The Recall Desk

Category

Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

28051–28075 of 40397

  • HighFDA (Devices)·Z-0816-2017·2016-12-28

    Medtronic Arrive Braided Transseptal Sheath Recalled for Liner Fragmentation

    Oscor, Inc. is recalling Medtronic Arrive Braided Transseptal Sheaths because a fragment of the inner liner may come out during insertion.

    Product
    MEDTRONIC ARRIVE BRAIDED TRANSSEPTAL SHEATH 10F C70¿ S79CM D84CM GLOBAL Catalog # 990079-070-000 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0900-2017·2016-12-28

    Exactech Recalls Equinoxe Reverse Shoulder Screw Kits Due to Disengaged Heads

    Exactech, Inc. is recalling 2,278 Equinoxe Reverse Shoulder Fixed Angle Torque Defining Screw Kits because disengaged screw heads may not be retained in the screwdriver.

    Product
    Equinoxe Reverse Shoulder Fixed Angle Torque Defining Screw Kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0818-2017·2016-12-28

    Medtronic Arrive Braided Transseptal Sheath Recalled Due to Liner Fragmentation

    Oscor, Inc. is recalling 55 units of Medtronic Arrive Braided Transseptal Sheaths because a fragment of the inner liner detached during insertion.

    Product
    MEDTRONIC ARRIVE BRAIDED TRANSSEPTAL SHEATH 10F C90¿ S61CM D66CM GLOBAL Catalog # 990061-090-000 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0893-2017·2016-12-28

    Ansell Recalls Gammex Surgical Gloves Due to Latex Contamination

    Ansell Healthcare Products is recalling 15 cases of Gammex Non-Latex Surgical Gloves because they incorrectly contained natural rubber latex gloves.

    Product
    Gammex Non-Latex Sensitive Neoprene Surgical Gloves Ansell Lanka (pvt) Ltd. Biyagama Export Processing Zone, Biyagama, Sri Lanka Ansell Healthcare Europe, N.V Internationaleiaan 55, 1070 Brussels, Belgium Manufactured for/ Fabrique' pour: Ansell Healthcare products…
    Category
    Medical Device
    Distribution
    5 states
  • HighFDA (Devices)·Z-0805-2017·2016-12-28

    Oscor Medical Sheath Recalled Due to Inner Liner Fragmentation

    Oscor, Inc. is recalling 110 units of the Adelante Breezeway sheath because a fragment of the inner liner detached during insertion.

    Product
    OSCOR ADELANTE BREEZEWAY 10F C90¿ S79CM D84CM Catalog # AB101078 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0792-2017·2016-12-28

    Oscor Medical Sheath Recalled Due to Inner Liner Fragmentation

    Oscor, Inc. is recalling 30 units of the Adelante Breezeway sheath because a fragment of the inner liner detached during insertion.

    Product
    OSCOR ADELANTE BREEZEWAY 8F C90¿ S61CM D66CM Catalog # AB081042 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0796-2017·2016-12-28

    Oscor Medical Sheath Recall Due to Inner Liner Fragmentation

    Oscor, Inc. is recalling 61 units of the Adelante Breezeway sheath because a fragment of the inner liner detached during insertion. The product was distributed in the US and internationally.

    Product
    OSCOR ADELANTE BREEZEWAY 8F C90¿ S79CM D84CM Catalog # AB081046 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0815-2017·2016-12-28

    Medtronic Arrive Braided Transseptal Sheath Recalled for Fragment Release

    Oscor, Inc. is recalling 59 units of Medtronic Arrive Braided Transseptal Sheaths because a fragment of the inner liner detached during insertion.

    Product
    MEDTRONIC ARRIVE BRAIDED TRANSSEPTAL SHEATH 10F C55¿ S79CM D84CM GLOBAL Catalog # 990079-055-000 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0885-2017·2016-12-28

    Siemens IMMULITE 2000 PSA Assay Recalled for Hook Effect Performance Issues

    Siemens Healthcare Diagnostics is recalling IMMULITE 2000 PSA assays because they do not meet the High-dose Hook Effect expectations stated in the Instructions for Use.

    Product
    IMMULITE 2000/IMMULITE 2000 XPi Systems Prostate-Specific Antigen (PSA) Assay, REF/Catalog Number L2KPTS2, SMN 10706281 and REF/Catalog Number L2KPTS6, SMN 10706282 (US); REF/Catalog Number L2KPS2, SMN 10380986 and REF/Catalog Number L2KPS6, SMN 10380996 (OUS); IVD --- Origin…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0791-2017·2016-12-28

    Oscor Medical Sheath Recalled for Inner Liner Fragmentation During Insertion

    Oscor, Inc. is recalling 87 units of the Adelante Breezeway sheath because a fragment of the inner liner detached during insertion.

    Product
    OSCOR ADELANTE BREEZEWAY 8F C70¿ S61CM D66CM Catalog # AB081041 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0775-2017·2016-12-28

    AngioDynamics Recalls Stiffened Micro-Introducer Kits Due to Detachment Risk

    AngioDynamics is recalling 996 boxes of Stiffened Micro-Introducer Kits because the rotating luer on the included sheath/dilator components may detach during use.

    Product
    ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO B PG, Item Number H787065970515, Catalog No.REF 06597051, 2) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO S PG, Item Number H787065970535, Catalog…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0804-2017·2016-12-28

    Oscor Medical Sheath Recalled Due to Inner Liner Fragmentation During Insertion

    Oscor, Inc. is recalling 124 units of the Adelante Breezeway 10F sheath because a fragment of the inner liner detached during insertion. The product was distributed worldwide.

    Product
    OSCOR ADELANTE BREEZEWAY 10F C70¿ S79CM D84CM Catalog # AB101077 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0817-2017·2016-12-28

    Oscor Recalls Medtronic Arrive Braided Transseptal Sheaths Due to Liner Fragmentation

    Oscor, Inc. is recalling 991 units of Medtronic Arrive Braided Transseptal Sheaths because a fragment of the inner liner detached during insertion.

    Product
    MEDTRONIC ARRIVE BRAIDED TRANSSEPTAL SHEATH 10F C90¿ S79CM D84CM GLOBAL Catalog # 990079-090-000 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0808-2017·2016-12-28

    Oscor Medical Sheath Recall Due to Inner Liner Fragmentation During Insertion

    Oscor, Inc. is recalling 33 units of the Adelante Breezway 10F Straight sheath because a fragment of the inner liner detached during insertion.

    Product
    OSCOR ADELANTE BREEZWAY 10F STRAIGHT S79CM D84CM Catalog # AB101219 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0777-2017·2016-12-28

    AngioDynamics Recalls Stiffened Micro-Introducer Kits Due to Detachment Risk

    AngioDynamics is recalling 7 boxes of Stiffened Micro-Introducer Kits because the rotating luer on the included sheath/dilator components may detach from the hub during use.

    Product
    ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T NON-ECHO B PG, Item Number H787065970685, Catalog No.REF 06597068, 2) 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T NON-ECHO S PG, Item Number H787065970705…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0798-2017·2016-12-28

    Oscor Medical Sheath Recall Due to Inner Liner Fragmentation During Insertion

    Oscor, Inc. is recalling 26 units of the Adelante Breezeway 8F Straight sheath because a fragment of the inner liner detached during insertion. The defect poses a risk of foreign body introduction into the patient.

    Product
    OSCOR ADELANTE BREEZEWAY 8F STRAIGHT S61CM D66CM Catalog # AB081210 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0814-2017·2016-12-28

    Medtronic Arrive Braided Transseptal Sheath Recalled for Inner Liner Fragmentation

    Oscor, Inc. is recalling 186 units of Medtronic Arrive Braided Transseptal Sheaths because a fragment of the inner liner was found during insertion.

    Product
    MEDTRONIC ARRIVE BRAIDED TRANSSEPTAL SHEATH 10F C120¿ S61CM D66CM GLOBAL Catalog # 990061-120-000 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0789-2017·2016-12-28

    Smiths Medical Recalls Acuvance 2 Safety Catheters Due to Valve Defects

    Smiths Medical is recalling 26,750 Acuvance 2 Safety Catheters because some devices have missing or incorrectly positioned side port valves.

    Product
    Acuvance¿2 Safety Catheter, Product No. 1720-AI Product Usage: The Acuvance¿ 2 Safety Catheter is a Peripheral Intravenous Catheter (PIVC). PIVCs are designed for single use venipuncture on all patient populations where access to the veins and arteries are required. The…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0786-2017·2016-12-28

    AbbVie Recalls 20 FR PEG Kits Due to Disconnection Reports

    AbbVie is issuing an urgent field correction for 20 FR PEG Kits due to increased reports of tube disconnections from Y-connectors.

    Product
    AbbVie PEG Kit 20 FR
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0773-2017·2016-12-28

    AngioDynamics Recalls Stiffened Micro-Introducer Kits Due to Component Defect

    AngioDynamics is recalling 80 boxes of Stiffened Micro-Introducer Kits because a component may detach during use.

    Product
    ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T ECHO B PG, Item Number H787065970465, Catalog No.REF 06597046, 2) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T ECHO S PG, Item Number H787065970485, Catalog…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0800-2017·2016-12-28

    Oscor Medical Sheath Recalled Due to Inner Liner Fragmentation

    Oscor, Inc. is recalling 22 units of the Adelante Breezeway sheath because a fragment of the inner liner detached during insertion.

    Product
    OSCOR ADELANTE BREEZEWAY 10F C55¿ S61CM D66CM Catalog # AB101072 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0801-2017·2016-12-28

    Oscor Medical Sheath Recalled Due to Inner Liner Fragmentation

    Oscor, Inc. is recalling specific lots of the Adelante Breezeway sheath because a fragment of the inner liner detached during insertion.

    Product
    OSCOR ADELANTE BREEZEWAY 10F C70¿ S61CM D66CM Catalog # AB101073 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Devices)·Z-0884-2017·2016-12-28

    Siemens IMMULITE PSA Assay Recalled for High-Dose Hook Effect Failure

    Siemens Healthcare Diagnostics is recalling IMMULITE PSA assays because they do not meet the high-dose hook effect expectations stated in the Instructions for Use.

    Product
    IMMULITE/IMMULITE 1000 Systems Prostate-Specific Antigen (PSA) Assay, REF/Catalog Number LKPTS1, Siemens Material Number (SMN) 10706279 (US); Catalog Number LKPS1, SMN 10380960 and Catalog Number LKPS5, SMN 10380949 (OUS); IVD --- Origin: UK Siemens Healthcare Diagnostics…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0890-2017·2016-12-28

    Medtronic Recalls Clearify Visualization Systems Due to Sterility Compromise

    Medtronic is recalling 230 Clearify Visualization Systems because the sterile barrier was breached, compromising product sterility. The recall covers worldwide distribution.

    Product
    Best Practices kits containing certain production lots of Clearify Visualization System Item Code: 21-345 Intended to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the scope as well as clean the scope lens
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0795-2017·2016-12-28

    Oscor Medical Sheath Recall Due to Inner Liner Fragmentation

    Oscor, Inc. is recalling 76 units of the Adelante Breezeway sheath because a fragment of the inner liner detached during insertion.

    Product
    OSCOR ADELANTE BREEZEWAY 8F C70¿ S79CM D84CM Catalog # AB081045 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
    Category
    Medical Device
    Distribution
    8 states

Looking for a different category? Browse all recalls.