-02

    Roche cobas p 612 Pre-Analytical Instrument Recall for Sharp Edge Injury

    Roche is recalling cobas p 612 Pre-Analytical Instruments because the sharp edge of the sample rack tray center guide rail caused four reported injuries.

    Product
    cobas p 612 Pre-Analytical Instrument calculator/data processing module, for clinical use Product Usage: The cobas 6000 series is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitative…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0104-2017·2016-11-02

    Roche RSD 800A Aliquoting System Recalled Due to Sharp Edge Injury Reports

    Roche is recalling 9,014 RSD 800A Aliquoting System calculator/data processing modules because four reports of injury were received from the sharp edge of the sample rack tray.

    Product
    RSD 800A Aliquoting System calculator/data processing module, for clinical use Product Usage: The Roche/Hitachi MODULAR Analytics System is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0316-2017·2016-11-02

    Aesculap Safil Suture Recall Due to Needle Detachment Risk

    Aesculap is recalling 180 units of Safil QUI + UNDY 5/0 sutures because the needle may detach from the thread during surgery.

    Product
    Safil QUI + UNDY 5/0 (1) 45CM DSMP13 Suture
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0314-2017·2016-11-02

    BBI Solutions Recalls SimpliRED D-Dimer Assay Kits Due to Performance Issues

    BBI Solutions is recalling SimpliRED D-Dimer assay kits because stability testing showed reduced sensitivity, which may cause false negative results and affect clinical decisions.

    Product
    SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin degradation products containing the cross-linked D-diner site in human whole blood
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0304-2017·2016-11-02

    Nova Medical Recalls Journey Rolling Walkers Due to Loose Screws

    Nova Medical is recalling 1,380 Journey Rolling Walkers because screws holding the side supports can loosen and fall out.

    Product
    4205 Journey Rolling Walker Product Usage: used to assist people in walking
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0305-2017·2016-11-02

    Hospira Symbiq Infuser Recall for Potential White Screen Error

    Hospira is recalling Symbiq One Channel Infusers because a white screen during KVO titration may allow premature programming, potentially causing therapy delays and significant injury.

    Product
    Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids Device is a general purpose infusion pump designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0114-2017·2016-11-02

    Sequent Medical Recalls VIA 27 Microcatheter Due to Re-sheathing Defect

    Sequent Medical is recalling 618 units of the VIA 27 Microcatheter because the device may not adequately re-sheath interventional devices as designed.

    Product
    VIA 27 Microcatheter, Part No. FG27-154-01, Model No. VIA-27-154-01 The VIA 27 is intended for the introduction of non-liquid interventional devices (such as stents/flow diverters) and infusion of diagnostic (such as contrast media) or therapeutic agents into the neuro…
    Category
    Medical Device
    Distribution
    11 states
  • HighFDA (Devices)·Z-0103-2017·2016-11-02

    Roche VS II Aliquoter System recalled due to sharp edge injury risk

    Roche is recalling VS II Aliquoter System calculator modules because a sharp edge on the sample rack tray caused four reported injuries.

    Product
    VS II Aliquoter System calculator/data processing module, for clinical use Product Usage: Elecysy 411 analyzer is a fully automated, random access, computer controlled analytical systems for quantitative and qualitative determinations of analytes in body fluids.
    Category
    Medical Device
    Distribution
    Distributed nationwide

Looking for a different category? Browse all recalls.