The Recall Desk

Agency

U.S. Food and Drug Administration

The FDA regulates food, drugs, biologics, medical devices, and tobacco. We aggregate Class I, II, and III enforcement actions across its three recall streams.

Visit the official FDA site →

About FDA recalls

The FDA is the single biggest source of US product recalls by volume. Its enforcement workload splits into three streams — Foods (everything from infant formula to leafy greens to nutritional supplements), Drugs (prescription medicines, over-the-counter products, and active pharmaceutical ingredients), and Medical Devices (everything from implants to home blood-pressure monitors). The Recall Desk ingests all three from the openFDA enforcement endpoint every six hours.

FDA classifies every recall as Class I, II, or III. Class I means there is a reasonable probability that using the product will cause serious health consequences or death — these are the recalls you should treat as urgent. Class II means the product may cause temporary or medically reversible harm. Class III means the product is unlikely to cause adverse health consequences but still violates FDA rules. We map these classifications into our 1–5 severity score using the rubric on the methodology page.

Medical-device recalls are the highest-volume stream by far — software-update letters, calibration corrections, and IFU updates all show up in enforcement reports alongside genuine hazard recalls. We try to surface the severe ones first, but when in doubt, click through to the FDA notice itself for the clinical detail.

34001–34025 of 88304

  • HighFDA (Devices)·Z-1935-2020·2020-05-20

    Stryker Power-PRO TL Ambulance Cots Recalled for Crash Test Non-Compliance

    Stryker is recalling 34 Power-PRO TL ambulance cots in the UK because they may not meet the crash test standard BS EN 1789.

    Product
    Stryker Power-PRO TL cot-powered ambulance cot Model Number 6550 - Product Usage: is a powered wheeled stretcher that is intended to support and transport the entire body of a traumatized, ambulatory or non-ambulatory human patient (includes infants and adults).
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1966-2020·2020-05-20

    Magstim Recalls Rapid2 Stim Interconnecting Cables Due to Potential Failure

    The Magstim Company Limited is recalling 71 Rapid2 PSU Stim Interconnecting Cables because the high-voltage cable may fail. The recall covers units distributed worldwide.

    Product
    Rapid2 PSU Stim Interconnecting Cable - Product Usage: indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1978-2020·2020-05-20

    Access Scientific Recalls BD PosiFlush Saline Syringes Due to Packaging Holes

    Access Scientific LLC is recalling 30 units of BD PosiFlush SF Saline Flush Syringes because holes in the packaging compromised sterility.

    Product
    BD PosiFlush SF Saline Flush Syringe 10 mL used in POWERWAND XL 5Fr 8cm Quick Kit. REF/UDI / 94114/10859821006220 - Product Usage: is used to gain access to the vascular system to sample blood and administer fluids intravenously. It may be used for power injection of contrast…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1975-2020·2020-05-20

    Access Scientific Recalls BD PosiFlush Saline Syringes Due to Packaging Holes

    Access Scientific LLC is recalling 6,066 units of BD PosiFlush SF Saline Flush Syringes because holes in the packaging compromised sterility.

    Product
    BD PosiFlush SF Saline Flush Syringe 10 mL used in POWERWAND XL 4Fr 10cm Quick Kit. REF/UDI / 94105/10859821006190 - Product Usage: is used to gain access to the vascular system to sample blood and administer fluids intravenously. It may be used for power injection of contrast…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1940-2020·2020-05-20

    Cook Inc. Recalls Formula 418 Renal Balloon-Expandable Stents

    Cook Inc. is recalling 10 Formula 418 Renal Balloon-Expandable Stents because the stent may dislodge from the delivery system before deployment.

    Product
    Formula 418 Renal Balloon-Expandable Stent, RPN FOR418-18-80-6-12, Order Number (GPN) G56611 Balloon-expandable stent made of 316L stainless steel with a slotted tube configuration. It is pre-mounted on an over-the-wire balloon catheter, which serves as the delivery system.…
    Category
    Medical Device
    Distribution
    4 states
  • HighFDA (Devices)·Z-1984-2020·2020-05-20

    Dover Silicone Foley Catheter Recalled for Temperature Sensor Inaccuracy

    FDA recalls 468 Dover Foley catheters with temperature sensors that failed accuracy tests. Incorrect readings could lead to misdiagnosis.

    Product
    Dover 100% Silicone Foley Catheter with Temperature Sensor 14FR 5CC2W, Product Code: 50514T - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1987-2020·2020-05-20

    Dover Foley Catheter Recalled for Temperature Sensor Calibration Failures

    Covidien is recalling Dover Foley catheters with temperature sensors that failed to meet calibration tolerances, posing a risk of incorrect diagnosis.

    Product
    Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 18FR 5CCCC2W, Product Code: 50518ICT - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1990-2020·2020-05-20

    Dover Silicone Foley Catheter Recalled for Temperature Sensor Inaccuracy

    FDA is recalling Dover silicone Foley catheters because temperature sensors failed to meet calibration standards, risking incorrect patient diagnoses.

    Product
    Dover 400mL, 100% Silicone Premium Urine Meter Foley Tray, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P16TSD - Product Usage: for Drainage of the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1954-2020·2020-05-20

    Cardiac Assist Recalls TandemLung Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling two TandemLung Kit units because an assembly error may prevent the pump from priming.

    Product
    TandemHeart pump is assembled into kits: TandemLung Kit - DL31, Product: 5730-3118 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1952-2020·2020-05-20

    Cardiac Assist Recalls TandemHeart ProtekDuo Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling 46 TandemHeart ProtekDuo Kits due to an assembly error that may cause failure to prime.

    Product
    TandemHeart pump is assembled into kits: ProtekDuo Kit - DL3, Product: 5720-3118 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1979-2020·2020-05-20

    Access Scientific Recalls Saline Flush Syringes Due to Packaging Holes

    Access Scientific LLC is recalling 570 units of BD PosiFlush SF Saline Flush Syringes used in POWERWAND XL catheters because holes in the packaging compromised sterility.

    Product
    BD PosiFlush SF Saline Flush Syringe 10 mL used in POWERWAND XL 4Fr 10cm Extended Dwell Catheter. REF/UDI / 94108/00859821006773 - Product Usage: is used to gain access to the vascular system to sample blood and administer fluids intravenously. It may be used for power injection…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1983-2020·2020-05-20

    Access Scientific Recalls Saline Flush Syringes Due to Packaging Holes

    Access Scientific LLC is recalling 110 units of BD PosiFlush SF Saline Flush Syringes because holes in the packaging compromised sterility.

    Product
    BD PosiFlush SF Saline Flush Syringe 10 mL used in POWERWAND XL 5Fr 8cm Max Barrier Kit. REF/UDI / 94116/10859821006237 - Product Usage: is used to gain access to the vascular system to sample blood and administer fluids intravenously. It may be used for power injection of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1951-2020·2020-05-20

    Cardiac Assist Recalls 38 TandemHeart ProtekDuo Kits Due to Assembly Error

    Cardiac Assist, Inc. is recalling 38 TandemHeart ProtekDuo Kits because an assembly error may prevent the pump from priming.

    Product
    TandemHeart pump is assembled into kits: ProtekDuo Kit - DL29, Product: 5720-2916 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1967-2020·2020-05-20

    BD Alaris PCEA Administration Set Recalled for Potential Leaking

    BD is recalling 110,350 Alaris PCEA Administration Sets because they may leak at the luer connection, potentially causing inadequate pain control.

    Product
    BD Alaris PCEA Administration Set, Microbore Tubing with Yellow Identification Stripe ASV, Catalog Number 30893-07 - Product Usage: Microbore Tubing with Yellow Identification Stripe Anti-siphon Valve Use Aseptic Technique Set cannot be used for gravity infusion. 1. Connect…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1956-2020·2020-05-20

    Merit Medical Recalls Custom Sheath Introducer Kits Due to Labeling Error

    Merit Medical Systems is recalling 15 Custom Sheath Introducer Kits because they were incorrectly labeled and contained the wrong sheath assembly.

    Product
    Custom Sheath Introducer Kit REF K21-00031 - Product Usage: intended use is to provide access and facilitate the percutaneous introduction of various devices into vein.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2007-2020·2020-05-20

    Edwards Lifesciences Recalls Sterile Oxymetry Oligon Catheter Kits

    Edwards Lifesciences is recalling 597 sterile catheter kits because their packaging was compromised, potentially affecting sterility.

    Product
    Sterile Oxymetry Oligon Catheter Kit. Pressure injectable. REF/UDI:XA3816SAFKT/ (01)00690103197181
    Category
    Medical Device
    Distribution
    15 states
  • HighFDA (Devices)·Z-1963-2020·2020-05-20

    DeRoyal Sterile Custom Kits Recalled Due to Packaging Holes

    DeRoyal Industries is recalling sterile custom kits packaged with BD PosiFlush syringes because holes in the packaging may compromise sterility.

    Product
    DeRoyal Sterile Custom Kits packaged as a Piggy Back with the BD PosiFlush SF Saline Flush Syringe, labeled as follows: a. H* Vein Pack Pgybk, 89-6209.03 b. H* Vein Pack, 89-6209.04 c. PICC Line Tray Pgybk, 89-6150.12 d. Umbilical Arterial Cath, 89-9370.01 e.…
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-1998-2020·2020-05-20

    ICU Medical Recalls Central Venous Catheters Due to Guidewire Issue

    ICU Medical is recalling 10 Central Venous Catheters because the included guidewire cannot pass through the needles.

    Product
    Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 41401-25.
    Category
    Medical Device
    Distribution
    19 states
  • ModerateFDA (Devices)·Z-1964-2020·2020-05-20

    Bard PowerPort ClearVUE Slim Implantable Port Recalled for Incorrect Catheter Lock Size

    Bard Peripheral Vascular Inc is recalling 200 units of PowerPort ClearVUE Slim Implantable Ports because the catheter locks may be 8Fr instead of the labeled 6Fr.

    Product
    BARD Access Systems PowerPort ClearVUE Slim Implantable Port With Smooth Septum and Attachable 6 F Polyurethane Open-Ended Single-Lumen Venous Catheter, Product #5676300 - Product Usage: All materials are biocompatible, can be used with virtually all injectable solutions…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1999-2020·2020-05-20

    ICU Medical Recalls Central Venous Catheters Due to Guidewire Issue

    ICU Medical is recalling 60 units of Model 41406-23 Central Venous Catheters because the included guidewires cannot pass through the needles.

    Product
    Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 41406-23
    Category
    Medical Device
    Distribution
    19 states
  • ModerateFDA (Devices)·Z-1928-2020·2020-05-20

    Edwards HemoSphere Oximetry Cable Recalled for Inconsistent Connector Crimping

    Edwards Lifesciences is recalling 35 HemoSphere oximetry cables due to intermittent communication caused by inconsistent crimping of connectors.

    Product
    Oxymetry Cable, Rxonly. REF/UDI: HEMOXSC100/(01)00690103196986 - Product Usage: The HemoSphere oximetry cable is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Food)·F-0940-2020·2020-05-20

    Natural by Nature 1% Milk Recalled for Antibiotic Residue

    Natural Dairy Products is recalling Natural by Nature 1% Milk due to a sample testing positive for antibiotic residue. The product was distributed to seven states.

    Product
    Natural by Nature 1 % Milk (gallon, 1/2 gallon, and quart)
    Category
    Food
    Distribution
    7 states
  • ModerateFDA (Devices)·Z-1942-2020·2020-05-20

    Karl Storz Endoscopy Recalls Insufflation Tubing Due to Ethylene Oxide Residues

    Karl Storz Endoscopy is recalling sterile insufflation tubing because it may contain higher ethylene oxide residues than the acceptable level.

    Product
    Insufflation tubing to be used with varies insufflator units. REF 031200-10; REF 031322-10; REF 031532-10; REF 031622-10
    Category
    Medical Device
    Distribution
    27 states
  • ModerateFDA (Devices)·Z-1970-2020·2020-05-20

    C.R. Bard Recalls SureStep Foley Tray Systems Due to Residue

    C.R. Bard is recalling 8,392 SureStep Foley Tray Systems because a white paper-type residue may be attached to the catheter.

    Product
    SureStep Foley Tray System - Product Usage: Intended for use in the drainage and/or collection and/or measurement of urine. Generally, drainage is accomplished by inserting the catheter through the urethra and into the bladder. However, drainage is sometimes accomplished by…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2002-2020·2020-05-20

    ICU Medical Recalls Central Venous Catheters Due to Guidewire Issue

    ICU Medical is recalling 15 units of Central Venous Catheters because the included guidewire cannot pass through the needles.

    Product
    Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 50456-01
    Category
    Medical Device
    Distribution
    19 states